Primary purpose of High PRF (HPRF) and Multi-Gate Doppler
Both techniques are used to measure higher Doppler shifts (faster blood velocities) with less or reduced aliasing.
High PRF (HPRF) Doppler
Fires rapid pulses using the timing of a shallower gate while focusing at a deeper depth to reduce aliasing; sums returning echoes from multiple gates into a single spectrum (creating range ambiguity / ambiguous gates).
Multi-Gate Doppler
Places multiple sample gates at equally spaced depths across a vessel and samples/displays each gate separately to measure higher velocities with less aliasing.
Fast Fourier Transformation (FFT)
The mathematical algorithm used by spectral Doppler processors to break complex returning echo signals into individual frequency shifts displayed over time.
Y-axis (vertical) on a spectral Doppler display
Velocity (cm/s) or Doppler frequency shift (kHz).
X-axis (horizontal) on a spectral Doppler display
Time.
Above vs. below the baseline on a spectral Doppler display
Direction of blood flow (above = flow toward the transducer; below = flow away from the transducer).
Brightness (spectral intensity) on a spectral Doppler display
Amplitude / strength of the returning RBC signal (representing how many red blood cells are moving at that specific velocity).
Frequency vs. amplitude on a spectral Doppler graph
Frequency (velocity) is shown by vertical height on the Y-axis (how fast blood moves), whereas amplitude is shown by pixel brightness (how many RBCs are moving at that speed).
Plug ("blunt") flow
A type of laminar flow in which all blood layers and RBCs move at the exact same velocity across the entire vessel; occurs at vessel entrances and in the ascending aorta.
Parabolic flow
A type of laminar flow with a bullet-shaped velocity profile where blood flows fastest in the exact center of the lumen and slowest near the vessel walls due to wall friction.
Turbulent / chaotic flow and eddy currents
Disorganized, swirling blood flow patterns (eddy currents) moving in many directions at varying speeds; occurs post-stenosis (distal to a narrowing) and in normal dilated areas like the carotid bulb.
Location of peak velocity vs. turbulence in a stenosis
Velocities peak inside (at the site of) the stenosis, while turbulent flow occurs distal (post) to the stenosis, producing clinical signs like a palpable thrill, audible bruit, or murmur.
Pressure gradient across a stenosis
Pressure is higher proximal to the stenosis and lower distal to it, creating a pressure gradient (ΔP).
Clear spectral window
The clean, empty black area beneath the systolic peak on a spectral Doppler tracing indicating smooth laminar flow, seen when a small sample volume is centered in the vessel.
Two causes of spectral broadening (absence of spectral window)
Modal velocity
The brightest portion of a spectral Doppler tracing, representing the velocity at which the greatest number of red blood cells are moving ("most encountered velocity").
Maximal velocity (Vmax)
The highest velocity tip of the spectral waveform (peak systolic velocity / fastest moving red blood cells).
Mean velocity (Vmean)
The mathematical average of all velocities within the traced spectral waveform contour; always less than maximal velocity (Vmax).
Velocity Time Integral (VTI)
A distance measurement (in cm) obtained by tracing a single Doppler spectral curve, indicating how far a column of blood travels during one heartbeat/flow period; used in echocardiography to calculate stroke volume and valve area.