ascending bellows
attached to base
descends during inspiration
collapse if circuit disconnects
slight resistance on inhalation 2-4 cmH2O
descending bellows
attached to top
compressed up during inspiration
refills due to gravity on bellows
will look likr working of circuit disconnected
where are asending bellows used
modern GE ventilators
where are descending bellows used
no longer in use
types of mechanical ventilation
conventional
high frequency ventilation
high frequency jet ventilation
high frequency oscillation
conventional ventilation
4-60 breaths/min
high frequency ventilation
60-110 breaths/min
high frequency jet ventilation
110-400 breaths/min
high frequency oscillation
400-2400 breaths per minutes
gas pushed into lung during inspiration and pulled out diring expiration
very low todal volumes less than dead space of lung
who is high frequency oscillation used in
neonates and adult patient populations to reduce lung injury
single circuit ventilator
piston
no bellows
no drive gas
double circuit ventilator
drive gas
pneumatic
ventilator drive gas
equal to minute ventilation (RRxTV)
GE Ohmeda drive gas
100% O2
regulated to 25 psi within ventilator engine
anesthetic gases inside bellows will be diluted if it tears
Drager Ventrui ventilator drive gas
entrains room air
30% O2 rest RO
piston driven ventilator
electrically driven piston
eliminates need for drive gas
in piston driven ventilator resovoir bag is not isolated from
ventilator and collects FGF entering breathing system
in piston driven ventilator breathing bag
will expand and contract during inspiration and expiration
modulates pressure increases in system
which machines use piston driven ventilator
drager fabius GS
Apollo ventilator
classification of positive pressure ventilators
inspiratory characteristics
cycling
inspiratory characteristics of positive pressure ventilators
consant flow generator
constant pressure generator
cyclining of positive pressure ventilators
time cycles ventilator
volume cycled ventilator
pressure cycled ventilator
flow cycled ventilator
volume controlled ventilation (VCV)
preset TV is delivered
constant flow rate
on fixed time interval regardless of patient efforts
in VCV practitioner sets
TV
RR
pressure limit (40 default)
PEEP
settings for VCV
TV
RR
pressure limit (40 default)
PEEP
I:E
T pause

VCV waveform characteristics
-pressure increases during inspiration
-peak pressure: reflects resistance
-plateau pressure: reflects lung compliance
-volume is linear increase

