A more detailed QC program for DR systems has been
developed by the
AAPM
acceptance testing is performed by a
medical physicist
daily testing is performed by a
technologist
weekly testing is performed by a
technologist
monthly testing is performed by a
technologist
semiannual/annual testing is performed by a
medical physicist
system adjustments are made by the
vendor or service personnel
dropped pixels
Areas of no data (usually appearing as black dots or lines) caused by damage to the active matrix array
electronic noise
Each detector element has a certain
amount of electronic noise associated with it (also known as
dark noise)
incomplete charge transfer
Caused by the collection element not sufficiently sending all data to the computer for processing, resulting in incomplete image formation
electronic memory artifact
Exposures are taken in too rapid a sequence, resulting in not enough time for each previous exposure to transfer the entire signal, creating an
artifact known as “ghosting.”
uneven amplification
In large detector arrays, several amplifiers are used that can vary in gain from one to the next. Flat-fielding or equalization is a software correction that is performed to equalize the response of each pixel to a
nonuniform X-ray beam.
backscatter
Scattering through the back of the cassette that can either create noise in the image or may create an image of the internal circuitry found in the back of the cassette
high capital cost
DR image receptors are very expensive; therefore, they must be rugged to handle the workload during radiographic examinations.
three levels of Quality Control and System Maintenance required for digital radiographic systems
routine, full inspection, system adjustment
there is no standard QC process that can
be used by all systems due to variety
since there is no standard QC process, QC techs must rely on the
manufacturer’s guidelines for specific QC procedures
There are some basic QC procedures that are
common to all manufacturer's systems
common QC procedure for portables
exposure switch cord should be a minimum of 6ft
basic QC common to all systems
DR AMA inspection and cleaning, display monitor inspection and cleaning, phantom image testing, wireless transmission inegrity
main components of a DR system include
image receptor, computer and software, workstation (display monitor)
image receptor
consists of an active-matrix array (AMA) or flat panel (FP)
computer or software
used for processing the image
workstation (display monitor)
used to display image and carry out post processing functions
The image receptors used must be kept clean and free of dirt, debris, blood, contrast media, etc. to
ensure they remain effective and provide accurate and quality images
The internal batteries on wireless DR cassettes should be inspected to ensure that they
are fully charged
wireless transmission integrity should be evaluated
daily (basically done without realizing)
An exposure of a phantom can be made on a wireless DR cassette and then viewed on the monitor to ensure that
the wireless signal was received uncorrupted
computer processor
Determines how fast the computer works (Should be at least an I7 or I9 (preferred) processor)
computer RAM
16-32 GB (preferred)
hard drive
solid state device (SSD), preferably swappable
software
program that allow the computer to process the image
software should contain productivity aids
positioning guides, image communication tools/web viewers/email/DICOM, technique/x-ray exposure setting assistance during image acquisition, auto report generation for radiologist
how often should you check display monitor peformance
at least monthly (recommended by ACR)
A high resolution monitor is used as the
primary display monitor for viewing of digital radiographic images
standard computer monitors are used for
secondary displays (clinical displays)
Cathode Ray Tube Displays (CRT)
They are no longer manufactured, but many are still in use today
Liquid Crystal Displays (LCD)
flat panel monitors that are the most used monitor for general computer
use as well as viewing diagnostic images
Plasma Screen
also a flat panel monitor, this displays a brighter image than the first two types but is more expensive
AAPM QC Acceptance Testing says to check
inventory, installation, function, clinical use
inventory check
did you get what you ordered (features, accessories)
installation check
is it correctly placed in the room
function check
does it work as advertised
clinical use check
are the images acceptable
Phantom images are obtained from
special test tools that can either be supplied by the digital system manufacturer, or a multipurpose test phantom that can evaluate any digital system can be purchased
Phantom image testing should be performed on
all digital systems
PACS are used to
store, archive, and retrieve digital images
PACS should be tested
quarterly
Two methods for performing digital fluoroscopy
image intensifier tube (digital fluoro), flat panel (digital fluoro)
digital fluoroscopy components
oX-ray Generator
oImage Intensifier – tube type or flat-panel
oCamera – only with tube-type II
oAnalog to Digital Converter
oComputer – creates image using a matrix
oWorkstation with high resolution monitors allows for post-processing
Nondigital components should be tested using
same
procedures as standard fluoroscopy
Digital imaging components are tested using
special test tools according to system manufacturer
All of the following types of QC tests can be done by technologist EXCEPT
a) Daily tests
b) Acceptance tests
c) Monthly tests
d) Weekly tests
Acceptance tests
All of the following are potential DR imaging problems EXCEPT
a) Electronic Noise
b) Back Scatter
c) Decreased Capital Costs
d) Electronic Memory Artifacts
Decreased Capital Costs
The primary monitor is the monitor that we view images on as technologist.
a) True
b) False
False
When performing Acceptance testing, the AAPM says to check which of the following?
a) Inventory
b) Installation
c) Clinical Use
d) All the above
All the above
As a technologist, you can perform which of the following daily QC tests?
a) Verify that the wireless DR systems are communicating without any interference
b) Phantom image testing and analysis
c) Image processing evaluation
d) Evaluate exposure indicator accuracy
Verify that the wireless DR systems are communicating without any interference