What primarily causes Scatter radiation?

Compton effect and some coherent scatter

2
New cards

What is contrast?

degree of difference in OD between areas of an image

3
New cards

What would only be in an ideal image?

Transmittance and absorption

4
New cards

What does scatter do?

Reduce image contrast by exposing parts of film

5
New cards

What are the main factors that contribute to scatter radiation?

ALL of the above

6
New cards

What decreases both Compton and photoelectric effect?

Energy (kVp)

7
New cards

T/F: Photoelectric decreases much faster than compton when increasing energy(kVp)?

True

8
New cards

Is there more compton or photoelectric at higher energies?

Compton

9
New cards

Why would we ever raise the kVp?

to decrease patient dose and increase transmittance

10
New cards

What occurs to patient dose is kVp is to low?

Increases

11
New cards

What occurs to transmittance if kVp is to low?

Not enough x-rays will go through to create image

12
New cards

What happen to scatter radiation as the field size increases?

Increases

13
New cards

What is the most common beam-restricting device?

Light-localizing variable aperture collimator

14
New cards

What is the first part that a collimator controls?

reduce and control off-focus (stem) radiation exiting the X-ray tube window

15
New cards

What is the benefit of a adjustable collimator?

Variable x-ray beam size

16
New cards

What are the devices used to reduce scatter?

B and C

17
New cards

What do beam-restricting devices do?

limit the size and shape of the primary X-ray beam, thereby reducing patient dose and scatter radiation

18
New cards

What do radiographic grids effect?

Remnant beam (after patient)

19
New cards

How much scatter can a high-quality grid reduce?

80-90%

20
New cards

What is the inter-space material in a grid made of?

Aluminum or plastic

21
New cards

What are the important dimensions of a grid?

ALL of the above

22
New cards

What determines the grid ratio?

Height of grid / interspace width

23
New cards

Which are the more effective grids?

High ratio

24
New cards

What is the downside to a higher ratio grid?

Increased patient radiation dose

25
New cards

What ratio do most systems use?

8:1 to 12:1

26
New cards

What is the range of grid rations?

5:1 to 16:1

27
New cards

What is grid frequency?

Grid strips per centimeter

28
New cards

What range of frequencies do most grids have?

25-45 lines per centimeter

29
New cards

What are the results of using high-frequency grids?

Less distinct grid lines on radiograph

Requires high exposure, higher patient dose

30
New cards

What is the principal function of a grid?

Improve image contrast by absorbing scatter

31
New cards

What is the contrast improvement factor? (k)

image contrast with grid / image contrast without grid

32
New cards

What value in contrast improvement factor indicates no improvement?

1

33
New cards

How does a higher grid ratio and grid frequency affect contrast improvement factor?

It increases the contrast improvement factor

34
New cards

What must be increased to maintain OD when using a grid?

mAs

35
New cards

What is the Bucky factor? (B)

The increase in exposure (mAs) necessary to compensate for radiation absorption when a grid is introduced

36
New cards

What is another name for the Bucky factor?

Grid factor

37
New cards

What is the equation for the Bucky factor?

Patient dose with grid / patient dose without grid

38
New cards

What is the simplest type of grid?

Parallel

39
New cards

What is a result of decreased SID?

there is more potential for grid cutoff

40
New cards

What can also influence grid cutoff?

X-ray field size

41
New cards

What occurs if there is short SID?

OD decreases towards the edge of image receptor

42
New cards

What is the goal of focused grids?

Minimize grid cutoff

43
New cards

How are strips aligned in a focused grid?

angled to match the divergent angle of the X-ray beam corresponding to a specific focal distance

44
New cards

What is the purpose of moving grids?

Eliminate grid lines

45
New cards

What kinds of grids will have less noticeable lines?

High grid frequency and moving grid mechanisms

46
New cards

What is the name for the moving grid?

Potter-Bucky Diaphragm

47
New cards

What is the largest problem with grids?

Misalignment (leading to grid cutoff)

48
New cards

What can cause misalignment of a grid?

ALL of the above

49
New cards

What is the air-gap technique?

alternate method of reducing scatter by increasing object distance

50
New cards

What grid is the air gap technique equivalent to?

8:1

51
New cards

What is the main disadvantage to air-gap technique?

Image magnification

52
New cards

What still needs to be increased with air-gap technique?

mAs

53
New cards

At what rate does mAs need to be increased with air-gap technique?

10% for every cm

54
New cards

When was computed radiography introduced?

1980s by Fujifilm Medical

55
New cards

What other names did the computed radiograph have?

“Photostimulable Phosphor Plate Radiography”

PSP plate radiography

56
New cards

What is photostimulable luminescence?

Receiving x-rays and storing the energy as trapped electrons, then releasing them as light when exposed to a different light source

57
New cards

How is Photostimulable luminescence accomplished?

Photostimulable phosphors (PSPs) in the plate

58
New cards

What are storage phosphor screens?

a latent image is stored for a period

59
New cards

What needs to occur once an image plate is exposed and the latent image is formed?

Stimulated to obtain the manifest image

60
New cards

What causes the PSP to emit light?

Finely focused laser beam scanning the plate

61
New cards

In what light region does the PSP emit light?

Blue

62
New cards

What is used to clear the image plate to prepare for a new image?

Intense white light

63
New cards

What contains the laser photomultiplier tubes and photodiodes?

Computed radiography reader

64
New cards

What does the Computed radiography reader do?

Removes PSP plate, scan it, erase it, and replace it

65
New cards

When was digital radiography introduced?

1980s by Fujifilm Medical after CR

66
New cards

What does digital radiography not have compared to computed radiography?

ALL of the above

67
New cards

What does digital radiography use?

Flat panel detectors (FPDs)

68
New cards

What is direct digital radiography?

Converts x-ray photons into electrical signal

69
New cards

What is indirect digital radiography?

Converts x-ray photons into visible light, then into an electrical charge

70
New cards

What uses indirect conversion of x-ray photons into visible light and to turn into electrical charge besides indirect flat planels?

Charged Coupled Device Detectors

71
New cards

What do direct flat panel detectors use?

Amorphous selenium photoconductor to convert x-rays into electrical signals

72
New cards

What processes the electrical signals in direct flat panel detectors?

Thin film transistor (TFT)

73
New cards

What kind of flat panel detectors are most commonly used in mammography where high resolution is required?

Direct

74
New cards

What does indirect flat panel detectors use?

Phosphor layer for producing visible light, which is read by a photodiode

75
New cards

What is a downside to CCD detectors?

Potential to produce high amount of image noise due to demagnification process

76
New cards

What is bit depth?

the number of discrete sampling levels used to interpret and convert analog signal into a digital one

77
New cards

What does a high bit depth allow for?

More accurate representation of the incoming signal; Requires more memory

78
New cards

What represents the number of bit depth levels?

2^n

79
New cards

Why is a small bit ranges beneficial?

Accuracy

80
New cards

What is another way to make bits smaller?

Shrink total range of energy values

Reduces dynamic range

81
New cards

What is the dynamic range of detector?

Range of exposure values over which the signal will be properly digitized

82
New cards

What will result from a low dynamic range?

Limit range of exposure values able to be represented

83
New cards

What will result from a high dynamic range?

Each bit covering a greater exposure value range

84
New cards

What is a DICOM image file

Standardized image file type for medical imaging

85
New cards

What is the filename extension for a DICOM file?

.dcm