1/34
Looks like no tags are added yet.
When using pulsed wave (PW) ultrasound, the frequency of the ultrasound is dependent upon ____________.
thickness of the crystal
speed of sound in through the crystal
the characteristics of the crystal
degree of dampening of the crystal
The correct answer is:
the characteristics of the crystal
When using continuous wave (CW) ultrasound, the frequency of the ultrasound is equal to the ____________.
speed of sound in through the crystal
thickness of the crystal
degree of dampening of the crystal
frequency of the electrical pulse
The correct answer is:
frequency of the electrical pulse
3. A multi-dimensional array uses electrical delays to optimize focus in the ________________ plane.
elevational
lateral and elevational
axial
lateral
Remember: Control of elevational plane is optimized with this technology.
The correct answer is:
lateral and elevational
4. What is the primary importance of the backing material?
Prevents excessive “ringing” of the crystal
Isolates the patient from an electrical signal
Allows improved transmission of the sound energy into the tissues
Limits outside electrical interference
Remember: Remember, a short pulse is a good pulse because it improves resolution.
The correct answer is:
Prevents excessive “ringing” of the crystal
*This allows a more significantly decreased pulse length.
A curvilinear array transducer uses a(n) ___________ focusing technique.
internal fixed
sequential
sonographer adjustable
electronic delay
Remember: Virtually all of todays transducers use electronic delays (phased) to focus the beam.
The correct answer is:
electronic delay

6. In the figure, the arrow points to the _____________________.
backing material
matching layer
electrical shielding
acoustical isolator
The correct answer is:
matching layer
*This has an acoustical impedance between the crystal and the skin and allows a more efficient transmission of the ultrasound energy into the tissues.

7. A 2-D transducer allows an instrument to simultaneously create an image array ___________.
using a computerized reconstruction technique
in real time
that simultaneously acquires B-mode and Doppler
in both a sagittal and transverse planes
Remember: 2D transducers can make 3D images, 1D transducers are usually used to make 2D images.
The correct answer is:
in both a sagittal and transverse planes

8. A _______________ transducer uses one (1) or a few piezoelectric crystals that are moved or oscillated to create a sector shaped image?
Question 8Answer
mechanical array
annular array
sequential array
curvilinear array
Remember: This was the original technology used to create a 2-D image but rarely used today.
The correct answer is:
mechanical array
A 4-D transducer acquires data using __________.
a post processing function
a 3 dimensional format in real-time
a highly focused picture element (pixel)
multiple areas within the scan line virtually simultaneously
Remember: 3D acquires data from all three planes and the 4th dimension is real time.
The correct answer is:
a 3 dimensional format in real-time
10. A 3-dimensional array acquires will have _________.
the best resolution
a slower frame rate
a very high frame rate
the poorest resolution
Remember: The acquistion of large numbers of sample volumes or data points will reduce frame rate.
The correct answer is:
a slower frame rate
11. What is bandwidth?
The range of frequencies within a pulse of ultrasound
The length of time the transducer crystal vibrates
The width or time of the electrical pulse
The range of frequencies returning to the transducer
Remember: A crystal does not emit a single frequency but actually a range of frequencies above and below the center frequency. This range is called the bandwidth.
The correct answer is:
The range of frequencies within a pulse of ultrasound
12. A continuous wave Doppler transducer would provide optimal ability to detect and display
low velocity blood flow
discrete blood flow in a large vessel
discrete blood flow in a small vessel
high velocity flow
Remember: The CW Doppler has a sample volume along the entire line of sight and will not alias.
The correct answer is:
high velocity flow
13. The property of a material that allows conversion of a mechanical compression to an electrical signal is referred to as the _____________.
Huygens principle
electromechanical effect
hydraulic pressure effect
piezoelectric effect
Remember: This special property allows a material to vibrate when an electrical signal is applied and also generates an eletrical signal when it is vibrated.
The correct answer is:
piezoelectric effect
*Most US transducers use this type of crystal.
14. A 3-dimensional array acquires data from __________.
fetal development studies
multiple areas within the scan line virtually simultaneously
a voxel or volume element
a highly focused picture element (pixel)
Remember: 3D acquires data from all three planes. A “3 dimensional volume”
The correct answer is:
a voxel or volume element
15. A(n) ____________ transducer utilizes many piezoelectric crystals arranged along the face of the transducer that are activated in a sequence creating a 2-D image with a fixed focus?
curvilinear array
mechanical array
linear sequential
annular array
Remember: The crystals are activated sequentially down the array. This was the beginning of the array technology used to create a 2-D image but there are superior methods to activate the crystals today.
The correct answer is:
linear sequential
*The crystals are activated sequentially down the array.
16. A ___________transducer would most likely be optimal for a cardiac exam.
sector
linear
mechanical
curvilinear
Remember: The small footprint allows intracostal access
The correct answer is:
sector
17. Which type of transducer allows the sonographer to beam steer?
Curvilinear array
Sector array
Linear array
Annular array
The correct answer is:
Linear array
18. A(n) _____________ transducer uses numerous concentric piezoelectric crystals to create a 2-D image?
Question 18Answer
annular array
mechanical array
curvilinear array
sequential array
Remember: Annular transducers are considered an array transducer due to multiple crystals.
The correct answer is:
annular array
*This has some advantages in that the size and shape of the crystals allows focusing on all planes at all depths but not widely used today.