VCV waveform identifier
square form for flow


time cycled volume controlled
-constant volume regardless of lung dynamics of patient
-pressure will vary with compliance/resistance of patient
pressure limited part of VCV
patient with high compliance: lower pressure to deliver VT
patient with low compliance: higher pressure to deliver VT
VCV stops delivering TV when
pressure limit is reached
pressure controlled ventilation (PCV)
pre set pressure is delivered
tidal volume delivered is based by rise time and pressuer
what does practitioner set for PCV
inspiratory pressure
rate
pressure limit PEEP
waveform of PCV
quickly achieves set pressure and maintains until exhalation begins gas flows highest at beginning
VT fluctuates with what in PCV
changes in compliance and resistance
higher resistance and decreased compliance causes
decreased TV
time cycled pressure controlled
constant pressure regardless of lung compliance and resistance
tidal volume will vary with compliance/resistance
pressure waveform of PCV
square pressure due to constant pressure
settings for PCV are
Pinsp
RR
I:E
PEEP
Pmax
Rise rate
PCV waveform characteristics
peak flow remains the same but total flow fluctuates
volume fluctuates depending on flow
pressure is the same
maximum pressure formula
inspiratory pressure + PEEP
pressure mode benefits
-assists with breath delivery in patients who are difficult to ventilate or susceptible to barotrauma
-facilitates more even gas distribution
-improves recruitment of closed alveolar units
-delivers consistent pressure regardless of leakeage
pressure mode applications
-danger of high PIP, PCV to limit pressure within airway
-if compliance is low, PCV can obtain higher TV
-compensates for leaks
-eliminated breathing circuit compliance
when do you use PCV to limit PIP to limit pressure
LMA
emphysema
neonates/infants
when do you use PCV tfor low compliance to get higher TV
pregnancy
laparoscopic surgery
morbid obesity
ARDS
pressure vs volume control
both only used in mechanical ventilating not spontaneous
pressure form is square in PCV
flow form is square in VCV
volume ventilation gas distribution
gas follows path of least resistance and majority of preset tidal volume enters more compliant lung
increasing TV in volume ventilation in not compliant lung can cause
increased pressure and barotrauma
pressure ventilation gas ventilation
pressure in lungs equal throughout
-high initial flow rate puts a large bolus of gas in upper airways early in cycle, as flow decelerates more laminar flow at end makes even distribution
synchronized intermittent mandatory ventilation (SIMV)
synchronizes ventilator delivered breaths with patients breaths within trigger window
often volume controlled
patient breath within trigger window vs out (SIMV)
breath within window volume controlled breath
breath out of window pressure support breath
SIMV-PC
Synchronized intermittent mandatory ventilation with pressure control
you set the inspiratory pressure not TV
patient breath within trigger window vs out (SIMV-PC)
breath within window pressure controlled breath
breath out of window pressure support breath
SIMV-PSV
Synchronized intermittent mandatory ventilation with pressure support ventilation
patient breath within trigger window vs out (SIMV-PSV)
breath within window volume controlled breath
breath out of window pressure support breath
in SIMV the vent will always
deliver set number of breaths regarless of how much the patient is breathing
when is SIMV used
during emergence or anytime patient is not paralyzed
SIMV breath timing
timing of mechanical breath coincides with spontaneous breath to prevent breath stacking
mandatory breath given if no breath taken within window
SIMV waveform
mandatory breath wave looks how it should with patient breaths displayed as little blips
SIMV PSV settings
TV
RR
Tinsp
Rise rate
Tpause
Psupport
PEEP
Pmax
Trigger window
Flow trigger
End of breath
SIMV-PC settings
Pinsp
Psupport
Trigger window
rise rate
RR
PEEP
flow trigger
Tinsp
Pmax
End of breath
SIMV-PSV vs SIMV-PC
SIMV benefits
improved comfort in awake patient
less chance of bucking
patient can breath spontaneously without concern for hypoventilation/atelectasis
hypoventilation
not enough ventilation to eliminate CO2
low TV or RR
atelectasis
not enough oxygenation where alveoli close or not enough for gas exchange
Pressure Support Ventilation (PSV)
augmentation of tidal volume of spontaneously ventilation patient
patient initiates every breath and machine adds pressure support
settings for PSVPro
Psupport
flow trigger
back up time
PEEP
end of breath
Pinsp
trigger window
Pmax
RR
Tinsp
Rise rate
PSVPro to SIMV-PC
on some machines in PSVPro if patient goes a certain period of time without initiating a breath the machine will convert to SIMV-PC
what machines have SIMV-PC as backup for PSVPro
Aestiva
pressure controlled volume guarantee (PCVG)
combination of PCV and VCV
delivered preset TV with lowest possible pressure
first breath of PCVG
patient compliance is determine and inspiratory pressure levels determined for subsequent breaths
volume controlled breath
subsequent breaths of PCVG
pressure controlled breaths using amount of pressure determined by the first breath to deliver set TV
pressure range for subsequent breaths
low end: PEEP +2
high end: 5cm below Pmax
inspiratory pressure change between breaths mac +/- 3 cmH2O
practioner sets for PCVG
TV
RR
I:E
PEEP
Pmax
Rise rate
safety features of PCVG
tidal volume not delivered warning
if TV not achievable at Pmax less 5cmH2O
VCV mode
initiated by: time
limit: volume
cycle: volume/ time
PCV mode
initiated by: time
limit: pressure
cycle: time
SIMV mode
initiated by: time/pressure
limit: volume
cycle: volume/ time
PSV mode
initiated by: pressure/flow
limit: pressure
cycle: flow/ time
alveolar ventilaton formula
(TV- Dead space) x RR
dead space formula
anatomic + alveolor
usually 1/3 of TV
actual volume delivered formula (old machines)
ventilator + FGF - compliance loss
compliance lost formula
peak pressure x machine compliance
how to adujust for compliance
fresh gas decoupling
fresh gas compensation
fresh gas decoupling
means to prevent the fresh gas flow from affecting the tidal volume by isolating the fresh gas flow so that it doesn't enter the breathing system during inspiration
how does fresh gas decoupling work
piston ventilator closes a one way (decoupling) valve, diverting FGF to manual breathing bag during inspiration
how to tell a machine has fresh gas decoupling
breathing bag moves during ventilator operation
which machines have fresh gas decoupling
apollo
drager (fabius)
fresh gas compensation
means to prevent the fresh gas flow from affecting the tidal volume by measuring the actual tidal volume and using this information to change the volume of gas delivered by the ventilator
which machines have fresh gas compensation
aisys
avance
aestiva
inspiratory pause
lungs held inflated at fixed pressure or volume between inspiration and expiration
value of inspiratory pause
-provides time for alveoli with slower time constants to fill more
-provides diagnostic info about peak airway pressure, airway resistance, and static lung compliance
alveolar inflation time constants
increased resistance, increased time constant, take longer to fill
use inspiratory pause
alveolar inflation time constants formula
compliance x resistance
what does peak pressure tell us
pressure in due to gas flow through the resistance of circuit and airways
P=QR
why would peak pressure be increased
due to resistance changes
tube kinked, bronchospasm, increased flow
what does plateau pressure tell us
pressure due to static lung compliance and volume
P=V/C
why would plateau pressure be increased
due to compliance changes
pulmonary edema, pneumothorax, increased TV
auto PEEP
new inspiration begins before exhalation is complete
auto peep can cause
hyperinflation
barotrauma
pneumothorax
decreased venous return
auto peep can occur in
copd
ards