it employs the visual detection of fluorescent dyes coupled (conjugated) to antibodies which react with the Ag present using fluorescent microscopy

fluorescent Ab Assay

2
New cards

he developed the immunofluorescence techniques for the first time

Albert Coons

3
New cards

Albert Coons used _____, a blue fluorescing compound, coupled to pneumococcal antiserum to detect bacterial antigens in tissue sections

Beta-anthracene

4
New cards

a histochemical or cytochemical techniqur for the detection and localization of antigens

immunofluorescence

5
New cards

FITC absorbs light or short blue light at wavelength ____

490-495nm

6
New cards

FITC emits its characteristic green color at ___

517nm

7
New cards

a chemical that can absorb electromagnetic energy of relatively short wavelngth, and almost instantaneously emit light at a longer frequency (lower energy level)

fluor or fluorochrome

8
New cards

a chemical compound able to re-emit light upon excitation

fluorophore

9
New cards

a fluorochrome that emits green light

fluorescein isothiocyanate (FITC)

10
New cards

a fluorochrome that emits red color

tetramethylrhodamine isothiocyanate

11
New cards

what is used to view fluorescence?

fluorescent microscope

12
New cards

who invented fluorescent microscope?

August Kohler, Carl Reichert, and Henrich Lehmann

13
New cards

the brightness of fluorescence observed by the eye depends on:

  1. the efficiency at which the dye converts incident light (absorbed light) into fluorescent light

  2. the concentration of the dye in the tissue specimen

  3. the intensity of the exciting (absorbed) radiation

14
New cards

a part of the fluorescent miscroscope that produces an intense beam

mercury vapor arc lamp

15
New cards

a part of the fluorescent miscroscope that limits the heat transfer from the light source

heat filter/infrared filter

16
New cards

a part of the fluorescent miscroscope that transmits only the desired wavelngth

exciter filter

17
New cards

a part of the fluorescent miscroscope that is designed to efficiently reflect excitation wavelength

dichroic mirrors

18
New cards

a part of the fluorescent miscroscope that provides a dark background against which the fluorescent object glow

darkfield condenser

19
New cards

a part of the fluorescent miscroscope that removes any remaining ultraviolet light which could damage the viewer's eyes, or blue & violet light which could reduce the images

barrier filter (lens)

20
New cards

factors affecting fluorescence measurements

  1. light scattering in the soln

  2. quenching of the solvent

  3. cuvettes

  4. fluorescence intensity

21
New cards

exciting light can collide with non-fluorochrome molecules in the sample, and cause the light to be scattered. The scattered light is of the same wavelength as the exciting light, and so contribute to the bg noise/reading

light scattering in the soln

22
New cards

it refers to the decrease in fluoroscent intensity; may occur when the fluorochrome interacts with molecules in the soln. or some of the radiant energy absorbed by the fluorochrome is transferred to the molecules rather than being emitted as fluorescence

quenching of the solvent

23
New cards

are used to measure the fluorescence and can contribute to the bg noise bcs some materials in it can fluoresce

cuvettes

24
New cards

decreases with increasing temp

fluorescence intensity

25
New cards

general stages involved in immunofluorescence technique

  1. preparation of immune antiserum or purified gamma globulin

  2. conjugation with fluorescent dye

  3. staining procedure

26
New cards

what is the principle of immunofluorescence technique?

specific antibody is conjugated with fluorescent comounds. The conjugated antiserum is added to cells or tissues and becomes fixed to antigens, thereby forming a stable immune complex. Non-antibody proteins are removed by washing and when the resultant preparation is observed in fluorescence microscope against a dark bg, aAg bound specifically to fluorescent aAb can be detected by virtue of the bright color of the Ab (for shot, Fr. Google: Ag-Ab binding using fluorescent detection)

27
New cards

types of qualitative fluorescent Ab technique

  1. direct immunofluorescnece test/single layer technique

  2. indirect immunofluorescence test/double layer technique

  3. ihibition technique

  4. complement staining technique

  5. biotin-avidin system

28
New cards

it is used to ID unknown Ag by using fluorescein labeled Ab

direct immunofluorescence test/single layer technique

29
New cards

what is the principle of direct immunofluorescence test/single layer technique

conjugated antiserum added directly to the tissue section or viable cell suspension. Subsequent washing in physiologic buffer saline removes excess uncombined labeled Ab. Mixture us then examined under the fluorescent microscope

30
New cards

it is performed to ID foreign substances (Ag) in tissues

direct immunofluorescence test

31
New cards

it is used to detect Ab in unknown sera or unlabeled sera

indirect immunofluorescence test/double layer technique

32
New cards

what is the advantage of indirect immunofluorescence test?

it eliminates the need to purify and individually conjugate each serum sample

33
New cards

what is the principle of indirect immunofluorescence test?

unlabeled Ab is exposed to Ag, followed by the addition of labeled antiglobulin sera directed against the globulin of the species used in the initial exposure. In this case, fluorescence indicates that the reaction between the Ag, the uknown Ab, and the labelled antiglobulin has taken place

34
New cards

to detect a wide variety of human Ag-Ab reactions, _____ is used

fluoriscinated rabbit or goat antihuman globulin

35
New cards

steps in direct immunofluorescence

  1. fix specimen on slide

  2. add Ab specific for the desired Ag

  3. look for fluorescence

36
New cards

steps in indirect immunofluorescence

  1. fix specimen on slide

  2. add Ag specific for the desired Ag

  3. add second Ab

  4. look for fluorescence

37
New cards

what is the principle of inhibition technique?

nlabeled Ab is added to Ag. Ag becomes saturated w/ unlabeled specific Ab. Labeled Ab is then added. No flluorescence is seen and the test remains non-reactive because no Ag sites remain to react w/ the labeled Ab

38
New cards

it is similar to the indirect technique, except that the labeled/conjugated aAb is directed against the complement components, often to guinea pig origin

complement staining technique

39
New cards

advantege of complement staining technique

a single conjugate (anti-pig complement) may be used to test the serum form any species

40
New cards

its fluorescence indicates that a reaction has occured between the unknown Ab and the Ag, resulting in the fixation of complement to the complex,, and that a subsequent reaction has occurred between the guinea pic complement and the anti-guinea pig complement

complement staining technique

41
New cards

a basic glycoprotein derived from egg albumin with MW of 68,000; it shows a remarkably high affinity for vitamin Biotin

Avidin

42
New cards

it can be easily coupled covalently to a protein (Ab) and then reacted with fluorochrome-coupled Avidin

Biotin

43
New cards

what is the principle behind Biotin-Avidin System

after reaction of Ag w/ labeled Ab, the Biotin-labeled second Ab is added. Since many molecules of Biotin can be coupled to an Ab, the subsequent addition of fluorochrome-labeled Avidin results in a firm bond with exceedingly bright fluorescence.

44
New cards

what is the advantage of Biotin-Avidin System

the lack of non-specific binding of fluorochrome-coupled Avidin to various substrates and general use of Avidin cojugates in binding to Biotin-labeled Ab regardless of their derivation

45
New cards

it measures the fluorescence spectra of microscopic samples or microscopic area of a larger object

microfluorometer