Course: BME 312 Biomedical Instrumentation II
Institution: Near East University, Faculty of Engineering, Department of Biomedical Engineering
Subject Matter: Cardiac Pacemakers
Location: Sylmar, CA, USA
Model: Victory SR 5610, Serial Number: 1723182
Components: Cardiac Pacemakers, Electrodes in heart, Pacemaker overview
Sources: SCIENCEphotoLIBRARY
Definition: An electric simulator that produces periodic electric pulses.
Purpose: Conducts stimuli to electrodes in the heart, causing it to contract.
Application: Used during disease states when the heart does not stimulate at the proper rate.
Definition: Free-running pacemakers that emit electric stimulus at a uniform rate regardless of the heart’s condition.
Control Mechanism: The oscillator sets the pulse rate.
Output: Pulse rate regulates the stimulating pulse to the heart via lead wires to cardiac electrodes.
Reliability: Blocks must be highly reliable.
Compatibility: Must be biocompatible and tolerable by body.
Protection: Essential to safeguard circuit components for reliable operation.
Environment: Designed to work in the corrosive environment of the body.
Size: Should occupy minimal volume or mass.
Hermetically sealed metal packages for protection.
Materials: Titanium, Stainless Steel.
Sealing Techniques: Electron beam or laser welding to avoid damage during sealing.
Reliability: Metal packages are more compact and reliable than polymer-based alternatives.
Historical Context: Primary cell batteries used in the 1970s; required replacement every two years.
Current Battery: Lithium Iodide batteries.
Characteristics: Increased lifespan, open-circuit voltage of 2.8 V, high reliability, but has high source resistance limitation.
Mechanism: Implemented via free-running oscillators.
Functionality: Advanced pacemakers utilize timer circuits with complex logic, quartz crystal control, and microprocessor-based circuits.
Purpose: Generates electrical stimulus for the heart.
Types: Produces constant-voltage or constant-current amplitude pulses optimal for myocardium stimulation.
Constant-voltage pulses: Range from 5.0 to 5.5 V with durations of 500-600 µs.
Constant-current pulses: Ranges from 8 to 10 mA, with pulse durations of 1.0 to 2.0 ms.
Typical pulse rate: 70 to 90 beats per minute.
Durability: Must be mechanically strong and able to withstand constant heart motion.
Electrical Insulation: Essential to prevent shunting of stimulating current.
Design: Interwound helical coils made from spring-wire alloy, housed in silicone-rubber or polyurethane.
Features: Coiling minimizes stress, while multiple strands provide redundancy against failure.
Benefits: Encapsulation maintains flexibility, insulation, and biocompatibility.
Unipolar Electrodes: One electrode in contact with the heart with a large indifferent electrode elsewhere.
Bipolar Electrodes: Two electrodes placed in or on the heart.
Epicardial Electrodes: Placed on the heart's surface.
Intramyocardial Electrodes: Buried within the heart wall.
Endocardial Electrodes: Pressed against the inner heart's surface.
Requirements: Materials must not dissolve, irritate heart tissue, or undergo electrolytic reactions.
Material Unity: Same materials as lead wires to avoid junctional problems.
Biological Interaction: Formation of a dense capsule around electrodes reduces interaction and increases stimulation threshold.
Common Materials: Platinum, Platinum-based alloys (with stainless steel, carbon, titanium, etc.).
Structure: Flexible electrodes coiled, providing good contact with heart wall.
Functionality: Stimulates heart wall through conducting bands; secured in place by silicone rubber.
Flexibility: Back support provides mechanical compliance between the electrode and heart wall.
Bipolar Stimulation: Involves pairs of electrodes for effective stimulation.
Components: Includes timing circuit, output circuit, electrodes, reset circuit, and amplifier.
Feedback Loop: Resets after each stimulus, awaiting a natural heartbeat.
Purpose: Amplifies natural beats to reset timer if they occur during the interval.
Components: Controlled by a sensor, controller circuit, pulse generator, control algorithm, lead wire & electrode system.
Sensor Role: Converts physiological variable into an electric signal as input to the controller circuit.
Responsiveness: Allows pacemaker to adjust pacing according to patient’s needs, remaining dormant when natural pacing is functional.
Placement: Can be inside the pacemaker or at another body site, connected by lead wires.
Variables:
Right Ventricle blood temperature (Thermistor)
ECG Signals (Stimulus-to-T wave interval, R-wave area)
Blood pH (Electrochemical pH Electrode)
Right ventricular pressure changes (Semiconductor strain gauge)
Venous blood oxygen saturation (Optical oximeter)
Intracardiac volume changes (Electric-impedance)
Respiratory parameters (Thoracic electric-impedance plethysmography)
Body vibration (Accelerometer)