tissue harmonics
generation of higher frequency waves as a sound pulse propagates through tissue in a nonlinear fashion
harmonics penetration vs. resolution
allows for a lower fundamental frequency (deeper penetration) and a higher harmonic frequency (better return signal resolution)
harmonic frequency sound waves
requires broadband transducers
secondary to tissue responses to the fundamental frequency
non-linear propagation of sound
reduces artifacts
two types of harmonic imaging mechanisms
tissue and contrast
tissue harmonics
not present as it leaves the transducer
created during transmission
nonlinear
weaker harmonic signal than UEA signal
contrast harmonics
needs UEAs (microbubbles)
created during reflection of bubbles
occurs only when enhancing agents are present and MI is >0.1
MI <0.3 recommended
nonlinear bubbles
stronger harmonic signal than tissue
fundamental frequency
frequency of the transducer
2nd harmonic
2x the fundamental frequency
ultrasound machines only process the ______ harmonic
second
the stronger the beam…
the more harmonics
midfield
best resolution and strength to produce harmonics
if the fundamental frequency is 1.8 MHz what is the harmonic ffrequency?
3.6 MHz (second harmonic, 2x)
the harmonic beam is relatively (narrow/wide) compared to the fundamental beam
narrow
one of the greatest advantages to harmonic imaging is…
clutter reduction
grating lobe reduction
lower energy grating lobes, improve lateral resolution and reduces lateral artifacts
axial resolution degradation
overlap of signals
longer PD
reduces bandwidth
pulse inversion harmonic imaging
multiple lines with different phases are transmitted
the lines of fundamental frequency add _________ and harmonic frequencies add ____________
destructively, contructively
advantages of tissue harmonic imaging
improves SNR
improves contrast and lateral resolution
reduces artifacts
why use UEAs
reflection from blood is very weak due to Rayleigh scattering
microbubbles
create an impedance mismatch
UEAs increase reflectivity from the blood by…
about 30dB
reflection(%) =
(impedance 2 - impedance 1/ impedance 2 + impedance 1)² x 100
agitated saline
assess septal defects (ASD, PFO)
assess TR
definity and optison (perfluorocarbon gas)
opacify heart chamber
wall motion abnormalities
filling defects
thrombus/ intracardiac masses
lumason (sulfur hexafluoride)
liver masses
agitated saline formula
0.5 mL air/ 8mL bacteriostatic saline
UEA requirements
safe
can’t reproduce
strong reflector
can pass through capillaries (small)
MI of < _____ is recommended to prevent early destruction of the microbubbles
0.3
MI >1
increased risk of bursting microbubbles, reducing effectiveness
(lower/higher) frequency transducers increase the risk of cavitation and destruction of microbubbles
lower