defines if pacing parameters are programmed through the mapping system or external EP stimulator
vein mode
acquiring and delineating geomtry within a distinct anatomical region
euclidean distance
shortest possible distance bw two points
pace all
routes stimuli to both sphere 9 catheter lattice and mini surface electrodes in a rapid burst once every pacing period
geodesic distance
measures distance following curvature of map
what is pace all useful for
testing phrenic nerve capture
pace minis
routes pacing stimuli to all sphere 9 catheter mini surface electrodes in a rapid sequential burst once every pacing period
what are pace minis useful for
testing for exit block from a PV
STIM
delivers pacing as programmed

what are these
atrial trigger
ventricular trigger
stim trigger
what does new map do
creates capability for a new geometry and EA information
what does remap do
initializes the map using the chamber and geometry from an existing source map
what does clone do
creates a full copy of an existing source map including chamber, rhtyhm, WOI, geo, and electrogram data
what does import do
creates a geo shell from a surface file saved on an external usb drive
respiration gating
the % of stable respiratory cycle that geometry and electrogram data is collected
voltage threshold
min voltage required for activation timing to be assigned to the point
activation threshold
how sharp a signal must be to be annotated, measured in V/s with the purpose to discriminate between near-field and far-field activity

which part is when info is collected and which part is when info is not collected
collected: white portion
not collected: orange portion
fast remap
initializes a map using the active map’s geometry, chamber, rhythm, and WOI
what does it mean when the outer ring is smaller? when larger?
the displayed cath location is near the surface of the map
displayed cath location is further from the surface of the map
dashed line vs solid line around the catheter
dashed: reduced local impedance
solid: increased local impedance
what do the orange and blue outlines mean on the sphere globe
orange: F curve deflection
blue: D curve deflection
if the ablation button is gray or black, what could that be due to?
poor location quality
rapid cath movement
respiration gating
triggers not detected
V trigger signals are inconsistent with stim trigger
errors on patches/displacement
Tag Connect Thresholds for PFA and RF
RF: 7-8mm
PFA: 5 mm

describe these
1.utilizes integral sensor temp rise above baseline value to determine ablation tags
2.when checked, ablation tags are shaped, positioned, and oriented based on sensors meeting temp criteria
3.when checked, ablation tags are displayed as transparent shapes with a white border
4.when checked, ablation tags within geo shell are obstructed
RF temp threshold
50 degrees C
PF temp threshold
one or two degrees above pt’s temp
impedance measurement
measured between the ablation electrode and the return electrodes
flow
real time flow rate ml/min being delivered by the HexaFlow irrigation pump, with target flow shown in bottom left
temperature measurement in treat mode
real time temp in C measured with the ablation catheter; max temp across all temp sensors
level measurement in treat mode
real time current level % being delivered by the HexaGen RF generator or the PFA generator
during RF, level is reported as?? what about PFA??
RF: % of the max current squared
PFA: % of the max current
blue tag connect line
inter - ablation distance is greater than the user defined criteria
blue ablation tag
temp criteria for an ablation was not met
grey tag connect line
interablation distance is lower than the manually set criteris

what do the two orange lines mean?
bright orange: max temp for al 9 sensors
dark orange: other temp sensors
what is the hierarchy of triggers when multiple are selected?
stim
atrial
Ventricle
is it an appropriate WOI if it spans 250ms but the TCL is 256ms?
yes because the WOI covers approximately 100% of the TCL
rec triggers selected for sinus
A and V
rec triggers selected for internally paced
S and V
rec triggers selected for Afib
V
rec triggers selected for aflutter or tach
A
WOI boundaries for sinus
left boundary: pre p wave
right boudnary: pre QRS
WOI boundaries for internally paced
left boundary: post pacing spike +10ms
right: pre QRS
WOI boundaries for afib
left: ~300ms pre QRS
right:pre QRS
WOI boundaries for aflutter or tach
physician preference or 100% TCL or 50/50 around trigger

where is the ablation


when 1 sensor exceeds the user defined threshold, the ablation tag is flattened into a disc centered on the sensor that met criteria






inter ablation distance
distance between the closest temp sensore that have met the user defined temp threshold