RT Protection
Scatter poses the greatest occupational hazard to the RT
The patient is the source of scatter in rad exams (fluoro, mobile, diagnostic)
At a 90 degree angle, at a distance of 3 feet from the patient, the scattered x ray intensity is approximately 0.1% of the original intensity of the primary beam
Never stand in the primary beam
ALARA Concept
Apply cardinal principles
Collimate the beam
In 1954 the National Committee on Radiation Protection (Later known as the National Council on Radiation Protection and Measurements) put forth the principle that radiation exposures should be kept “as low as reasonably achievable” with consideration for economic and societal factors
The ALARA concept is accepted by all regulatory agencies
The ALARA concept adopts an extremely conservative model with respect to the relationship between ionizing radiation and potential risk
Occupational Risk
Keep effective dose limits as low as possible (do not keep dose in upper range of limit, ex: annual dose right below or at 50 mSv)
The risk to a radiographer from radiation exposure may be equated with occupational risk in other industries that are generally considered reasonable safe
NCRP proposed that radiation protection programs for radiation workers be designed to prevent individual workers from having a total external plus internal cumulative EfD in excess of their age in years times 10 mSv
Dose Reduction Methods
Avoid repeat exams
The patient as a source of scattered radiation
Scattered radiation occupational hazard
Filtration of the diagnostic beam
Protective apparel
Technical exposure factors
Patient Restraint
RT should never hold
First choice - Use mechanical restraining devices
Last resort - Use family member or friend of pt, provide shielding, instructions for position to hold; other hosp personnel should not be used to hold routinely
Distance: Inverse Square Law
(d2/d1)2
The ISL also implies that if a radiographer moves closer to a source of radiation, the radiation exposure to the radiographer drastically increases
Ex: if the radiographer stands 2 feet away from an x ray source instead of 6 feet away the radiographer’s radiation exposure increases by a factor of 9
Shielding
Structural Protective Barriers
Most common materials; lead and concrete
Accessory protective devices: Lead impregnated vinyl
Ex: Apron, gloves, protective eye wear
Protective Structural Barriers
Doors to X-ray room
Must be closed during exposures; attenuation equivalent to 1/32” lead
Floors and ceiling
Usually dose not require shielding if made of concrete (Provides adequate shielding from scatter radiation)
Primary Protective Barrier
Prevents direct, un-scattered radiation
Located perpendicular to the line of travel to the x-ray beam
1/16” lead and extends 7 feet upward from floor
Secondary Protective Barrier (Leakage and Scatter)
Ceiling is secondary protective barrier
Should overlap the primary barrier by 1/2”
Consist of 1/32” lead
Control-booth barrier - Protection of radiographer
Must extend 7 feet upward from floor
Lead glass window - 1.5mm lead equivalent
Clear lead-plastic secondary protective barrier
7 feet upward from the floor, lead equivalencies from 0.3- 2mm
Ex: special procedures open view window
Typically 0.5mm lead equivalencies
Accessory Protective Devices: Lead Aprons and Gloves
At 100 kVp a protective lead apron must be equivalent to at least 0.25mm thickness lead equivalent
A lead apron of 0.5 or 1 mm lead equivalent offer much more protection
0.25 mm lead would be appropriate for mammography
0.5mm lead equivalent is most widely used and recommended for diagnostic imaging
Wrap around is recommended for interventional studies
Accessory Protective Devices
Neck and Thyroid Shields
Minimum of 0.5mm lead equivalent
Lead gloves
0.25mm lead equivalent is recommended
Protective eyeglasses
Minimal lead equivalent of 0.35mm
Care of Protective Garments
Inspect annually for cracks
Inspect visually and fluoroscope or take radiographs using high kVp
Store on properly designed racks (hanging) or lay flat on table (not folded on x-ray unit)
Remove from service if crack exists
RT Protection During Fluoro
Protective Curtain
A protective curtain of at least 0.25 mm lead equivalent or retractable panel mounted on side of table of at least 0.5mm lead equivalent should be positioned between the fluoroscopist and patient to minimize scatter to personnel during fluoroscopy
Standing position
Standing at right angles to the patient decreases radiologist and RT dose
Mobile Fluoro and RT Protection During Mobile Procedures
C-arm Fluoro (Mobile)
Scatter radiation is more intense with x-ray tube over patient
Portable X-ray Machine
Length of Exposure Cord
