Exam 2 Intro to Systems Lecture 6: Vent Modes

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/100

encourage image

There's no tags or description

Looks like no tags are added yet.

101 Terms

1
New cards

ascending bellows

attached to base

descends during inspiration

collapse if circuit disconnects

slight resistance on inhalation 2-4 cmH2O

2
New cards

descending bellows

attached to top

compressed up during inspiration

refills due to gravity on bellows

will look likr working of circuit disconnected

3
New cards

where are asending bellows used

modern GE ventilators

4
New cards

where are descending bellows used

no longer in use

5
New cards

types of mechanical ventilation

conventional

high frequency ventilation

high frequency jet ventilation

high frequency oscillation

6
New cards

conventional ventilation

4-60 breaths/min

7
New cards

high frequency ventilation

60-110 breaths/min

8
New cards

high frequency jet ventilation

110-400 breaths/min

9
New cards

high frequency oscillation

400-2400 breaths per minutes

gas pushed into lung during inspiration and pulled out diring expiration

very low todal volumes less than dead space of lung

10
New cards

who is high frequency oscillation used in

neonates and adult patient populations to reduce lung injury

11
New cards

single circuit ventilator

piston

no bellows

no drive gas

12
New cards

double circuit ventilator

drive gas

pneumatic

13
New cards

ventilator drive gas

equal to minute ventilation (RRxTV)

14
New cards

GE Ohmeda drive gas

100% O2

regulated to 25 psi within ventilator engine

anesthetic gases inside bellows will be diluted if it tears

15
New cards

Drager Ventrui ventilator drive gas

entrains room air

30% O2 rest RO

16
New cards

piston driven ventilator

electrically driven piston

eliminates need for drive gas

17
New cards

in piston driven ventilator resovoir bag is not isolated from

ventilator and collects FGF entering breathing system

18
New cards

in piston driven ventilator breathing bag

will expand and contract during inspiration and expiration

modulates pressure increases in system

19
New cards

which machines use piston driven ventilator

drager fabius GS

Apollo ventilator

20
New cards

classification of positive pressure ventilators

inspiratory characteristics

cycling

21
New cards

inspiratory characteristics of positive pressure ventilators

consant flow generator

constant pressure generator

22
New cards

cyclining of positive pressure ventilators

time cycles ventilator

volume cycled ventilator

pressure cycled ventilator

flow cycled ventilator

23
New cards

volume controlled ventilation (VCV)

preset TV is delivered

constant flow rate

on fixed time interval regardless of patient efforts

24
New cards

in VCV practitioner sets

TV

RR

pressure limit (40 default)

PEEP

25
New cards

settings for VCV

TV

RR

pressure limit (40 default)

PEEP

I:E

T pause

26
New cards
<p>VCV waveform characteristics</p>

VCV waveform characteristics

-pressure increases during inspiration

-peak pressure: reflects resistance

-plateau pressure: reflects lung compliance

-volume is linear increase

knowt flashcard image


27
New cards

VCV waveform identifier

square form for flow

knowt flashcard image


<p>square form for flow</p><img src="https://assets.knowt.com/user-attachments/877f3435-0085-44f8-9081-3ecce374d47f.png" data-width="50%" data-align="center" alt="knowt flashcard image" style="display: block; width: 50%; margin-left: auto; margin-right: auto;"><p></p>
28
New cards

time cycled volume controlled

-constant volume regardless of lung dynamics of patient

-pressure will vary with compliance/resistance of patient

29
New cards

pressure limited part of VCV

patient with high compliance: lower pressure to deliver VT

patient with low compliance: higher pressure to deliver VT

30
New cards

VCV stops delivering TV when

pressure limit is reached

31
New cards

pressure controlled ventilation (PCV)

pre set pressure is delivered

tidal volume delivered is based by rise time and pressuer

32
New cards

what does practitioner set for PCV

inspiratory pressure

rate

pressure limit PEEP

33
New cards

waveform of PCV

quickly achieves set pressure and maintains until exhalation begins gas flows highest at beginning

