what do film-cassette combos gold

an image intensifying screen

2
New cards

what happens to the screen when exposed to light

glows briefly when exposed to light

3
New cards

what is most of an image formed by (film-cassette)

by light not by xrays

4
New cards

when is non-screen film radiography used

dental radiography

5
New cards

what is the cassette

rigid holder for the screen and film

  • Protects film from light


6
New cards

what happens if the film-screen has bad contact

decreased contrast and detail

7
New cards

why does the cassette need regular inspection and cleaning

  • Check rubber, hinged and latches

  • Visually check the screen for dirt/damage

  • Clean using distilled water and completely dry before used

  • Avoid light-leak artifacts


8
New cards

what are intensifying screens

suspension of phosphor crystals in a binder (absorb xray photons and emit them as light)

  • Screen emulsion is glued to plastic polymer base


9
New cards

what converts xrays to light on the intensifying screen

phosphor

10
New cards

what is film more sensitive to

light than xrays

11
New cards

fluorescence

instantaneous = good

12
New cards

phosphorescence

continues after stimulation has ceased = bad → “screen lag” or afterglow

13
New cards

how must film-screen combos match

speed, colour, manufacturer

14
New cards

what does speed of the screen mean

the speed to convert Xrays to light

  • slower speed = more energy needed to make an image


15
New cards

characteristics of blue system film-screen

screens emit blue light

  • Film is sensitive to blue light

  • BUT physically the film looks green


16
New cards

characteristics of green system film-screen

screens emit green light

  • Film is sensitive to green light

  • BUT physically the film look blue/purple


17
New cards

if a screen speed is slow…

more Xrays needed = thinner emulsion = increased spatial resolution

18
New cards

if a screen speed is fast…

fewer Xrays needed = thicker emulsion = decreased resolution/detail

19
New cards

is calcium tungstate emulsion old or new; slow or fast

older/slower

20
New cards

is rare earth emulsion old or new; slow or fast

newer/faster

21
New cards

what colour do Rare earth phosphors glow

green

  • Needs green receptive film


22
New cards

what colour do Calcium Tungstate phosphors glow

blue

  • Needs blue receptive film


23
New cards

what will happen if wrong film-screen combo

grey film lacking in contrast and will need to vastly increase the mAs to get an image

24
New cards

structure of radiographic film

  1. base

  2. emulsion

  3. supercoat

  4. anticurl / antihalation layer


25
New cards

what does the base layer consist of

MYLAR (gelo) non-tearing, non-flammable, translucent, and flexible

26
New cards

when is a single emulsion film used

high resolution imaging (CT and MRI)

  • Requires higher doses of radiation to make image


27
New cards

which side of a single emulsion film is dull and shiny

Emulsion side is dull; anticurl layer is shiny

28
New cards

which layer does a double emulsion film not have

no anticurl layer

  • both sides are emulsion

  • lower doses of radiation needed for image


29
New cards

purpose of the supercoat layer on film

hard protective gelatin layer to protect from processing

30
New cards

purpose of the anticurl/antihalation layer

single emulsion film only; prevents reflection/shadowing of image

  • Is removed during processing


31
New cards

what does the emulsion contain which helps with latent image formation

photosensitive elements, which are silver halides and a high quality gelatin (completely translucent, easily absorbs moisture)

32
New cards

what does the gelatin-silver halide network contain

sensitivity specks (tiny physical imperfections)

33
New cards

function of sensitivity specs in the gelatin-silver halide network

absorb different amounts of energy

34
New cards

what happens when silver halide molecules are activated

they are attracted to and congregate in the sensitivity specks

35
New cards

how do the specs congregate

Degree of attraction if proportional to energy (light/xrays) received which varies across the film

36
New cards

what is a latent image

a lattice network of silver halide particles

  • Very delicate and easily damaged


37
New cards

what will a high detail film have more of

sensitivity specs = more detailed image

38
New cards

what does processing turn silver halide into

  • clumps of black metallic silver → density on film


39
New cards

purpose of grids

reduce scatter radiation, improve image quality

40
New cards

when are grids used in film vs digital

film = if body part is >10 cm

digital = for all X-rays

41
New cards

where is the grid located

above the grid tray

42
New cards

which way do grid lines run

parallel to length of table

43
New cards

what is the grid ratio

ratio of height of the lead strips to the distance (interspace) between them

44
New cards

parallel grid line

vertical lead strips

  • more xrays absorbed around the edges, less in the middle


45
New cards

focused grid lines

angled lead strips

  • similar amount of primary beam is absorbed across the image

  • Must put the grid in the correct way and correct SID


46
New cards

what is a common step for error and artifacts to occur

processing

47
New cards

when is a dark room required for preocessing

manual or automatic processing

48
New cards

purpose of the safelight (RED)