The exposure switch must allow the RT to remain at least 6ft from the x-ray tube during the exposure
RT must stand at maximum distance from pt, tube and useful beam
RT should attempt to stand at a right angle to the x-ray beam
Diagnostic X-ray Suite Protection Design
Radiation Shielding Categories
Primary Radiation
Secondary Radiation
Scatter
Leakage
Over Shielding: It is better to spend extra money up front and over shield than to not meet requirements later, if regulatory standards change. Follows ALARA principle
Radiation Absorbent Barrier Design Considerations
The mean energy of the x-rays that will strike the barrier
Whether the barrier is of primary or secondary nature
The distance from the x ray source to a position of occupancy 0.3m from the barrier
The workload of the unit
The use factor of the unit
The occupancy factor behind the barrier
The intrinsic shielding (Tube housing attenuation) of the x ray unit
Whether the area beyond the barrier is “controlled” or “uncontrolled”
Calculation Considerations
Workload (W)
Inverse Square Law
Use Factor (U) (beam direction factor)
Occupancy Factor (T)
Workload (W)
Describes weekly radiation use of a diagnostic x ray unit
Radiation “on time”
Indicates the level of radiation activity in a room
Amount of use of the x ray unit; the time the x ray tube is energized/week
Includes the voltage & mA of the generator, daily patient load (# of exams/wk) measured in milliampere-minutes per week (mA-min/wk)
Busy room = 1500-1000 mA-min/wk
Private office = 100 mA-min/wk
Formula: W = mAs/patient x # of patients/hr x # hrs/wk/ 60 min
Use Factor (U) or Beam Direction Factor
Quantity that indicates the proportional amount of time during which the x ray beam is energized and directed toward a particular barrier (wall, floor, ceilings)
Full Use (U = 1)
Floors of radiation rooms except dental installations, doors, walls, and ceilings of radiation rooms exposed routinely to the primary beam
Partial Use (U = 1/4)
Doors and walls of radiation rooms not exposed routinely to the primary beam; also floors of dental installations
Occasional Use (U = 1/16)
Ceilings of radiation rooms not exposed routinely to the primary beam; because of the low use factor, shielding requirements for a ceiling usually determined by secondary rather than primary beam considerations
Time of Occupancy Factor (T)
Used to indicate the fraction of time that an area of the hospital (outside the exposed area) may be occupied by a given individual
Designed to overestimate the fraction of time that a given individual might actually spend in an area; more than ensures that exposure is kept low
Two Types:
Uncontrolled areas
Areas occupied by non rad workers (general public), those subject to a dose limit of 10 mrem/week; shielding lowers the dose equiv limit to 2mrem/wk
The walls must have 1-2 tenth value layers more than controlled areas
Included hallways, waiting rooms, stairways, exterior of hosp, toilets, shops, nurses stations, offices, kids play areas
Takes on different values depending on purpose of area
Controlled areas
Areas occupied by radiation workers; those who wear dosimeters
Must reduce the exposure rate/ dose equiv limit to < 100 mrem/wk
T value = 1
New Approaches to Shielding
More rigorous analysis of workload using kVp actually used
Abandon use of adding one HVL if leakage and scatter barrier requirement are similar - use exact calculations
More accurate estimate of use factor to reflect true percentage of time the beam is directed at various barriers
More realistic occupancy (T) values for low occupancy areas
Workload: More realistic use of contemporary survey data
Leakage and scatter: Explicit barrier calculations
Use factor: Adjusted for beam direction data reflecting actual usage patterns
Occupancy factor: Realistic assumptions of occupancy of low occupancy areas
Pregnant/Potentially Pregnant RT
Declaration of pregnancy
Provide written notification of pregnancy or suspected pregnancy to supervisor
Termination or involuntary leave of absence or altered work schedule not allowed
Counseling (during pregnancy while continuously employed)
Review of monitoring records (previous rad exposure hx)
Review of dose limit during pregnancy (for fetus)
DL for period of preg 0.5 rem (5 mSv) DL/mo 50 mrem (0.5 mSv)
Personnel Monitoring Devices
2nd dosimeter worn at waist level under apron to monitor fetal exp
Must maintain exposure level on a separate record from mother and identify as fetal exp
Review of proper radiation protection principles and practices
Utilize ALARA
Cardinal principles of radiation
Protective apparel ( 0.5mm lead apron; wrap around type can be used)
Read and sign form acknowledging rad risk