Automatic Exposure Control (AEC)

A system used to consistently control the amount of radiation reaching the image receptor by terminating the length of exposure.

2
New cards

How does AEC work?

radiation is transmitted through the patient and hits the IR and converted into an electrical signal that trips the system and terminates the exposure time

3
New cards

How many photocell/sensors/detectors are there for AEC?

usually a set of 3 radiation measuring detectors but could have as many as 5, arranged in some specific manner

4
New cards

Configuration for photocell/sensors/detectos for AEC

dependent on the anatomy of interest the radiographer selects it determining which ones of the usual 3 actually will meaure radiation exposure reaching the IR

5
New cards

Photocell/Sensors/Detectors

the part of the AEC that detects when the desired amount of radiation has been recieved to achieve the desired IR brightness

6
New cards

The position of detectors varies in two ways what are they?

Entrance Type & Exit Type

7
New cards

What is cotrolled by the AEC device?

Time

8
New cards

What is controlled by the techonologist when AEC is used?

kVp & mA, not time

9
New cards

Entrance Type

AEC device positioned between the patient and the IR

-most efficient

10
New cards

Most common type of position of the detectors?

Entrance Type

11
New cards

Exit Type

detectors are positioned after the patient and image receptor

12
New cards

How does AEC funtion?

exit radiation from the patient interact with the detector, producing a current, the current produced is the detector charges the x-ray timer and when the x-ray timer recieves a predetermined amount of charge from the detector (photcell), the x-ray exposure is terminated

13
New cards

Stregth of current & Exit Radiation interacting with the detector relationship

directly proportional

-small current = small amount of exit radiation (vice versa)

14
New cards

Exit Type will use what device to convert radiation into electricity?

Phototimers

-no longer used

15
New cards

Entrance Type will use what device to convert radiation into electricity?

Ionization Chamber Systems

16
New cards

When employing an AEC device, the radiographer will select what on the contro planel?

kVp, mA, detecto arrangement, exposure arrangement, backup timer