Cuffed Tracheostomy Tube with Disposable Inner Cannula prevalence
Most common type used currently with a Cuffed
Cuffed Tracheostomy Tube with Disposable Inner Cannula feature
Most common with a removable inner cannula
Cuffed Tracheostomy Tube change requirement
No need to change unless cuff patency is in question.
-many hospitals change weekly
Fenestrated Tracheostomy Tube inner cannula
Contains a removable inner cannula
Fenestrated Tracheostomy Tube inner cannula and upper airway
When inner cannula is removed & balloon is deflated, patient may breathe through upper airway
Fenestrated Tracheostomy Tube used for
Often used for weaning purposed
Fenestrated Tracheostomy Tube mechanical ventilation
With cuff inflated & inner cannula inserted mechanical ventilation is possible
Silver Jackson Tracheostomy Tube permanence
May be used as a permanent trach
Silver Jackson Tracheostomy Tube material
Cuffless & made from sterling silver
-Silver is more durable & easier to clean and less likey to cause infection
Silver Jackson Tracheostomy Tube usage
Not very common usage these days
Communi-Trach (Pitt Speaking Tube)
Facilitates speech with an inflated cuff; blows O2 by vocal cords via fenestration.
Communi-Trach speech mechanics
Patient may speak without using their own expiratory gas flow; often difficult to coordinate; speech is not normal sounding.
Specialized Weaning Devices purpose
Maintain patency of stoma during weaning.
Specialized Weaning Devices examples
Olympic Trach-Talk, Passy-Muir Tracheostomy Valve, Trach Button, Kistner Button.
Olympic Trach-Talk
A Briggs adapter with a spring-loaded valve. Allows patient to inspire through trach, but must expire through upper airway. Trach cuff must be deflated.
Passy-Muir Tracheostomy Valve
Small one-way valve that facilitates speech. Cuff must be deflated. Considered the gold standard of speaking valves.
Switch to speaking or weaning device challenges
Difficult for some patients due to fear, anxiety, or muscular weakness
Starting specialized weaning device use strategy
Start small, coaxing the patient to use the device a few minutes a day & then build up
Trach Button purpose
Used to maintain the stoma after tracheostomy has been removed (decannulation)
Trach Button components
Consists of inner cannula, plug, IPPB adapter, spacers
Trach Button plug function
-Plug closes the stoma so patient can breathe through upper AW
-IPPB adapter allows mechanical ventilation (no cuff)
-Spacers adapt to varying neck thickness
Kistner Button definition
Trach button w/one-way valve (no cuff = no seal = no mechanical ventilation)
Kistner Button mechanical ventilation
Can't be used with mechanical ventilation
Kistner Button valve function
One-way valve forces patient to exhale through the upper airway (Facilitates for speech)
Kistner Button cough benefit
Allows for a more effective cough
Secretion Evacuation Devices examples
Suction Regulators, Suction Catheters
Suction Regulators function
Use a single-stage regulator to reduce the high negative pressure from the supply line to safe physiological levels
Suction Regulators pressure limit
Generally, we don't need to go over 120mmHg
Suction Regulators pressure reading
Must occlude to accurately read pressure!
Suction Catheters purpose
Long, narrow catheters designed to remove secretions, blood, or vomit from a patient's airway
Suction Catheters diameter restriction
Never use a catheter that has a diameter > ½ the diameter of the patients AW (either anatomical or artificial)
Suction Catheters technique
Insert the catheter & apply suction as catheter is being withdrawn
-No more than 10 seconds
Suction Catheters pre-oxygenation
Pre-hyperoxygenate always hyperventilate if necessary
Wright Respirometer mechanism
Is portable and uses rotating vanes & a gear mechanism
Wright Respirometer accuracy
Most accurate b/t 10-20 LPM
Wright Respirometer flow limit
Flow > 300 LPM may damage vein!
-Inertia may have some effect (an object in motion will stay in motion until something stops it)
Peak Flowmeters purpose
Pneumotachometer used to measure PEFR @ bedside
Wright peak flowmeter
Permanent/reusable
Mini-Wright peak flowmeter
Cheaper/lighter/plastic version
Assess peak flowmeter
Disposable; very common today; vertical
Asthma Care Plan - Green Zone
All is good, no symptoms/ PEFR 80-100%/ Take controller meds
Asthma Care Plan - Yellow Zone
Coughing and wheezing w/ increased WOB/Symptoms come on exertion, relieved with medication/ Peak flow 50-100%/ Take controller meds and rescue meds as needed, and call MD if in yellow zone for over 24 hours
Asthma Care Plan - Red Zone
Difficulty breathing, coughing and wheezing not helped by medication/ Trouble walking or talking due to symptoms/ Meds not working/ PEFR less than 50%/ CALL MD NOW, if no answer GO TO ER
CO-Oximetry Principle of Operation
Tells us if what is bound to Hb is O2 or something else
Pulse Oximetry Principle of Operation
Tells us if something is bound to Hb
Hgb Binding
Reversibly binds w/ O2: HbO2 , RHb
Carboxyhemoglobin (COHb)
Bound to CO, Affinity = 200-250 x
Methemoglobin (MetHb)
Cannot bind w/ O2, Brown in color
CO-Oximetry purpose
Determines fractional Hb using spectrophotometry
CO-Oximetry RBC handling
RBCs are hemolyzed to release Hb
CO-Oximetry isobestic point
Wavelength where 2 or more variants (Hb, HbO2, HbCO) absorb same amount of light
CO-Oximetry concentration calculation
By knowing amount of light absorbed concentration can be determined
End-Tidal CO2 principle of operation
Capnography
End-Tidal CO2 sensor positions
Mainstream, Sidestream, Proximal diverting
Colormetric End Tidal CO2 Determination
Principle of operation