Post-Treatment Process
Terminate Treatment: Return Arterial Blood
Steps to return Arterial blood:
1 Prepare supplies & apply PPE
2 Reduce pump speed
3 Stop blood pump
4 Move the plastic hemostat from saline "T" to above saline "T"
5 Open saline line clamps
6 Return blood through arterial line - wait until light pink
7 Close clamps
Terminate Treatment: Return Venous Blood
Steps to return Venous blood:
1 Move the plastic hemostat from above saline "T" to below saline "T"
2 Start blood pump and increase speed to 150-200 mL/min
3 Observe patient and machine carefully
4 Pinch and release blood tubing between blood pump and dialyzer until blood in venous bloodline and fistula needle changes to light pink
5 Reduce blood pump speed to minimum then stop pump
6 Clamp venous bloodline clamp