Objectives
discuss development of fluoroscopy
differentiate between fluoroscopic and radiographic imaging
Describe the components of an image intensifier
Explain the operation of an image intensifier
Explain the operation of an image intensifier in magnification mode and its effect on image quality and patient exposure
Contect Spects
exposure factors
fluoroscopy time
Automatic Brightness Control
Receptor Positioning
Magnification Mode
Air Kerma Display
Dose or Time Documentation
Minimum Source-to-Skin Distance
History of Fluoroscopy
invented bu Thomas C. Edison
1st fluoroscope available commercially in 1896
Produced a faint image
Dark Adapt: operator would sit in a darkened room OR wear adaption googles with red lenses
Dynamic radiographic examination
views while standing in the path of the x-ray beam as it exits the patient
high dose to the operator
1950’s: Image intensifier
Electronic vacuum tube
brightened image
eliminated the need to dark-adapt
improved visible anatomic details
indirect viewing
low mA (.5- 5 mA)
Round or Oval FOV
Spatial resolution: 1-2 Lp/mm
Converts:
remnant beam → light
light → electrons
electrons → light
Input Phosphor
absorbs x-ray photons and emits light photons
Cesium Iodide (CsI)
scintillating phosphor
emits light
Closest to patient
larger phosphor
15 - 30 cm
Photocathode ( photo → light, cathode → source of electrons)
emits electrons in response to light stimulus
photoemission
need many light photons to create1 electron
Cesium and antimony
Curved surface to ensure that all electrons emitted travel same distance to output phosphor
makes sure all electrons make it towards cathode
Electrostatic Focusing Lenses
Negative charged plates
repels electrons and focuses them toward the output phosphor
electrostatic repulsion
Focal Point
where electrons merge into a point
Accelerating Anode
constant positive charge of 25 kV
Attracts electrons toward the output phosphor
Output Phosphor
absorbs electrons and emits light
1 electron emits many light photons
Silver-activated zinc cadmium sulfide
increase light brightness compared to brightness at input phosphor
brightness gain
smaller phosphor
2,5 cm
Image Intensification
Input Phosphor absorbs the remnant beam and emits light in response
emitted light exposes the PHOTOCATHODE which emits electrons
Brightness Gain
ability of the image intensifier to convert x-ray to light and increase the brightness of the image
Brightness Gain = Flux Gain x Minification Gain
Higher Brightness Gain = Greater Efficiency of the Image Intensifier
Image Intensifiers have. brightness Gains of 50000 - 200000
As the image intensifier ages it loses its efficient to convert light
more radiation is needed to produce the same level of brightness
Flux Gain
conversion of efficiency of the I.I.
Light photons ‘in’ vs light photons ‘out’
Flux gain = # of output light photons / # of input of x-ray photons
Minification Gain
Expresses the degree an image is made smaller from input to output phosphor
diameter of inout phosphor vs diameter of output phosphor
brighter image output
electrons concentrated into smaller surface area
minification gain = ( d1/d2)²
Automatic Brightness Control
ABC
maintains the overall appearance of intensified image
contrast and brightness
automatically adjusts the kVp and mA
operator can select a desired brightness level at the control panel
can be slow to respond to change sin patient tissue thickness and density as it moves
brightness lag
Magnification Mode
“tri-focus” image intensifier
adjustable inout phosphor size
various sizes
most common: 25/17/12 cm
increases patient dose
Camera Tube
attached to image intensifier
Converts light signal from output phosphor to an electronic signal
Electron Gun: heated filament that supplies a constant electron current
Electrostatic Grids: Accelerates and focuses the electron bea,
control grid
Outside of tube:
electromagnet coils:
controls size and position of electron beam
Target assembly: 3 layers
window: outside layer
signal plate: conducts video signal
target plate: photoconductive layer; conducts electrons when illuminated
Spot Film
‘Static’ imaging - one time frame
receptor located between patient and image intensifier
receptor i sparked out of beam path
stored in a lead lined shroud until needed
until will shift from fluoroscopy mode to radiographic mode
Milliampaerage
Fluoroscopic mode:
low mA
2-5 mA
Radiographic Mode:
High mA
100-1200 mA
Photospot Camera
allows radiologist to view image while it is being recorded
Photospot v Spot film
photospot: image intensifier
Spot film: image receptor
Photospot: fluor FOV
Spot Film: Image Receptor FOV
Cineradiography
Dynamic recording
PinCushion Effect
curved edges of image
distortion caused by unequal magnification
Mottle
insufficient information present to create image
increase mA to compensate
Air Kerma
kinetic energy released per unit mass in air
should not normally exceed 5 cGy/min at the position wher ethe beam enters the patient
must not exceed 10 cGy/min
air kerma from leakage radiation. measured at 1 m from the source, does not exceed 0.1 rad in any 1 hour
Dose Display
updated every 6 sec
irradiation time in minutes and tenths of minutes shall be continuously displayed
flouro and radiation dose
Personnel Protection
only persons whose presence is necessary shall be in the fluoroscopy room during exposures
Source-to Skin Distance
SSD
must not be less than 30 cm
should not be less than 38 cm (15 in.)
Bucky Slot Cover
protective device
automatically covers the bucky opening durng exposure
at least 0.25 mm Pb equivalent
High Level Control
requires positive action to start and maintain the high exposure rate
continuous audible signal shall accompany the high rate