Vascular Physics Overview
Focus on image optimization and direction of blood flow.
Doppler effect relevant to ultrasound and real-life scenarios (e.g., weather, sound).
Doppler Effect
Describes the change in frequency or wavelength of a wave in relation to an observer moving relative to the wave source.
Important in ultrasound for assessing blood flow velocity.
Sound Wave Generation
Sound waves produced by piezoelectric crystals in the ultrasound transducer.
Electric stimulation causes the piezoelectric elements to vibrate, producing sound waves.
Frequency based on the size of piezoelectric elements: smaller = higher frequency, larger = lower frequency.
Standard ultrasound frequencies: 2 to 17 MHz, up to 18 MHz for detailed imaging.
Doppler Shift
Occurs due to the difference between transmitted frequency and received frequency from moving blood cells.
Positive Doppler shift: moving towards transducer (higher returning frequency).
Negative Doppler shift: moving away from transducer (lower returning frequency).
Null: no shift when frequencies are equal.
Correct angling of the vessel important for accurate measurements (ideally below 60 degrees).
Spectral Display
Doppler shifts displayed; negative below baseline, positive above.
Sample volume must be correctly placed to ensure accurate data.
Spectral inversion may occur; check settings for proper display of flow direction.
Beam Angle and Frequency Shift
0-degree angle: maximum frequency shift and accuracy in measuring blood flow velocity.
90-degree angle: minimal frequency shift; often results in missed flow detection (black area on display).
Beam steering can adjust the angle of the transducer to optimize blood flow assessment.
Hemodynamics Factors
Three main factors affecting resistance to blood flow:
Diameter/radius of the vessel.
Length of the vessel.
Viscosity (thickness) of the blood.
Poiseuille's law relates these properties to blood flow resistance.
Doppler Imaging Techniques
Continuous Wave Doppler: utilizes two piezoelectric elements. Advantages: no aliasing, detects flow across multiple vessels. Disadvantages: does not isolate specific vessels.
Pulsed Wave Doppler: sends pulses and receives echoes over time, allowing for depth-specific analysis.
High pulse repetition frequency (PRF) is essential to avoid aliasing; relates to Nyquist limit (frequency exceeding half the PRF leads to aliasing).
Artifacts in Doppler Imaging
Mirror image artifact: occurs when sound reflects off strong reflectors, misrepresenting vessel locations.
Elevation plane focus artifacts: averaging surrounding tissue with the object of interest, leading to misinterpretation.
Intrinsic spectral broadening: can be caused by improper sample volume size leading to incorrect velocity readings.
Image Optimization Techniques
Ensure TGC and gain settings are adjusted for optimal image clarity.
Maintain perpendicular beam placement to anatomical structures for best projection.
Use appropriate transducer frequency for the area being examined.
Utilize auto-optimization features of modern machines where applicable.
Minimize speckle artifacts through proper adjustments in imaging settings.
Adjust wall filters to avoid missing critical low-velocity measurements.
Panoramic imaging is useful for large anatomical areas or extensive vessels.
Conclusion
Continuous reevaluation and adjustment of imaging parameters are necessary for accurate ultrasound diagnostics.
Technology and collaboration among sonographers can lead to improved imaging techniques and understanding.