what is milliamperage (mA) directly proportional to
radiation quantity exiting x-ray tube
what is milliamperage (mA) inversely related to and why
exposure time - to maintain exposure to IR
what is exposure time (s) directly proportional to
radiation quantity exiting x-ray tube
what is exposure time (s) inversely related to and why
mA - to maintain exposure to IR
what does increase in mAs cause
increased exposure to IR
what does decrease in mAs cause
reduced exposure to IR
what is kilovoltage (kVp) directly related to
radiation quality
quality exiting x-ray tube
what is kilovoltage (kVp) inversely related to
radiographic contrast
what is mAs directly proportional to
quantity of radiation reaching IR
how does mAs affect quantum noise
increased mAs = decreased noise
decreased mAs = increased noise
inversely proportional relation
what mA and exposure time values affect mAs
double s to fixed mA
double mA to fixed s
what is relation between mA and exposure time when maintaining mAs
mA double, exposure time halved
exposure time doubled, mA halved
what happens to exposure time is mA is halved when maintaining
halved
what technical factor is image brightness not dependent on
mAs
kVp
what technical factor is image contrast not dependent on
kVp
what level of technical factors is recommended in digital imaging and why
high kVp low mAs
reduces quantum noise which cannot be fixed in postprocessing unlike contrast and brightness
what are challenges with mAs
generators
patients age and condition
thickness of anatomic part
quantum noise
how do generators affect mAs
single phase produce less radiation with same mAs compared to high frequency
what is exposure indicator
numerical value on processed image indicating level of exposure IR received
how to mAs be adjusted for repeat exposures
factor of 2
what does kVp affect in IR and how
exposure to IR - alters amount and penetrating ability of x-ray beam
what does kVp affect in body and how
penetration and absorption in anatomic tissue
increased kVp = more penetration and absorption
decreased kVp = less penetration and absorption
directly proportional
what does kVp affect in an image and how
subject contrast
increased kVp = decreased subject contrast
decreased kVp = increased subject contrast
inversely proportional
why does high kVp produce lower contrast
increase penetration = less absorption, more transmission = less variation in beam intensity exiting patient = low subject contrast
why does low kVp produce high contrast
decreased penetration = more absorption, less transmission = more variations in beam intensities exiting patients = high subject contrast
what technical factors increase exposure to IR
increase mAs, exposure time, or kVp
what technical factors decrease exposure to IR
decreased mAs, exposure time, or kVp
what relation of technical factors maintain exposure to IR
increase mA and proportionally decrease time
increase time and proportionally decrease mA
increase kVp 15% and decrease mAs by half
decrease kVp by 15% and double mAs
what could be considered when selecting mAs in digital IR
keeping the values within IR exposure latitude
why is kVp not typically manipulated when varying IR
it affects subject contrast
what can overcome negative impacts of high kVp
computer processing
windowing displayed imaged
in what situations should kVp be adjusted to maintain exposure
portable
mobile x-ray
what is the 15% rule
altering the kVp by 15% has the same effect as doubling or halving the mAs
how to use 15% rule without maintaining mAs
multiple by 1.15 for 15% increase
multiple by .85 for 15% decrease
how to use 15% rule with maintaining mAs
multiple by 1.15 for 15% increase, divide mAs by 2
multiple by .85 for 15% decrease, multiply mAs by 2
what are secondary factors that affect images
focal spot size
SID
OID
alignment of CR to anatomy to IR
grids
beam restriction
generator output
tube filtration
patient factors
what does size of focal spot affect
unsharpness
spatial resolution
what does increasing focal spot size do
increase unsharpness
decrease spatial resolution
what does decreasing focal spot size do
decrease unsharpness
increase spatial resolution
what does SID affect
amount of radiation reaching patient and IR
what is the relationship between radiation intensity and distance
increasing SID decreases intensity reaching IR
decreasing SID increases intensity reaching IR
inverse square law
how does SID affect size distortion and recorded detail/spatial resolution
increased SID = decreased magnification and increased spatial resolution
decreased SID = increased magnification and decreased spatial resolution
formula for maintaining exposure with SID and mAs
mAs1 / mAs2 = (SID1)2 / (SID1)2
what happens with increased OID
decrease in intensity reaching IR
less scatter reaching IR due to beam divergence
potentially increased contrast
increased magnification
reduced spatial resolution
what is the purpose of air gap
reduce amount of scatter radiation reaching IR
how does air gap work
increasing OID (10-15cm) causes scatter to miss receptor thus reducing fog and creating sharper images
what are advantages of air gap
improved image contrast without needing gird
no grid artifacts to affect image quality
improved clarity
what are disadvantages of air gap
increased magnification/distortion due to increased OID
increased exposure due to compensating for OID to have good image quality
what does increasing OID affect
decreased exposure to IR
increased contrast and magnification
decreased spatial resolution
what does decreasing OID affect
increase exposure to IR
decrease contrast and magnification
increase spatial resolution
what does inverse square law state
intensity of x-ray beam is inversely proportional to the square of the distance from the source
what does SID affect in inverse square law
quantity of radiation reaching IR
what is the formula for inverse square law and SID
i1 / i2 = (D2)2 / (D1)2
what is magnification factor
to observe the effect of distance on size distortion
how to calculate Magnification Factor (MF)
SID / SOD
how to calculate source to object distance (SOD)
SID - OID
how to calculate object size
image size - MF
what does MF of 1 mean
no magnification
what can cause shape distortion of anatomic area
inaccurate CR alignment to tube, part, or IR
what does reducing scatter that reaches IR affect
image quality
what does reducing scatted do
increase contrast
when are grids recommended
for body parts thicker than 10 cm
60 kVp or higher
how to grids reduce amount of radiation reaching IR
absorb some transmitted radiation exiting patient
what does a more efficient grid do
create better contrast
what is grid conversion formula
mAs1 / mAs2 = grid ratio1 / grid ratio2
what is the conversion factors for grid ratios
no grid = 1
5:1 = 2
6:1 = 3
8:1 = 4
12:1 = 5
16:1 = 6
what does a larger field size affect
increases amount of tissue irradiated
production of more scatter
increased amount of radiation reaching IR
what does a smaller field size affect
reduced amount of tissue radiated
reduced amount of scatter produced
reduced amount of radiation reaching IR
what do high frequency generators require
lower exposure technique settings
when does tube filtration affect radiation quality
excessive or insufficient filtration
what are different type of body habitus
hypersthenic
sthenic
hyposthenic
asthenic
what does thickness of the body part affect
beam attenuation (increases with thickness)
scatter production
image contrast