Rad-103: Ch 7 Exposure Technique Factors

0.0(0)
Studied by 0 people
call kaiCall Kai
learnLearn
examPractice Test
spaced repetitionSpaced Repetition
heart puzzleMatch
flashcardsFlashcards
GameKnowt Play
Card Sorting

1/71

encourage image

There's no tags or description

Looks like no tags are added yet.

72 Terms

1
New cards

what is milliamperage (mA) directly proportional to

radiation quantity exiting x-ray tube

2
New cards

what is milliamperage (mA) inversely related to and why

exposure time - to maintain exposure to IR

3
New cards

what is exposure time (s) directly proportional to

radiation quantity exiting x-ray tube

4
New cards

what is exposure time (s) inversely related to and why

mA - to maintain exposure to IR

5
New cards

what does increase in mAs cause

increased exposure to IR

6
New cards

what does decrease in mAs cause

reduced exposure to IR

7
New cards

what is kilovoltage (kVp) directly related to

radiation quality

quality exiting x-ray tube

8
New cards

what is kilovoltage (kVp) inversely related to

radiographic contrast

9
New cards

what is mAs directly proportional to

quantity of radiation reaching IR

10
New cards

how does mAs affect quantum noise

increased mAs = decreased noise

decreased mAs = increased noise

inversely proportional relation

11
New cards

what mA and exposure time values affect mAs

double s to fixed mA

double mA to fixed s

12
New cards

what is relation between mA and exposure time when maintaining mAs

mA double, exposure time halved

exposure time doubled, mA halved

13
New cards

what happens to exposure time is mA is halved when maintaining

halved

14
New cards

what technical factor is image brightness not dependent on

mAs

kVp

15
New cards

what technical factor is image contrast not dependent on

kVp

16
New cards

what level of technical factors is recommended in digital imaging and why

high kVp low mAs

reduces quantum noise which cannot be fixed in postprocessing unlike contrast and brightness

17
New cards

what are challenges with mAs

generators

patients age and condition

thickness of anatomic part

quantum noise

18
New cards

how do generators affect mAs

single phase produce less radiation with same mAs compared to high frequency

19
New cards

what is exposure indicator

numerical value on processed image indicating level of exposure IR received

20
New cards

how to mAs be adjusted for repeat exposures

factor of 2

21
New cards

what does kVp affect in IR and how

exposure to IR - alters amount and penetrating ability of x-ray beam

22
New cards

what does kVp affect in body and how

penetration and absorption in anatomic tissue

increased kVp = more penetration and absorption

decreased kVp = less penetration and absorption

directly proportional

23
New cards

what does kVp affect in an image and how

subject contrast

increased kVp = decreased subject contrast

decreased kVp = increased subject contrast

inversely proportional

24
New cards

why does high kVp produce lower contrast

increase penetration = less absorption, more transmission = less variation in beam intensity exiting patient = low subject contrast

25
New cards

why does low kVp produce high contrast

decreased penetration = more absorption, less transmission = more variations in beam intensities exiting patients = high subject contrast