34
New cards

VT fluctuates with what in PCV

changes in compliance and resistance

35
New cards

higher resistance and decreased compliance causes

decreased TV

36
New cards

time cycled pressure controlled

constant pressure regardless of lung compliance and resistance

tidal volume will vary with compliance/resistance

37
New cards

pressure waveform of PCV

square pressure due to constant pressure

38
New cards

settings for PCV are

Pinsp

RR

I:E

PEEP

Pmax

Rise rate

39
New cards

PCV waveform characteristics

peak flow remains the same but total flow fluctuates

volume fluctuates depending on flow

pressure is the same

40
New cards

maximum pressure formula

inspiratory pressure + PEEP

41
New cards

pressure mode benefits

-assists with breath delivery in patients who are difficult to ventilate or susceptible to barotrauma

-facilitates more even gas distribution

-improves recruitment of closed alveolar units

-delivers consistent pressure regardless of leakeage

42
New cards

pressure mode applications

-danger of high PIP, PCV to limit pressure within airway

-if compliance is low, PCV can obtain higher TV

-compensates for leaks

-eliminated breathing circuit compliance

43
New cards

when do you use PCV to limit PIP to limit pressure

LMA

emphysema

neonates/infants

44
New cards

when do you use PCV tfor low compliance to get higher TV

pregnancy

laparoscopic surgery

morbid obesity

ARDS

45
New cards

pressure vs volume control

both only used in mechanical ventilating not spontaneous

pressure form is square in PCV

flow form is square in VCV

46
New cards

volume ventilation gas distribution

gas follows path of least resistance and majority of preset tidal volume enters more compliant lung

47
New cards

increasing TV in volume ventilation in not compliant lung can cause

increased pressure and barotrauma

48
New cards

pressure ventilation gas ventilation

pressure in lungs equal throughout

-high initial flow rate puts a large bolus of gas in upper airways early in cycle, as flow decelerates more laminar flow at end makes even distribution

49
New cards

synchronized intermittent mandatory ventilation (SIMV)

synchronizes ventilator delivered breaths with patients breaths within trigger window

often volume controlled

50
New cards

patient breath within trigger window vs out (SIMV)

breath within window volume controlled breath

breath out of window pressure support breath

51
New cards

SIMV-PC

Synchronized intermittent mandatory ventilation with pressure control

you set the inspiratory pressure not TV

52
New cards

patient breath within trigger window vs out (SIMV-PC)

breath within window pressure controlled breath

breath out of window pressure support breath

53
New cards

SIMV-PSV

Synchronized intermittent mandatory ventilation with pressure support ventilation

54
New cards

patient breath within trigger window vs out (SIMV-PSV)

breath within window volume controlled breath

breath out of window pressure support breath

55
New cards

in SIMV the vent will always

deliver set number of breaths regarless of how much the patient is breathing

56
New cards

when is SIMV used

during emergence or anytime patient is not paralyzed

57
New cards

SIMV breath timing

timing of mechanical breath coincides with spontaneous breath to prevent breath stacking

mandatory breath given if no breath taken within window

58
New cards

SIMV waveform

mandatory breath wave looks how it should with patient breaths displayed as little blips

59
New cards

SIMV PSV settings

TV

RR

Tinsp

Rise rate

Tpause

Psupport

PEEP

Pmax

Trigger window

Flow trigger

End of breath

60
New cards

SIMV-PC settings

Pinsp

Psupport

Trigger window

rise rate

RR

PEEP

flow trigger

Tinsp

Pmax

End of breath

61
New cards

SIMV-PSV vs SIMV-PC

62
New cards

SIMV benefits

improved comfort in awake patient

less chance of bucking

patient can breath spontaneously without concern for hypoventilation/atelectasis

63
New cards

hypoventilation

not enough ventilation to eliminate CO2

low TV or RR

64
New cards

atelectasis

not enough oxygenation where alveoli close or not enough for gas exchange

65
New cards

Pressure Support Ventilation (PSV)

augmentation of tidal volume of spontaneously ventilation patient

patient initiates every breath and machine adds pressure support

66
New cards

settings for PSVPro

Psupport

flow trigger

back up time

PEEP

end of breath

Pinsp

trigger window

Pmax

RR

Tinsp

Rise rate

67
New cards

PSVPro to SIMV-PC

on some machines in PSVPro if patient goes a certain period of time without initiating a breath the machine will convert to SIMV-PC

