Ideal phantom
5 cm greater on sides
10 cm greater than FS in depth
Same physical and electron density
Electron Density
pe = pm * Na (Z/A)
Na = Avagadros nunumber 6.022×10²³
PDD equation
PDD = (Dd/ Ddmax) *100
PDD is dependent on
Energy
Field size
Increases with field size
Depth
SSD
Field Size
The distance between the 50% isodose line for a detector on the surface of a phantom 100 cm away from the source and beneath a layer of buildup material thickness roughly equal to dmax
TAR
= Dd/Df.s.
Measured under SAD conditions
TAR is dependent on
Energy
Field size
Depth
Independent of SSD
Back Scatter factor (BSF)
= Ddmax/Dfs at the level of dmax
= TAR at dmax
Scatter Air Ratio (SAR)
The ration of scattered dose at a given point in the phantom and the dose in free space at the same point.
Independent of the distance
Dependent on energy, depth, and field size
SAR(d,r) = TAR(t,r) - TAR (d,0)
Field Size
The width of the 50% point on a profile
Open field size
Refers to the open field size as defined by the collimators
Used for Sc table look up
Effective field size
Field size after blocking by either cerrobend of MLC
Used for Sp and TMR lookup
Collimator Scatter Factor (Sc)
represents the scatter contribution from the surface area of the collimator and the scatter from the head especially the exposure of the flattening filter
Increases with FS
Measured in air
Measured with build up cap
Phantom Scatter Factor (Sp)
Represents the scatter component from the phantom material (patient)
Cannot be measured
Increases with field size
Total Scatter Factor (Scp)
Measured at dmax
Output factors be defined at a depth of 10cm
Tissue Maximum Ratio (TMR)
= Dd/Ddmax under SAD conditions
Increases with
E
FS
Decreases with
Depth beyond the dmax depth
Independent of SSD
Scatter Maximum Ratio (SMR)
= TMR(total)- TMR(primary)
Off Axis Ratios (OARs)
Function of
Energy
Distance off central axis
Depth
Higher for lower energy at shallow depths
Jaw percent transmission
< 1%
MLC Intraleaf percent transmission
2%
MLC Interleaf percent transmission
4%
Cerrobend percent transmission
3-5%
Waterscanner Set-Up for profiles
Chambers
Depth planes
Speed
Profiles are obtained for
Open fields
Wedged fields
Penumbra
Dose in this region decreases rapidly
Dependent on
Source size
Distance from the source
Source to diaphragm distance
Isodose curves are measured by
Ion chambers
Solid state detectors
Film
Beam direction
Factors
Normal tissues
Inhomogeneties
Immobilization and Set-up
Should be separated by 15 degrees
Field size
The width of the 50% point on a profile
LINAC Calibrated SAD, then isocentric set-up
No not require a SSD correction
LINAC calibrated SAD, then SSD
Set ups do require a SSD correction
LINAC calibrated SSd, then isocentric (SAD)
Set ups do require a SSD correction
LINAC calibrated SSD, then SSD
Set ups do not require a SSD correction
ISL correction
(100/100+dmax)2
Collimators
Define the field dimensions
Transmission through is < 1%
Flattening filter
Make beam intensity uniform at a depth of 10 cm
Thickest in the middle
Hardens the CAX
Integral dose
Measure of total energy absorbed in treated volume
Increases for multiple field plans for lower energy and shorter SSD
Higher for IMRT
Number of fields
Increase will increase
Conformality
Integral dose
Need for wedges
Axial scanning
Better resolution
Longer scan times
Can overlap scans
Helical scanning
Can control the pitch
Resolution excelent
Air CT number
-1000
Lung CT number
-300
Fat CT number
-90
Water CT number
0
White matter CT number
30
Gray matter CT number
40
Muscle CT number 50
50
Bone CT number
1000
Window
Width (in CT numbers) of the ramp that is used to map CT numbers to brightness
As width decreases, the contrast in the image is enhaced
Level
CT number at the center of the window
MRI T1
Wait time is short
Relaxation time: rate of nuclei realignment (longitudinal)
Anatomical detail
MRI T2
Wait time is longer
Relaxation time: Rate of nuclei loss of synchronization with each other’s spin (Transverse)
Pathologic change
T1
Anatomical
T1 Gd
Anatomical
Contrast enhancing
T2
Pathological
PD
Proton density
Shows water concentration
fMRI
Function MRI
Sensitive to magnetic properties
Blood
MRI process summary
Applied Rf excites protons in tissue
Protons equilibrate and release Rf signals
Rf signals applied and Rf signals collected can be varied to create different images
Dark: Low signal collected
Bright: High signal collected
MRI advantages
Imaging acquisition in all planes
Transverse, sagittal, oblique
Superior for CNS
MRI disadvantages
Time acquisition
Motion artifacts
Geometric distortion at periphery
No electron density information
Positron emission tomography (PET)
Assess
Physiology
Biochemistry
Metabolism
Lymphatic superiority for cancer diagnosis
F-18 Half life
110 minutes
C-11 half life
20 minutes
N-13 half life
10 minutes
O-15 half life
2 min
PET principle of operation coincidence detection
511 keV photons
180 degrees
Timing coincidence
High SNR
PET Performance Factors
Detector normalization
Spatial resolution
Uniformity
Contrast
PET Resolution
4-5 mm
Ultrasound principles of operation
Pulse of ultra sonic is propagated into the patient from a transducer placed on the skin
Back-scattered echo signal is recorded by same transducer
Image is constructed from back scattered echo strength
kV Cone Beam CT Strengths
Large field of view
Quick acquisition time
Higher contrast than MV
Utilize contrast agents
Low imaging dose for high contrast areas
Independent imaging systems
kV Cone Beam CT Limitations
Shading artifacts from large scatter volumes
Clipping artifacts from metal implants
Motion artifact due to scan > 5 sec
MV Cone Beam CT Steps
Patient is positioned
200-240 degree arc scan
Reconstruction and registration
Alignment evaluation
Couch translations