Ideal phantom

  • 5 cm greater on sides

  • 10 cm greater than FS in depth

  • Same physical and electron density


2
New cards

Electron Density

pe = pm * Na (Z/A)

  • Na = Avagadros nunumber 6.022×10²³


3
New cards

PDD equation

PDD = (Dd/ Ddmax) *100

4
New cards

PDD is dependent on

  • Energy

  • Field size

    • Increases with field size

  • Depth

  • SSD


5
New cards

Field Size

The distance between the 50% isodose line for a detector on the surface of a phantom 100 cm away from the source and beneath a layer of buildup material thickness roughly equal to dmax

6
New cards

TAR

= Dd/Df.s.

  • Measured under SAD conditions


7
New cards

TAR is dependent on

  • Energy

  • Field size

  • Depth

  • Independent of SSD


8
New cards

Back Scatter factor (BSF)

= Ddmax/Dfs at the level of dmax

= TAR at dmax

9
New cards

Scatter Air Ratio (SAR)

The ration of scattered dose at a given point in the phantom and the dose in free space at the same point.

  • Independent of the distance

  • Dependent on energy, depth, and field size


SAR(d,r) = TAR(t,r) - TAR (d,0)


10
New cards

Field Size

The width of the 50% point on a profile

11
New cards

Open field size

Refers to the open field size as defined by the collimators

  • Used for Sc table look up


12
New cards

Effective field size

Field size after blocking by either cerrobend of MLC

  • Used for Sp and TMR lookup


13
New cards

Collimator Scatter Factor (Sc)

represents the scatter contribution from the surface area of the collimator and the scatter from the head especially the exposure of the flattening filter

  • Increases with FS

  • Measured in air

  • Measured with build up cap


14
New cards

Phantom Scatter Factor (Sp)

Represents the scatter component from the phantom material (patient)

  • Cannot be measured

  • Increases with field size


15
New cards

Total Scatter Factor (Scp)

  • Measured at dmax

  • Output factors be defined at a depth of 10cm


16
New cards

Tissue Maximum Ratio (TMR)

= Dd/Ddmax under SAD conditions

  • Increases with

    • E

    • FS

  • Decreases with

    • Depth beyond the dmax depth

  • Independent of SSD


17
New cards

Scatter Maximum Ratio (SMR)

= TMR(total)- TMR(primary)

18
New cards

Off Axis Ratios (OARs)

Function of

  • Energy

  • Distance off central axis

  • Depth

Higher for lower energy at shallow depths


19
New cards

Jaw percent transmission

< 1%

20
New cards

MLC Intraleaf percent transmission

2%

21
New cards

MLC Interleaf percent transmission

4%

22
New cards

Cerrobend percent transmission

3-5%

23
New cards

Waterscanner Set-Up for profiles

  • Chambers

  • Depth planes

  • Speed


24
New cards

Profiles are obtained for

  • Open fields

  • Wedged fields


25
New cards

Penumbra

  • Dose in this region decreases rapidly

  • Dependent on

    • Source size

    • Distance from the source

    • Source to diaphragm distance


26
New cards

Isodose curves are measured by

  • Ion chambers

  • Solid state detectors

  • Film


27
New cards

Beam direction

Factors

  • Normal tissues

  • Inhomogeneties

  • Immobilization and Set-up

Should be separated by 15 degrees

28
New cards

Field size

The width of the 50% point on a profile

29
New cards

LINAC Calibrated SAD, then isocentric set-up

No not require a SSD correction

30
New cards

LINAC calibrated SAD, then SSD

Set ups do require a SSD correction

31
New cards

LINAC calibrated SSd, then isocentric (SAD)

Set ups do require a SSD correction

32
New cards

LINAC calibrated SSD, then SSD

Set ups do not require a SSD correction

33
New cards

ISL correction

(100/100+dmax)2

34
New cards

Collimators

Define the field dimensions

  • Transmission through is < 1%


35
New cards

Flattening filter

Make beam intensity uniform at a depth of 10 cm

  • Thickest in the middle

  • Hardens the CAX


36
New cards

Integral dose

Measure of total energy absorbed in treated volume

  • Increases for multiple field plans for lower energy and shorter SSD

  • Higher for IMRT


37
New cards

Number of fields

Increase will increase

  • Conformality

  • Integral dose

  • Need for wedges


38
New cards

Axial scanning

  • Better resolution

  • Longer scan times

  • Can overlap scans


39
New cards

Helical scanning

  • Can control the pitch

  • Resolution excelent


40
New cards

Air CT number

-1000

41
New cards

Lung CT number

-300

42
New cards

Fat CT number

-90

43
New cards

Water CT number

0

44
New cards

White matter CT number

30

45
New cards

Gray matter CT number

40

46
New cards

Muscle CT number 50

50

47
New cards

Bone CT number

1000

48
New cards

Window

Width (in CT numbers) of the ramp that is used to map CT numbers to brightness

  • As width decreases, the contrast in the image is enhaced


49
New cards

Level

CT number at the center of the window

50
New cards

MRI T1

  • Wait time is short

  • Relaxation time: rate of nuclei realignment (longitudinal)

  • Anatomical detail


51
New cards

MRI T2

  • Wait time is longer

  • Relaxation time: Rate of nuclei loss of synchronization with each other’s spin (Transverse)

  • Pathologic change


52
New cards

T1

Anatomical

53
New cards

T1 Gd

  • Anatomical

  • Contrast enhancing


54
New cards

T2

Pathological

55
New cards

PD

  • Proton density

  • Shows water concentration


56
New cards

fMRI

  • Function MRI

  • Sensitive to magnetic properties

  • Blood


57
New cards

MRI process summary

  • Applied Rf excites protons in tissue

  • Protons equilibrate and release Rf signals

  • Rf signals applied and Rf signals collected can be varied to create different images

  • Dark: Low signal collected

  • Bright: High signal collected


58
New cards

MRI advantages

  • Imaging acquisition in all planes

    • Transverse, sagittal, oblique

  • Superior for CNS


59
New cards

MRI disadvantages

  • Time acquisition

  • Motion artifacts

  • Geometric distortion at periphery

  • No electron density information


60
New cards

Positron emission tomography (PET)

Assess

  • Physiology

  • Biochemistry

  • Metabolism

Lymphatic superiority for cancer diagnosis


61
New cards

F-18 Half life

110 minutes

62
New cards

C-11 half life

20 minutes

63
New cards

N-13 half life

10 minutes

64
New cards

O-15 half life

2 min

65
New cards

PET principle of operation coincidence detection

  • 511 keV photons

  • 180 degrees

  • Timing coincidence

  • High SNR


66
New cards

PET Performance Factors

  • Detector normalization

  • Spatial resolution

  • Uniformity

  • Contrast


67
New cards

PET Resolution

4-5 mm

68
New cards

Ultrasound principles of operation

  • Pulse of ultra sonic is propagated into the patient from a transducer placed on the skin

  • Back-scattered echo signal is recorded by same transducer

  • Image is constructed from back scattered echo strength


69
New cards

kV Cone Beam CT Strengths

  • Large field of view

  • Quick acquisition time

  • Higher contrast than MV

  • Utilize contrast agents

  • Low imaging dose for high contrast areas

  • Independent imaging systems


70
New cards

kV Cone Beam CT Limitations

  • Shading artifacts from large scatter volumes

  • Clipping artifacts from metal implants

  • Motion artifact due to scan > 5 sec


71
New cards

MV Cone Beam CT Steps

  • Patient is positioned

  • 200-240 degree arc scan

  • Reconstruction and registration

  • Alignment evaluation

  • Couch translations