Types of xray equipment
Diagnostic
Therapeutic
Diagnostic
Diagnostic ranges
10-1200 milliamperes (mA)
0.001-10 seconds
25-150 kilovoltage peak (kVp)
Therapeutic
Radiation oncology
Capable of X-ray and electron therapy
meV energies
Tables
Uniform, radiolucent surface
Flat surface most common
4 way table top motions
Variable heights
Fixed and tilting models
Tables (must be)
Easy to clean
Free of crevices
Difficult to scratch
Tube supports
Numerous configurations
Floor-to-ceiling
Floor mounted
Mobile
C-arm
Upright receptor units (called)
Wall stand or upright bucky
Upright receptor Units (functions)
Supports image receptor in a vertical orientation
Vertical travel (standard)
Tiliting or non-tilting
Numerous uses
Upright receptor Units (Uses)
Chest
Upright abdomen
AC joints
Cervical spine standing orthopedics
Upright sinuses
Specialized Diagnostic Equipment
Mammography
Tomography
Dental/panoramic
Computed tomography (CT)
Radiation therayu simulators
Urologic
Incoming line current
60Hz, AC
Most common way for electricity to be supplied to buildings
Types of generators
Single-phase power generator
3-phase power (or multiphase) generator
High-frequency generators
Capacitor discharge units
Battery operated units
Single phase power generator
Inefficient for xray production
Primarily for residential properties
3 phase power (or multiphase) generator
Uses 3 times as much power compared to single-phase
Common in commerical and industrial facilities
High frequency generators
Most popular and efficient for stationary xray rooms
Allows incoming line voltage to convert from 60Hz to thousads of Hz
Capacitor discharge units
Small parts mobile unit
Battery operated that be be recharged from incoming line current periodically when parked
Takes longer to discharge (possible leakage radiation
Battery-operated units
No leakage possible
More stable and effiicient
Transformer (does what)
Transfers electrical energy from one AC circuit to another circuit (s), either stepping up (increasing) the voltage or stepping down (decreasing) the voltage.
Transforms the intensity of AC voltage and current
Transformers (ohms law)
Work on the principle of mutual induction and ohms law (current supplied to primary coil, then induced voltage in secondary coil)
Transformers (Types)
Autotransformer
High coltage step up transformer
Filament step down transformer
Autotransformer
(Primary to low voltage side)
Controlled by kVp setting
Supplied power to other parts of circuit
High voltage step up transformer
(Secondary or high tension side)
Raises incoming line voltage to kVp
Sends signal to xray tube for exposure
Filament step down transformer
(filament circuit)
Controlled by the mAs setting
Thermionic emission
Exposure Switch
Deadman switch
Located behind the control window (console) on the generator console on the low-voltage side of the main circuit.
Depress in one motion to max tube life
X-ray exposure initiated automatically after rotor is at correct rpm
Separate rotor switches also used (less common with modern equipment)
Mobile equipment required 6ft cord
Timer Circuit
Regulates exposure termination
Electronic timers (most sophisticated)
Capable of very short exposure times (0.001 sec)
Milliampere-second (mAs) timers
Terminates exposure based on a predetermined mAs value
AEC timer (automatic exposure control