Volatile content
• when the presence of moisture or other volatile material may become critical
• Loss on Drying, Water Determination (for substances that appear to contain water as the only volatile constituent), Residual Solvents
Disintegration
• measures the time it takes for the dosage unit to disintegrate in an aqueous medium
• (exceptions) except for those intended to be chewed and for delayed- or extended-release products
basket-rack
assembly with 6 tubes held vertically upon a 10- mesh stainless steel wire screen, and water bath
Disintegration parameters
29-32 cycles/min; at least 15 mm below the surface of the fluid on the upward stroke and descends to NLT 25 mm from the bottom of the vessel on the downward stroke
disks
permitted only where specified or allowed in the monograph
o If is not specified on the monograph, there is no need to use the disk. The disk is used to prevent the dosage form to float.
immersion fluids maintained at 37 ± 2°C
distilled water (uncoated tablets, and effervescent granules and tablets), 0.1 M HCl or simulated gastric fluid and pH 6.8 buffer or simulated intestinal fluid (for delayed-release tablets and capsules)
first test
6 units must disintegrate, if not repeat the test on 12 additional units
second test
(additional 12 units) not fewer than 16 of the total of 18 tablets are disintegrated
Tablet friability
• determines the ability of tablets to withstand mechanical stresses and their resistance to chipping and surface abrasion
• applicable to most compressed, uncoated tablets
Friabilator
a tumbling apparatus where tablets are exposed to rolling or repeated shocks
rotate 100 times
o specifications: nmt 1% loss; for new formulations, nmt 0.8%; no capping or chipping
sample size: for tablets weighing ≤ 650 mg
For friability, use number of tablets weighing as near as possible to 6.5 grams; for tablets weighing over 650 mg – 10 tablets
Tablet breaking force
measures the mechanical integrity of tablets, which is the force required to cause them to fail (i.e., break) in a specific plane
Monsanto (Stokes) hardness tester
based on compressing tablets between two jaws via a spring gauge and screw
Pfizer hardness tester
the vertically mounted tablet was squeezed in a device that resembled a pair of pliers
Strong-Cobb hardness tester
the breaking load was applied through the action of a small hydraulic pump
Modern tester
more preferred because they employ mechanical drives, strain gauge-based load cells for force measurements, and electronic signal processing
Modern testers units
kiloponds or newtons {1 kilopond (kp) = 1 kilogram-force (kgf) = 9.80 newtons (N)}, Strong-Cobb Unt (SCU)