Toe pressures are a more accurate method of evaluating -
distal limb & foot perfusion in patients with falsely elevated segmental limb pressures
Toe pressures give similar -
clinical results as ABIs
During a toe pressure a digital cuff is placed at -
base of big toe
During a toe pressure BP is obtained using a -
photoplethysmography sensor and a TBI is reported
TBI stands for -
Toe-Brachial Index
digital cuffs used on the toes have a diameter of about -
2.0 to 2.5 cm
during toe pressure a PPG transducer is applied to -
digit distal to digital cuff
PPG transducers will detect waveforms from -
microcirculations arterial pulses (small pulses)
A waveform is obtained with the -
PPG transducer
Toe pressure (PPG transducer) normal waveform demonstrates -
brisk systolic upstroke
well-defined peak
concave shape on return to baseline during diastole
Toe pressure examination method -
cuff inflated until waveform disappears
systolic pressure is defined at point when wave form reappears
during a toe pressure the waveform is used to determine -
systolic pressure
Normal Toe-Brachial index (TBI) -
≥ 0.8
In toe pressures TBIs are useful when -
ankle vessels are incompressible
calcified vessel in PTA or DPA
toe pressure of 50 mm Hg indicates -
adequate pressure for healing
Calcifed ankle vessels would be an indication to do a -
TBI
TBIs are used to evaluate the -
healing process