19. A transducer can be defined as ________________
a device that allows diagnostic images
a mechanism of electrical conversion
a device that converts one form of energy into another
a computer program that allows image formation from an electrical signal
Remember: There are many kinds of transducers
The correct answer is:
a device that converts one form of energy into another
20. What is the primary importance of the matching layer?
Prevents excessive “ringing” of the crystal
Isolates the patient from an electrical signal
Limits outside electrical interference
Allows improved transmission of the sound energy into the tissues
Remember: The acoustical impedance between the crystal and the tissue is very large and therefore most of the energy would be reflected allowing very little to pass into the body. A matching layer has an acoustical impedance between the crystal and the tissues allowing for improved transmission.
The correct answer is:
Allows improved transmission of the sound energy into the tissues
*A matching layer allows for improved transmission.
21. A linear array transducer uses a(n) ___________ focusing technique.
internal fixed
sequential
sonographer adjustable
electronic delay
Remember: Virtually all of todays transducers use electronic delays (phased) to focus the beam.
The correct answer is:
electronic delay
22. What pulse characteristics results in the more narrow bandwidth?
A higher amplitude center frequency
A long pulse length
The width or time of the electrical pulse
A short pulse
Remember: The bandwidth is the range of frequencies within the pulse.
The correct answer is:
A long pulse length
*The longer the pulse, the more narrow the bandwidth
23. What would be the optimal way to adequately evaluate a deep structure?
Decrease the frequency
Increase the duty factor
Change the PRF
Increase the frequency
Remember: The lower the frequency, the less attenuation allowing increased penetration of the US energy.
The correct answer is:
Decrease the frequency

24. The electronic delays depicted in this image will result in ________________.
beam steering and focusing
a trapezoidal beam shape
focus adjustment
beam steering
Remember: The slope results in beam steering, while the curvature allows focusing.
The correct answer is:
beam steering and focusing
*There is a slope as well as a curvature depicted in the electronic delays.

25. A malfunction in a linear array would result in a __________ ?
loss of information along a vertical line beneath the malfunction
loss of information horizontally through the image
loss of Doppler capabilities along that scan line
loss of image information
The correct answer is:
loss of information along a vertical line beneath the malfunction
*This would result in a loss of data from the malfunctioning scan line.

26. A sector transducer creates a ______________ field of view.
fan shaped
rectangular
variable
trapezoidal
Remember: A fan shape is also known as a sector shape or pie shape.
The correct answer is:
fan shaped
*Also referred to as triangular.


27. In the figure, the arrow points to the _____________________.
matching layer
electrical shielding
acoustical isolator
backing material
Remember: A short pulse is a good pulse!
The correct answer is:
backing material
*The backing layer stops the crystal from excessive vibrations helping to create a short pulse.

28. What best defines “Quality Factor”?
The degree of “dampening” of the pulse
The ultimate resolution of the sound beam
The range of frequencies both above and below the main frequency
A unitless number related to bandwidth
Remember: Involves the pulse duration (time required to transmit the pulse) and the pulse repetition period which includes the transmit AND receive (listening) time
The correct answer is:
A unitless number related to bandwidth
29. A phased array transducer activates the crystal elements ______________ .
in various sequences
in random order and then re-assembled upon reception
all at once
one at a time down the array
Remember: Phased refers to the time delays used to create a “synthesized” ultrasound wave.
The correct answer is:
in various sequences
*The timing of the electrical pulse applied to the crystals is varied to create various beam characteristics.
30. The ability of a phased array transducer to focus the beam is based on ______________.
Snell's Law
Law of Refractivity
Huygen’s Principle
Bernoulli's Principle
Remember: The same principle that allows beam steering allows a focusing of the beam at specific depths created by a converging of the wavefronts created by the timing delays.
The correct answer is:
Huygen’s Principle
*Yes. Huygen's principle forms the basis of our ability to focus the beam electronically

31. A(n) _________________ transducer will have the best axial resolution.
linear sequential array, 6 MHz
linear phased array, 7.5 MHz
convex array, 4 MHz
annular array, 6 MHz
Remember: The higher the frequency, the shorter the wavelength and the shorter the SPL.
The correct answer is:
linear phased array, 7.5 MHz
32. When using a 2-6 MHz curvilinear array transducer to image an abdominal structure at a depth of 16 cm, the sonographer should adjust the frequency to ____________.
6 MHz
2 Mhz
5 MHz
4 MHz
Remember: The higher the frequency, the greater the attentuation so a lower frequency will most likely improve the image of a deep structure.
The correct answer is:
2 Mhz
*While not absolute, the lower frequency is likely to be optimal.
33. A(n) _______________ transducer utilizes many piezoelectric crystals that are arranged in a rectangular pattern and uses beam steering to create a triangular shaped image?
annular array
sector array
sequential array
mechanical array
Remember: A primary advantage of the sector array is the small footprint aiding access is some applications such as intercostal imaging in cardiac evaluation.
The correct answer is:
sector array
34. Dynamic receive focus employes the use of ________________ during the listening or receive cycle.
all the crystal elements
a small precisely located element
only the elements used during transmission
multiple elements which vary based on the depth of the signal
Remember: In addition to using multiple crystal to generate the active scan line, multiple crystals can be used to recieve the reflected energy.
The correct answer is:
multiple elements which vary based on the depth of the signal