1
New cards

Complete Ventilator Cycle 4 Phases

1.Starting Inspiration

2.Inspiration

  1. Ending Inspiration

  2. Expiration


2
New cards

What are some triggers

Pressure, Volume, Flow or time

3
New cards

How do you calculate TCT

60/RR

4
New cards

What is the usual ventilator trigger?

Time based trigger on Respiratory Rate

5
New cards

Bias flow

A constant flow of gas delivered to the patient, aiding in triggering and sustaining ventilation during mechanical support.

6
New cards

How to notice patient efforts

It’s most noticeable on pressure waveform if you see a drop below baseline

7
New cards

What to do if you notice patient efforts

Decrease the pressure sensitivity to make it easier for mandatory breaths to occur

8
New cards

Normal Trigger

Inspiration starts within an acceptable delay (usually 100ms) after the start of the patient’s inspiratory effort

9
New cards

Early Trigger

Inspiratory effort appears after the machine triggers inspiration

10
New cards

Late Trigger

Inspiration starts beyond an acceptable delay after the start of an inspiratory effort

11
New cards

False Trigger

Trigger event not caused by the patient’s respiratory effort

12
New cards

What kind of trigger uses proximal flow sensor

Volume Triggers

13
New cards

4 cycle variables

Pressure, Flow, Volume, Time

14
New cards

Cycle

Ends Inspiratory phase and simultaneously starts expiratory phase

15
New cards

High Pressure limit

A precautionary limit that is set before setting patient up to ventilator. If reached inspiration will terminate and an alarm will occur

16
New cards

Esens

The ventilator terminates inspiration and starts expiration when flow decays to a certain percentage. (Usually set between 20%-30%)

17
New cards

Spontaneous Breath

Patient has full control over timing and volume of breath. Can be assisted or unassisted

18
New cards

Patient Triggering

Flow, pressure, volume, electrical impulse, esophageal monitoring.

19
New cards
20
New cards
21
New cards
22
New cards
23
New cards
24
New cards
25
New cards
What happens if high pressure is detected during a ventilator breath, regardless of the intended cycling variable?
The ventilator immediately cycles to expiration, even if volume, time, or flow cycling was intended.