When and who developed the first commercial fluoroscope
1896 by Thomas Edison
The first commercial fluoroscope allowed for _________ radiographic examination
Dynamic
with the first fluoroscope, radiation dose was _____ and the image was _____
High; dim
When was the image intensifier introduced
1950s
What did the image intensifier do to the image
Brightened and made indirect viewing possible
What is the difference between the mA used in image intensified fluoroscopy and mA used in radiographic mode
Image intensified uses low mA
Radiographic mode uses higher mA
What mA is used in image intensified fluoroscopy
0.5-5 mA
What mA is used in radiographic mode
50-1200 mA
Low mA allows for
increased fluoro time
Control panel includes a timer that buzzes audibly when ______ of xray fluoroscopic time has been used
5 min
The continuous beam is active using a
Deadman's switch
where is tube located in fluoro
under the table
An electronic vacuum tube that converts the remnant beam to light, then electrons, then back to light, increasing the light intensity in the process.
image intensifier
Basic parts of the image intensifier
input phosphor
photocathode
Focusing lenses
accelerating anode
output phosphor
The input phosphor faces the patient, absorbs the remnant beam, and emits
Light
Emitted light exposes the photocathode, which emits
Electrons in proportion to light intensity
______ are needed to result in one emitted electron.
Many light photons
order of image intensifier
Input phosphor to light
Light to electrons
Back to light
Increases light intensity
Emitted electrons are accelerated to the output phosphor by the _______ and "focused" on the output phosphor by the _______
Accelerating anode; electrostatic focusing lenses
What is the result from image intensifiers
Increased image brightness
an expression of the ability of an image intensifier tube to increase the brightness of the image.
brightness gain
Traditionally, brightness gain was found by
multiplying the flux gain by the minification gain
It is now common practice to express the increase in brightness with
conversion factor
An expression of the luminance at the output phosphor divided by the input exposure rate.
conversion factor
What is the efficiency of the image intensifier with a higher conversion factor?
Greater
What happens to the image intensifier's ability to increase brightness?
Deteriorates with tube age
Brightness gain formula
brightness gain = minification gain x flux gain
flux gain formula
Flux Gain = # of Output Light Photons/# of Input X-ray Photons
A function of the fluoroscopic unit that maintains the overall appearance of the intensified image (contrast and brightness).
Automatic brightness control (ABC)
ABC automatically adjusts
kVp, mA, or both
T or F
With automatic brightness control, the operator can select a desired brightness level
True
What increases voltage to the electrostatic focusing lens?
Magnification mode
With image intensifier, magnification mode is going to do what to patient dose
Increase
Magnification factor formula for fluoro
Full size input diameter/selected input diameter
What does magnification do to fluoroscopist's ability to see small structures and patient dose
Improves ability to see small structures
Increases patient dose
What does magnification mode do to spatial resolution
improves
Spatial resolution in a typical fluoro system is
4-6 lp/mm
misrepresentation of the true size or shape of an object
distortion
Unequal magnification of an image creates a
pincushion appearance
Pincushion appearance causes
Loss of brightness around periphery
What results when insufficient information is present to create the image
image noise
Solution for insufficient information present to create image
Increase mA
The viewing system now
Television monitor
series of optical lenses that focus the image from the output phosphor on the camera tube
Optical lens system
What diverts some of the light to the desired spot-filming device
Beam-splitting mirror
What converts the electronic signal into a visible image
Television monitor
The television monitor converts electronic signal to
Visible image
Spatial resolution of image intensifier
5 lp/mm
Spatial resolution of monitor
1-2 lp/mm
A static imaging process using a radiographic cassette to obtain an image
cassette spot filming
Cassette spot filming is done using a
Film camera
A static imaging system that is used with an optical lens system incorporating a beam-splitting mirror
Photospot camera
What did early versions of digital fluoroscopy add between camera tube and monitor
ADC and computer
What replaced the camera tube and improved digital fluoroscopy
Charge-coupled device (CCD) with ADC
A device that takes the video (analog) signal & divides it into a number of bits (1s &0s) that the computer "understands"
analog-to-digital converter (ADC)
The number of bits that the signal is divided into determines
Contrast resolution
What is necessary for the computer to process and display the image
ADC
What may be used in the place of an image intensifier
flat panel detector
Indirect capture detector
cesium iodide and amorphous silicon
Direct capture detector
amorphous selenium
T or F
FPDs are much lighter and more compact, and they produce a digital signal directly
True
What replaced image intensifiers
FPD
Current dynamic versions of FPD are capable of up to?
60 frames per second
Advantages of using flat panel detectors in place of an image intensifier
- They reduce tower size, bulk, and weight.
- They replace spot filming and other devices.
- They do not degrade with age, are more durable, have better contrast resolution, have higher detective quantum efficiency (DQE), have a wider dynamic range, present a rectangular field providing more information, and have the postprocessing options common to digital images.
- They do not exhibit most image artifacts seen with image intensifiers
FPD reduce the patient's radiation dose by as much as
50%
FPD fluoroscopy operate in radiographic mode of what mA
50-1200 mA
T or F
FPD operate in radiographic mode and additional radiographic images are not needed
True
does FPD have a higher or lower spatial resolution than image intensified systems
Higher
What must be charted
Fluoro time
Cumulative air kerma
Dose Air Product
What allows for virtual collimation without exposing the patient
Last image hold (LIH)
Frame averaging reduces patient dose and image noise by
averaging multiple frames together
Frame averaging does what to spatial resolution
Decreases
automatically adjusts the tube current, voltage, filtration, and pulse with to maintain exposure to the flat panel detector.
Automatic exposure rate control (AERC)
the selection of a smaller field of view (FOV).
electronic magnification
a design of the unit that rapidly turns the x-ray beam on and off during operation.
pulsed fluoroscopy
What does pulsed fluoroscopy do to patient dose
Decreases
In digital systems, what does increases magnification do to patient dose
No increase
how many pulses occur per second of operation; how many exposures occur per second.
Pulse rate
the length of each pulse.
pulse width
What will save single images or a fluoro sequence loop to memory. Then can be post processed and maintained in the patient's permanent record
Last image hold (LIH) and fluoro loop save
What specifies the intensity of x-rays at a given point in air at a known distance from the focal spot or source of x-rays and the cumulative or total air kerma will be monitored during the fluoroscopic procedure.
Air kerma
Air Kerma stands for
kinetic energy released in matter
Other dose displays that provide indicators of patient radiation risk and should be documented in the patient's record.
DAP and KAP
Liquid crystal layer sandwiched between polarizing layers
Liquid Crystal Display (LCD)
Contains a layer of gas filled pixels, and when the gas in ionized, they release UV radiation, simulating phosphor coating in the cell, producing visible light
Plasma monitors
In plasma monitors, when gas in pixels is ionized, they release
UV radiation
Produces equipment to monitor equipment performance and minimize patient dose.
QC programs
Quality control is a team effort among the
Radiographer, radiologist, and medical physicists
T or F
The radiographer may be responsible for the operational inspection of the equipment.
True
Who should conduct and interpret the performance inspection and testing of the equipment as well as oversee the entire quality control monitoring program.
Medical physicist