Automatic Exposure Control (AEC)
A system used to consistently control the amount of radiation reaching the image receptor by terminating the length of exposure.
How does AEC work?
radiation is transmitted through the patient and hits the IR and converted into an electrical signal that trips the system and terminates the exposure time
How many photocell/sensors/detectors are there for AEC?
usually a set of 3 radiation measuring detectors but could have as many as 5, arranged in some specific manner
Configuration for photocell/sensors/detectos for AEC
dependent on the anatomy of interest the radiographer selects it determining which ones of the usual 3 actually will meaure radiation exposure reaching the IR
Photocell/Sensors/Detectors
the part of the AEC that detects when the desired amount of radiation has been recieved to achieve the desired IR brightness
The position of detectors varies in two ways what are they?
Entrance Type & Exit Type
What is cotrolled by the AEC device?
Time
What is controlled by the techonologist when AEC is used?
kVp & mA, not time
Entrance Type
AEC device positioned between the patient and the IR
-most efficient
Most common type of position of the detectors?
Entrance Type
Exit Type
detectors are positioned after the patient and image receptor
How does AEC funtion?
exit radiation from the patient interact with the detector, producing a current, the current produced is the detector charges the x-ray timer and when the x-ray timer recieves a predetermined amount of charge from the detector (photcell), the x-ray exposure is terminated
Stregth of current & Exit Radiation interacting with the detector relationship
directly proportional
-small current = small amount of exit radiation (vice versa)
Exit Type will use what device to convert radiation into electricity?
Phototimers
-no longer used
Entrance Type will use what device to convert radiation into electricity?
Ionization Chamber Systems
When employing an AEC device, the radiographer will select what on the contro planel?
kVp, mA, detecto arrangement, exposure arrangement, backup timer