4 Conditions for production of X-rays
1) Source of electrons
2) Target
3) High-voltage
4) Vacuum
Thermionic Emisson
Electrical process of liberating electrons from wire filament
The filament is heated to a very high temperature due to creating resistance with small diameter
“Boiling off” electrons create electron cloud around filament
Principle of incandescents light bulb
Very important x-ray design
Grid-Based Tubes
Precise control of the thermionic cloud
Brief change of focusing cup from negative to positive
Permits very short, rapid sequencing of exposure
Commonly used in:
Angiography
Pulsed fluoroscopy
Tube Failure (cathode)
Tube arching (like short circuit)
Filament breakage
Tube arcing
Vaporized tungstem collection envelope
Filament breakage
‘Boost’ and hold vs. repeated tube boosting
Filaments become increasingly thin due to vaporization over time (too much heat)
Tube failure (Anode)
Ball bearings worn (overuse) at high temp.
Melting of anode disks (rotor failure)
Stator fails and rotor ceases to move
Tube failure (anode)(causes)
Overload (high temp)
Poor power conditions
Mechanical failure
Target Area
Portion of anode that electrons stream contacts
Referred to by several names
Point of source of X-ray photons
Target Area (several names)
Target, focus point, focal point, focal track
Production fo X-rays (4 basic things)
Source of electrons by thermionic emissions
Means of accelerating electrons (kVp)
Means of decelerating electrons (target)
A vacuum (controlled environment)
2 focal points
Actual focal spot
Effective focal spot
Actual focal spot
Physical area where electrons strike the target
Controlled by: length of filament + target angle
Small actual focal spot mA limited
Effective Focal spot
The area projected out of the tube toward the patient.
Controlled by actual focal spot
Spatial Resolution (Actual vs Effective focal spot)
Smaller focal spot = Increased (better) resolution
Lines Focus Principle
Relationship between actual focal point and effective focal point
Size of effective focal point depend on angle of anode
The most common angle of the target
12 degrees
Anode Heel Effect
Results in variation of x-ray beam intensity along longitudinal axis of x-ray beam (can vary 45%)
More intense under cathode side
Intenional positioning of body parts can take advantage of heel effect
Warm-up procedure
Only sometimes
Gradually warms up anode
Prevents craking glass envelope in older machines
Maintains vacuums
Envelope (functions)
Supports anode/cathode assemblies
Maintains a vacuum to avoid debris and disturbances
Envelope (Modern tubes now metal)
10” long
6” central diameter
2” peripheral diameter
Envelope (Tube window)
Area constructed for the X-ray beam to exit
Protective housing
Metallic lead-lined cast steal
Supports X-ray tube
Controls leakage and scatter radiation (lead-lined)
Isolates high voltage
Protective Housing (mechanisms to cool tube)
Cooling oil surroundings tube
Cooling fans
Water colling in high end tube designs
Off-Focus Radiation
Undesirable part of beam produced away from the target and therefore considered off-focus
Contributes up to 25% of the total primary beam
Produces ghosting of image and reduces image quality
Can be corrected with post-processing in digital radiography
Extending X-ray tube life
Follow tube warm-up procedures
Avoid frequency and successive boosting of tube
Use low mAs settings
Avoid rough handling of x-ray tube head
Listen for unusual sounds
Cathode Assembly (includes)
Filaments
Focusing cup
Associated wiring
Filament
Dual focus
Thermionic emissions
Coil of thoriated tungsten
Greater current (is radiative)
Provides resistance (to make heat)
Focusing Cup
Composed of nickel
Low negative potential applied
Compresses/narrow thermionic cloud (space charge)
Two working systems
Focusing cups (2 working systems)
Space charge effect
Saturation Current
Space Charge Effect
Congested electrons limiting current flow
Limits exposure miliamperes (mA) to 1000-1200
Saturation current
Typically controlled by X-ray circuitry (kVp)
Higher kVp gives a bigger push of electrons toward the anode
Eliminates the space charge effect
Anode Assembly (Functions)
Target surface for X-ray production
Conducts high voltage (closed circuit pathway)(recyles energy)
Serves as primary thermal conductor (allows heat to be safely conducted)
Anode assembly (components)
Anode (with target), stator, rotor
Rotating Anode (types)
Stationary anode (only used for low-energy machines) (e.g, dental units)
Rotating anode (modern X-ray machines
Rotating anode (Tungsten-rhenium alloy)
High atomic #
High melting point
Heat-conducting ability
the rotating anode is made of tungsten-rhenium
Rotor
Ferromagnetic bars arranged in cylindrcal pattern
Inside stator + envelope
Copper cylinder connected to anode disk by molybdenum stem
Turns when stator energized (3000-12000)
Balls bearings (inside rotor)
Silver plated
Reduces surface contact/friction
Stator
Wraps around rotor outside envelope
Electromegnets consisting of copper windings
Stator failure
Suboptimal anode rotation speed (rotor stops turning)
Immediate melting on target results from too much heat (bullet heat)