Filters to eliminate blue and green light spectrum

49
New cards

order of film processing

  1. developer

  2. fixer

  3. washing/drying


50
New cards

which processing step converts the latent image to visible image

developer

51
New cards

what does developer convert

exposed (colourful) silver halide crystals to black metallic silver

52
New cards

what factors does developer depend on

Time vs Temp

  • too hot or too long = darkened image


53
New cards

which processing step removes unexposed silver halide crystals

fixer

54
New cards

what is the main ingredient in fixer

ammonium thiosulphate

  • Is acidic (pH of 4.1-4.4) → low pH preserves thiosulphate chemicals and facilitates hardening


55
New cards

what does fixer harden

the gelatin coating, prevents swelling of the gelatin coating after processing

56
New cards

which processing step removes chemical

washing/drying

57
New cards

what can poor washing lead to

yellowing due to residual ammonium thiosulphate

58
New cards

what does washing harden

the emulsion → increased resistance to scratches

59
New cards

what light are wash artifacts most visible by

reflected light vs transmitted light

60
New cards

manual processing

films are clipped onto hangers and manually dipped into tanks containing the developer, fixer, and wash solutions

  • Hung to dry

  • Extra rinse step

  • Film washing and drying takes much longer


61
New cards

automatic processing

rollers to transport film

  • Uses higher temperatures (36 vs 20)

  • Shorter drying time

  • Machine needs to be maintained REGULARLY


62
New cards

advantages to digital radiography (DR)

  • No film/dark room: save $$ on film and chemicals

  • Eliminates many sources of error/artifacts

  • Faster: image in seconds

  • Digital storage and transmission

  • Higher quality images


63
New cards

disadvantages to digital radiography (DR)

  • $$ to install

  • Wide range of technique: too easy to retake images to get a good one → dose creep

  • Technique charts may have higher kVp settings than needed

  • Increased risk of scatter and radiation exposure

  • High kVp compromised contrast for bone


64
New cards

advantages to computed radiography (CR)

  • Eliminates wet processing and film: faster and cheaper

  • Cassette cheaper to replace than true digital

  • Image is stored digitally


65
New cards

disadvantages to computed radiography (CR)

  • Only one plate at a time

  • Time consuming

  • If cassette not read ASAP will fade overtime (6-8 hours)


66
New cards

process of CR

  1. After stimulation with light/xrays, 50% of phosphor is converted to an unstable state

  2. Screen then processed by the CR reader

  3. Laser is used to read the image; converts electrons back to their stable state so the latent image becomes visible

  4. Can only process one plate at a time


67
New cards

how to reset a CR plate

flash it with a bright white light

  • Should be flashed every morning to refresh electrons


68
New cards

indirect DR conversion

converts xray to visible light, then to electronic signal

69
New cards

what layers are involved in indirect DR conversion

  1. Scintillation phosphor layer → structure of crystals helps to focus the light which improves image sharpness

  2. Photodiode layer → light converted into electrons

  3. Readout → Charged Couple Device (CDC) system

  4. TFT/Flat Panel system


70
New cards

direct DR conversion

converts directly to electronic signal (Xrays → electrons)

71
New cards

layers of direct DR conversion

Only one layer does the conversion → an amorphous selenium photoconductor

72
New cards

what do all DR conversions have

  • individual Detector Elements (DELs)

  • A Thin Film Transistor (TFT)

  • An analogue to digital converter


73
New cards

what are Detector Elements (DELs)

Each one has a way to collect radiation and convert it to electrons

74
New cards

what does a Thin Film Transistor (TFT) do

controls and amplifies electron flow in each DEL; like a switch

75
New cards

what does an analogue to digital converter do

converts electrons into a digital signal that computer can use

76
New cards

windowing

changes contrast

77
New cards

levelling

changes density (brightness)

78
New cards

contrast

difference in density between two object

79
New cards

contrast resolution

number of shades of grey (dynamic range)

  • Much better for digital vs film


80
New cards

A higher grid ratio means that:

More scatter radiation is absorbed

81
New cards

can screens can be replaced or moved from cassette to cassette

yes

82
New cards

If using rare earth screens, what spectrum of light must the film be sensitive to?

green range

83
New cards

what can a humid environment cause to screens

deterioration of screens resulting in screen artifacts

84
New cards

how is a CR plate read

using a fine beam laser

85
New cards

what is the CR plate sensitive to

xrays and light

86
New cards

why should silver be recovered if possible

  • It is a heavy metal and will contribute to pollution

  • It is not an expensive process

  • It can be done in the clinic

  • It is actual silver and is valuable


87
New cards

what is DICOM

a standard for the creation and exchange of digital medical images

88
New cards

which intensifying screens produce 3-4x the amount of visible light

rare earth phosphors

89
New cards

at what point is the image latent until

developing/processing

90
New cards

where is the photosensitive part of the xray film contained

within the emulsion layer and consists of silver salts such as silver bromide suspended in gelatin