1/25
Looks like no tags are added yet.
Q: What is mass spectrometry?
it is a powerful instrumental method of analysis.
It can be used to:
- find the abundance and mass of each isotope in an element allowing us to determine its relative atomic mass
- find the relative molecular mass of substances made of molecules
Q: What are the steps in Mass spectrometry?
Vaporisation, Ionisation, Acceleration, Deflection (Ion Drift), Detection
Q: What are the two main techniques of ionisation?
Electrospray ionisation.
Electron impact ionisation
Q: What are the 3 key points about electrospray ionisation?
- a gentler technique and prevents fragmentation.
- Adds a H⁺, so Mr +1
-it's typically used for polymers and biological materials like DNA.
Q: What is electron impact ionisation used for?
is used for elements and substances with low formula mass (that can be inorganic or organic molecules).
When molecules are ionised in this way, the 1+ ion formed is known as a molecular ion.
Q: How does electrospray ionisation happen? (4 marks)
- the sample X is dissolved in a volatile solvent (eg water or methanol)
- it is injected through a fine hypodermic needle to give a fine mist (aerosol).
- the tip of the needle is attached to the positive terminal of a high-voltage power supply.
- the particles are ionised by gaining a proton (ie an H⁺ ion which is simply one proton) from the solvent as they leave the needle producing XH⁺ ions (ions with a single positive charge and a mass of Mr + 1).
Q: What's an equation for electrospray ionisation if the sample is X?
X + H^+ —> XH^+

Q: How does electron impact ionisation work? (4 marks)
- the sample is first vaporised and then hit with electrons from an electron gun.
- The electrons knock off electrons from the molecule
- The molecules are now positively charged ions.
- This method often causes the sample to fragment.
Q: What is the equation that shows electron impact ionisation?
X + e^- —> X^+ + 2e^-

Q: What happens during Acceleration in the mass spectrometer? (2 marks)
- Molecules are accelerated to all have the same kinetic energy,so the speed is dependent on the mass of the molecule.
- Lighter particles move faster and are detected before heavier particles
Q: What happens during the ion drift stage of the Mass Spectrometer? (2 marks)
- the positive ions travel through a hole in the negatively charged plate into a tube.
- the time of flight of each particle through this flight tube depends on its velocity which in turn depends on its mass.
Q: During the ion drift stage what equation can be used to determine the time of flight?
Time of flight = Distance / Velocity
t = time of flight (s)
d = length of flight tube (m)
𝑣 = velocity of the particle (m s⁻¹)
Q: Which ions of the three isotopes of magnesium 24Mg+ , 25Mg+ , 26Mg+ would reach the detector first?
24Mg⁺
Q: What is the Detection stage of Mass Spectrometry? (3 marks)
- the ions hit a negatively charged electric plate.
- when they hit the detector plate, the positive ions are discharged by gaining electrons from the plate. This generates a movement of electrons and hence an electric current that is measured.
- the size of the current gives a measure of the number of ions hitting the plate.
Q: A computer uses the data to produce a mass spectrum. What does this show? (3 marks)
- the mass to charge (m/z) ratio
- and abundance of each ion that reaches the detector.
- given that all ions produced by electrospray ionisation and most of the ions by electron ionisation have a 1+ charge, the m/z is effectively the mass of each ion.
Q: On the spectrum what is the x-axis, and what is the y-axis?
The x-axis is mass/charge ratio.
The y-axis is % abundance.
Q: What does the main peak on the spectrum indicate?
- Molecular ion peak.
- This is the peak of the greatest mass/charge ratio.
- This represents the mass/charge value of the molecule we are analyzing.
Q: What are the smaller peaks that cluster around the molecular ion peak?
- These are from the same molecules but with different isotopes in them.
- The isotopic molecules have different masses and so different mass/charge ratio values.
Q: Why do smaller and significantly lighter peaks in the spectrum occur?
- Because of fragmentation.
- The molecule can fragment in the spectrometer
- Does not usually happen during electrospray ionisation.
Q: Once we have the mass spectrum, what can we use it to calculate?
-The relative atomic mass.
- Relative atomic mass is the average weighted mass of an atom relative to carbon-12.
- The key word to look at here is "average".
- This is because its value is calculated taking into account all of its isotopes and their relative abundances
Q: How can the relative atomic mass (Ar) be calculated?
combined mass of all isotopes
combined abundance of all isotopes
Q: If an isotope's abundance is given as a percentage, how does this make the calculation slightly easier?
So long as we have considered all the isotopes, the sum of all the percentages must add to 100%.
Q: For molecules that are ionised by electron impact how can the molecular ion be identified?
the signal with the greatest m/z value is from the molecular ion and its m/z value gives the relative molecular mass.
Q: What other small peaks may be present around the molecular ion, if electron impact ionisation has been used?
due to molecular ions that contain different isotopes.
Q: When using electron impact ionisation (but not with electrospray ionisation), why may there also be peaks at lower m/z?
due to fragments caused by the break up of molecular ion.
Electron impact
Q: What ionisation has been used to produce this spectrum?