Eddy Clifford Jerman

“father of radiology”, American inventor and early expert in the techniques of radiology. one of the first people to focus on the details that created a quality xray like exposure and positioning

2
New cards

anatomically programed, fixed KVP, variable KVP, proportional anatomy approach, AEC

what are the 5 different types of technique charts?

3
New cards

anatomically programmed technique chart

technologist uses a control screen to select the anatomic part, projection, and relative size (S, M, L) of the part to be imaged. Unit uses microprocessor data to select KVP and mAs.

4
New cards

critical

it is ___ that all radiographers use a particular technique chart by the same method and in exactly the same manner

5
New cards

phantom images, optimal image, build chart, phantom testing using chart, clinical trials, clinical fine tuning, ongoing fine tuning

what are the steps to building a technique chart?

6
New cards

phantom images

series of test exposures made using a variety of technical factors above and below the average levels to determine the optimal quality image

7
New cards

fixed KVP system

optimum KVP is selected and remains constant for the given body part and thickness

8
New cards

optimum KVP

maximum KVP that achieve full penetration of all anatomy of interest resulting in the most desired level of contrast

9
New cards

uniform contrast, wide exposure latitude, lower mAs prolongs tube life

advantages to a fixed KVP system

10
New cards

lower contrast images and more scatter

disadvantages to a fixed KVP system

11
New cards

low KVP because low scatter

high contrast is caused by

12
New cards

high KVP b/c increased scatter

low contrast is caused by

13
New cards

optimal KVP

maximum KVP level that will produce images with appropriate contrast and ensure proper penetration

Produces lower contrast (wide latitude) and the minimum patient dose

14
New cards

5 cm rule (fixed KVP)

for every 5 cm of thickness in a body part, mAs needs to be halved or doubled

15
New cards

30%

On old school film what was the minimum change in mAs between steps?

16
New cards

Variable KVP charts

KVP is changed based on measurement of body part thickness and mAs is specified for each body part

17
New cards

allows for small changes; changes contrast

Advantages then disadvantages of variable KVP systems

18
New cards

15% rule (variable KVP)

changing KVP by 15% gives the appearance of doubling/halving mAs so if KVP increases mAs should be halved and if KVP decreases mAs should by doubledh

19
New cards

historical development example

2 KVP per CM (2kvp x CM) + minimum KVP

20
New cards

stepped variable KVP system

combination of adjusting kvp and using the 4-5 CM rule of doubling mAs

21
New cards

proportional anatomy approach

A proper radiographic technique can be derived from the known technique of another body part (ex. 1 foot is about 2 hands so double technique used for hands to get the technique used for foot)

22
New cards

AEC system

selection of exposure time factors are used thereby mAs selections are eliminated

professional judgment must still be used for proper mA kVp and distance chosen

23
New cards

AEC

help the radiographer in producing CONSISTENT images from patient to patient regardless of the size resulting in

decreased repeat rate

decreased patient exposure

increased department efficiency

24
New cards

Automatic exposure cntrol

tech sets kVp and mA initiating the exposure. Time is determined by the machine and stops when IR has sufficient density terminating the exposure

25
New cards

less then 5 mAs or on a free plate

when can AEC not be used

26
New cards

photo timers and flat ionization chambers

what are the two types of AEC devices?