Meaghan Piretti Registry Prep- Equipment

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/50

encourage image

There's no tags or description

Looks like no tags are added yet.

51 Terms

1
New cards

The incoming power supply of the x-ray circuit is?

A/C

2
New cards

How many volts of power is coming in the x-ray circuit?

220 volts

3
New cards

The fuses in the x-ray circuit are also known as?

circuit breaker

4
New cards

The job is to protect equipment from excessive current?

fuses (AKA circuit breaker)

5
New cards

The fuses (AKA circuit breaker) is located?

on the primary side

6
New cards

what measures the incoming line voltage?

line voltage meter (AKA compensator)

7
New cards

What detects a surge/drop?

line voltage meter (AKA compensator)

8
New cards

What maintains the incoming 220 volts to the autotransformer?

line voltage meter (AKA compensator)

9
New cards

How many coils does the autotransformer have?

1

*TIP= Auto equals self

10
New cards

What does the autotransformer operate on?

self induction

11
New cards

What type of power supply do all transformers work off of?

A/C

12
New cards

The role/job is kVp selection not production

autotransformer

13
New cards

The autotransformer is located?

on the low voltage/ primary circuit

14
New cards

What determines the line voltage sent to the step up transformer?

Autotransformer

15
New cards

On the x-ray circuit, where do you select your mA?

mA selector/ Rheostat

16
New cards

What is also referred to as a variable resistor?

mA selector/ Rheostat

17
New cards

What regulates amperage- regulates thermionic emission?

mA selector/ Rheostat

18
New cards

Where is the mA selector/ Rheostat located?

Low voltage-filament circuit

19
New cards

The step up transformer works off what type of induction?

mutual induction

20
New cards

How many wires does the step up transformer have?

2

21
New cards

On the step up transformer, what side has more turns?

secondary side

* think your stepping up more turns

22
New cards

The step up transformer increases volts to?

kilovolts

23
New cards

Do we need high or low volts to produce xrays?

high

24
New cards

The step down transformer has more turns on what side?

primary side

25
New cards

What 2 jobs does the step down transformer have?

Decreases voltage & Increases amperage

*TIP= That is why it is next to mA selector. The "A" in mA, and the "A" in amperage

26
New cards

The step down transformer works off what type of induction?

mutual induction

27
New cards

The step down transformer has how many wires?

2

28
New cards

Something that we have to hit before we get to our x-ray tube because the xray tube only works on D/C?

Rectifier

29
New cards

What changes A/C to pulsed D/C?

Rectifier

*TIP= it rectifies the problem

30
New cards

How many solid state semiconductors diodes does the rectifier have?

4

31
New cards

Is more voltage ripple good or bad?

bad

32
New cards

Single phase has what percentage ripple?

100%

33
New cards

3 phase, 6 pulse has what percentage ripple?

14%

34
New cards

3 phase, 12 pulse has what percentage ripple?

4%

35
New cards

High frequency has what percentage ripple?

-1%

36
New cards

Total heat produced during an x-ray expsure?

heat units

37
New cards

Tube housing is made to absorb certain levels of?

Heat units

38
New cards

What is the heat unit classification for single phase?

kVp X mA X s X 1

39
New cards

What is the heat unit classification for 3 phase, 6 pulse?

kVp X mA X s X 1.35

40
New cards

What is the heat unit classification for 3 phase, 12 pulse?

kVp X mA X s X 1.41

41
New cards

What is the heat unit classification for high frequency?

kVp X mA X s X 1.45

42
New cards

What provides extremely short exposure times by taking advantage of tube loading potential?

falling load generator

ex.) Lat T-spine w/ breathing technique we need something with short exposure times

43
New cards

What are 4 thing that we need to produce x-rays?

1.) Source of free electrons

2.) A means of acceleration

3.) A way to focus electrons

4.) A means to decelerate electrons

44
New cards

Where do we get free electrons?

thermionic emission

45
New cards

Filament coil heats up and is determined by?

mA

46
New cards

What side is the filament coil on, anode or cathode?

cathode (-)

47
New cards

A means for accelerating electrons is your tube potential, also known as?

kVp (energy)

48
New cards

What focuses electrons toward the anode (+) ?

focusing cup

49
New cards

What is a focusing cup made of?

molybdenum or nickel

50
New cards

Rapidly stopping electrons, also known as a sudden deceleration is done by the?

Anode (+)

51
New cards

What is the anode made of?

Tungsten or Rhenium