68
New cards

what machines have SIMV-PC as backup for PSVPro

Aestiva

69
New cards

pressure controlled volume guarantee (PCVG)

combination of PCV and VCV

delivered preset TV with lowest possible pressure

70
New cards

first breath of PCVG

patient compliance is determine and inspiratory pressure levels determined for subsequent breaths

volume controlled breath

71
New cards

subsequent breaths of PCVG

pressure controlled breaths using amount of pressure determined by the first breath to deliver set TV

72
New cards

pressure range for subsequent breaths

low end: PEEP +2

high end: 5cm below Pmax

inspiratory pressure change between breaths mac +/- 3 cmH2O

73
New cards

practioner sets for PCVG

TV

RR

I:E

PEEP

Pmax

Rise rate

74
New cards

safety features of PCVG

tidal volume not delivered warning

if TV not achievable at Pmax less 5cmH2O

75
New cards

VCV mode

initiated by: time

limit: volume

cycle: volume/ time

76
New cards

PCV mode

initiated by: time

limit: pressure

cycle: time

77
New cards

SIMV mode

initiated by: time/pressure

limit: volume

cycle: volume/ time

78
New cards

PSV mode

initiated by: pressure/flow

limit: pressure

cycle: flow/ time

79
New cards

alveolar ventilaton formula

(TV- Dead space) x RR

80
New cards

dead space formula

anatomic + alveolor

usually 1/3 of TV

81
New cards

actual volume delivered formula (old machines)

ventilator + FGF - compliance loss

82
New cards

compliance lost formula

peak pressure x machine compliance

83
New cards

how to adujust for compliance

fresh gas decoupling

fresh gas compensation

84
New cards

fresh gas decoupling

means to prevent the fresh gas flow from affecting the tidal volume by isolating the fresh gas flow so that it doesn't enter the breathing system during inspiration

85
New cards

how does fresh gas decoupling work

piston ventilator closes a one way (decoupling) valve, diverting FGF to manual breathing bag during inspiration

86
New cards

how to tell a machine has fresh gas decoupling

breathing bag moves during ventilator operation

87
New cards

which machines have fresh gas decoupling

apollo

drager (fabius)

88
New cards

fresh gas compensation

means to prevent the fresh gas flow from affecting the tidal volume by measuring the actual tidal volume and using this information to change the volume of gas delivered by the ventilator

89
New cards

which machines have fresh gas compensation

aisys

avance

aestiva

90
New cards

inspiratory pause

lungs held inflated at fixed pressure or volume between inspiration and expiration

91
New cards

value of inspiratory pause

-provides time for alveoli with slower time constants to fill more

-provides diagnostic info about peak airway pressure, airway resistance, and static lung compliance

92
New cards

alveolar inflation time constants

increased resistance, increased time constant, take longer to fill

use inspiratory pause

93
New cards

alveolar inflation time constants formula

compliance x resistance

94
New cards

what does peak pressure tell us

pressure in due to gas flow through the resistance of circuit and airways

P=QR

95
New cards

why would peak pressure be increased

due to resistance changes

tube kinked, bronchospasm, increased flow

96
New cards

what does plateau pressure tell us

pressure due to static lung compliance and volume

P=V/C

97
New cards

why would plateau pressure be increased

due to compliance changes

pulmonary edema, pneumothorax, increased TV

98
New cards

auto PEEP

new inspiration begins before exhalation is complete

99
New cards

auto peep can cause

hyperinflation

barotrauma

pneumothorax

decreased venous return

100
New cards

auto peep can occur in

copd

ards