Intra-aortic Balloon Pump

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/13

encourage image

There's no tags or description

Looks like no tags are added yet.

14 Terms

1
New cards

What are the indications for IABP therapy?

-acute MI w/ cardiogenic shock, post-coronary artery bypass graft (CABG)

2
New cards

What are the contraindications for IABP therapy?

-aortic insufficiency/disease, severe peripheral vascular disease, or aortic regurg

3
New cards

Increase coronary perfusion & decrease the workload on the heart

Goals of IABP therapy

4
New cards

Pressure and EKG

What are the triggers for IABP?

5
New cards

Proper IABP Placement

-distal tip = 2-3 cm below the subclavian artery

-Proximal end = above renal arteries

-at the level of the carina

6
New cards

How to verify correct placement

-intact left radial pulse

-ensure adequate urine output (0.5 mL/kg/hr)

-CXR

7
New cards

How to manually inflate the balloon pump

-occurs with pump power failure

-manual pumping = inflate every 3-5 min (knock off clots)

-determine amount to inflate balloon with based on size of ballon placed

8
New cards

-keep balloon pump running

-switch to pressure trigger

What do you do during a cardiac arrest?

9
New cards

-balloon has ruptured

-very bad thing

-alert receiving facility

What do rust colored flakes in the line mean?

10
New cards

Normal IABP waveform

knowt flashcard image
11
New cards

Early Inflation

-inflation prior to aortic valve closure

-HARMFUL

-’U’shape

-forces blood back into LV

<p>-inflation prior to aortic valve closure </p><p>-HARMFUL </p><p>-’U’shape </p><p>-forces blood back into LV </p>
12
New cards

Late inflation

-inflation after aortic valve closure

-’W’ shape

<p>-inflation after aortic valve closure </p><p>-’W’ shape </p>
13
New cards

Early Deflation

-deflation of the balloon before systole

-’cliff’ shape

<p>-deflation of the balloon before systole </p><p>-’cliff’ shape </p>
14
New cards

Late Deflation

-HARMFUL

-deflation is occurring as the aortic valve is opening

-'widened appearance’ shape

-Lethal Dose

<p>-HARMFUL </p><p>-<span>deflation is occurring as the aortic valve is opening</span></p><p>-'widened appearance’ shape </p><p>-<strong>L</strong>ethal <strong>D</strong>ose </p>