tissue harmonics

generation of higher frequency waves as a sound pulse propagates through tissue in a nonlinear fashion

2
New cards

harmonics penetration vs. resolution

allows for a lower fundamental frequency (deeper penetration) and a higher harmonic frequency (better return signal resolution)

3
New cards

harmonic frequency sound waves

requires broadband transducers

secondary to tissue responses to the fundamental frequency

non-linear propagation of sound

reduces artifacts

4
New cards

two types of harmonic imaging mechanisms

tissue and contrast

5
New cards

tissue harmonics

not present as it leaves the transducer

created during transmission

nonlinear

weaker harmonic signal than UEA signal

6
New cards

contrast harmonics

needs UEAs (microbubbles)

created during reflection of bubbles

occurs only when enhancing agents are present and MI is >0.1

  • MI <0.3 recommended

nonlinear bubbles

stronger harmonic signal than tissue

7
New cards

fundamental frequency

frequency of the transducer

8
New cards

2nd harmonic

2x the fundamental frequency

9
New cards

ultrasound machines only process the ______ harmonic

second

10
New cards

the stronger the beam…

the more harmonics

11
New cards

midfield

best resolution and strength to produce harmonics

12
New cards

if the fundamental frequency is 1.8 MHz what is the harmonic ffrequency?

3.6 MHz (second harmonic, 2x)

13
New cards

the harmonic beam is relatively (narrow/wide) compared to the fundamental beam

narrow

14
New cards

one of the greatest advantages to harmonic imaging is…

clutter reduction

15
New cards

grating lobe reduction

lower energy grating lobes, improve lateral resolution and reduces lateral artifacts

16
New cards

axial resolution degradation

overlap of signals

longer PD

reduces bandwidth

17
New cards

pulse inversion harmonic imaging

multiple lines with different phases are transmitted

18
New cards

the lines of fundamental frequency add _________ and harmonic frequencies add ____________

destructively, contructively

19
New cards

advantages of tissue harmonic imaging

improves SNR

improves contrast and lateral resolution

reduces artifacts

20
New cards

why use UEAs

reflection from blood is very weak due to Rayleigh scattering

21
New cards

microbubbles

create an impedance mismatch

22
New cards

UEAs increase reflectivity from the blood by…

about 30dB

23
New cards

reflection(%) =

(impedance 2 - impedance 1/ impedance 2 + impedance 1)² x 100

24
New cards

agitated saline

assess septal defects (ASD, PFO)

assess TR

25
New cards

definity and optison (perfluorocarbon gas)

opacify heart chamber

wall motion abnormalities

filling defects

thrombus/ intracardiac masses

26
New cards

lumason (sulfur hexafluoride)

liver masses

27
New cards

agitated saline formula

0.5 mL air/ 8mL bacteriostatic saline

28
New cards

UEA requirements

safe

can’t reproduce

strong reflector

can pass through capillaries (small)

29
New cards

MI of < _____ is recommended to prevent early destruction of the microbubbles

0.3

30
New cards

MI >1

increased risk of bursting microbubbles, reducing effectiveness

31
New cards

(lower/higher) frequency transducers increase the risk of cavitation and destruction of microbubbles

lower