26
New cards

what technical factors increase exposure to IR

increase mAs, exposure time, or kVp

27
New cards

what technical factors decrease exposure to IR

decreased mAs, exposure time, or kVp

28
New cards

what relation of technical factors maintain exposure to IR

increase mA and proportionally decrease time

increase time and proportionally decrease mA

increase kVp 15% and decrease mAs by half

decrease kVp by 15% and double mAs

29
New cards

what could be considered when selecting mAs in digital IR

keeping the values within IR exposure latitude

30
New cards

why is kVp not typically manipulated when varying IR

it affects subject contrast

31
New cards

what can overcome negative impacts of high kVp

computer processing

windowing displayed imaged

32
New cards

in what situations should kVp be adjusted to maintain exposure

portable

mobile x-ray

33
New cards

what is the 15% rule

altering the kVp by 15% has the same effect as doubling or halving the mAs

34
New cards

how to use 15% rule without maintaining mAs

multiple by 1.15 for 15% increase

multiple by .85 for 15% decrease

35
New cards

how to use 15% rule with maintaining mAs

multiple by 1.15 for 15% increase, divide mAs by 2

multiple by .85 for 15% decrease, multiply mAs by 2

36
New cards

what are secondary factors that affect images

focal spot size

SID

OID

alignment of CR to anatomy to IR

grids

beam restriction

generator output

tube filtration

patient factors

37
New cards

what does size of focal spot affect

unsharpness

spatial resolution

38
New cards

what does increasing focal spot size do

increase unsharpness

decrease spatial resolution

39
New cards

what does decreasing focal spot size do

decrease unsharpness

increase spatial resolution

40
New cards

what does SID affect

amount of radiation reaching patient and IR

41
New cards

what is the relationship between radiation intensity and distance

increasing SID decreases intensity reaching IR

decreasing SID increases intensity reaching IR

inverse square law

42
New cards

how does SID affect size distortion and recorded detail/spatial resolution

increased SID = decreased magnification and increased spatial resolution

decreased SID = increased magnification and decreased spatial resolution

43
New cards

formula for maintaining exposure with SID and mAs

mAs1 / mAs2 = (SID1)2 / (SID1)2

44
New cards

what happens with increased OID

decrease in intensity reaching IR

less scatter reaching IR due to beam divergence

potentially increased contrast

increased magnification

reduced spatial resolution

45
New cards

what is the purpose of air gap

reduce amount of scatter radiation reaching IR

46
New cards

how does air gap work

increasing OID (10-15cm) causes scatter to miss receptor thus reducing fog and creating sharper images

47
New cards

what are advantages of air gap

improved image contrast without needing gird

no grid artifacts to affect image quality

improved clarity

48
New cards

what are disadvantages of air gap

increased magnification/distortion due to increased OID

increased exposure due to compensating for OID to have good image quality

49
New cards

what does increasing OID affect

decreased exposure to IR

increased contrast and magnification

decreased spatial resolution

50
New cards

what does decreasing OID affect

increase exposure to IR

decrease contrast and magnification

increase spatial resolution

51
New cards

what does inverse square law state

intensity of x-ray beam is inversely proportional to the square of the distance from the source

52
New cards

what does SID affect in inverse square law

quantity of radiation reaching IR

53
New cards

what is the formula for inverse square law and SID

i1 / i2 = (D2)2 / (D1)2

54
New cards

what is magnification factor

to observe the effect of distance on size distortion

55
New cards

how to calculate Magnification Factor (MF)

SID / SOD

56
New cards

how to calculate source to object distance (SOD)

SID - OID

57
New cards

how to calculate object size

image size - MF

58
New cards

what does MF of 1 mean

no magnification

59
New cards

what can cause shape distortion of anatomic area

inaccurate CR alignment to tube, part, or IR

60
New cards

what does reducing scatter that reaches IR affect

image quality

61
New cards

what does reducing scatted do

increase contrast

62
New cards

when are grids recommended

for body parts thicker than 10 cm

60 kVp or higher

63
New cards

how to grids reduce amount of radiation reaching IR

absorb some transmitted radiation exiting patient

64
New cards

what does a more efficient grid do

create better contrast

65
New cards

what is grid conversion formula

mAs1 / mAs2 = grid ratio1 / grid ratio2

66
New cards

what is the conversion factors for grid ratios

no grid = 1

5:1 = 2

6:1 = 3

8:1 = 4

12:1 = 5

16:1 = 6

67
New cards

what does a larger field size affect

increases amount of tissue irradiated

production of more scatter

increased amount of radiation reaching IR

68
New cards

what does a smaller field size affect

reduced amount of tissue radiated

reduced amount of scatter produced

reduced amount of radiation reaching IR

69
New cards

what do high frequency generators require

lower exposure technique settings

70
New cards

when does tube filtration affect radiation quality

excessive or insufficient filtration

71
New cards

what are different type of body habitus

hypersthenic

sthenic

hyposthenic

asthenic

72
New cards

what does thickness of the body part affect

beam attenuation (increases with thickness)

scatter production

image contrast