MRI Gradients, Coils, and Quality Assurance – Lecture Vocabulary

Gradient Fundamentals

  • Gradient = any magnetic component that varies over space (and/or time) and is super-imposed on the static main field (B_0).
  • MRI system contains three orthogonal, independently-driven gradient sets:
    • Z (longitudinal/head–feet)
    • Y (anterior–posterior)
    • X (left–right)
  • Purpose: locally increase or decrease B_0 so that frequency (or phase) of proton precession becomes position-dependent → essential for slice selection, phase encoding, frequency/readout encoding.
  • Gradients are separate from: main superconducting coils, passive/active shielding coils, and shim coils.

Gradient Coil Designs

  • Z-gradient: Helmholtz pair – two identical circular windings mounted at each bore end; produces symmetric field that rises toward each coil.
  • X & Y gradients: “Pear-saddle” coils – saddle-shaped windings creating transverse linear fields.
  • All gradient sets are resistive; field exists only while current is applied (unlike persistent superconductive main coils).
  • Gradient shield windings placed outside primary gradient conductors confine fringe gradient field inside bore → reduces eddy currents in cryostat & limits external RF noise.

Gradient Field Mathematics & Example

  • Example: B0 = 3.0\,\text{T}. A z-gradient generates a linear variation Gz = \pm1.5\,\text{T} across bore:
    • At + side: B{tot}=3+1.5=4.5\;\text{T}, 3+1=4.0\;\text{T}, 3+0.5=3.5\;\text{T} • At isocenter (0 point): B{tot}=3.0\;\text{T}
    • At − side: 3-0.5=2.5\;\text{T} … 3-1.5=1.5\;\text{T}
  • Reversing gradient current flips polarity; same magnitude, opposite sign.
  • “Linear” property: equal distance left/right of isocenter gives equal magnitude but opposite sign change (e.g., \Delta B = \pm1.0\,\text{T} at ±10 cm).

Gradient System Performance Parameters

  • Gradient Amplitude: peak strength per unit distance
    G_{max} in \text{mT/m} or \text{G/cm} (10 \text{mT/m}=1\,\text{G/cm}).
  • Rise Time: time from 0 to G_{max} (ms).
  • Slew Rate: S = G_{max}/\text{rise time} → units \text{mT/m/ms} or \text{T/m/s}; combines power & speed; major spec quoted by vendors (e.g., 120\,\text{mT/m/ms}).
  • Duty Cycle: % of each TR that gradient is energized.
  • Balanced Gradient: second gradient of equal magnitude & opposite polarity applied to re-phase spins and cancel phase dispersion.

Biological & Safety Considerations

  • Rapid switching induces electric fields in conductive tissue → peripheral nerve stimulation, muscle twitch, phosphenes (optic nerve), potential cardiac arrhythmia.
  • Duty-cycle & slew-rate limits in software avert PNS and ventricular fibrillation.
  • Acoustic noise: gradient vibration against cryostat; up to >100 dB; louder on high-field/higher-slew systems.

Pulse Sequence Context

  • Typical spin-echo: 90° RF → TE/2 → 180° RF → echo; gradient “blocks” timed for
    • Slice-select,
    • Phase-encode,
    • Frequency(read)-encode.
  • Four separate gradient events can be toggled within 20\,\text{ms}, causing audible “bang–bang”.

Hardware Requirements Synergy

  • High B_0 alone is insufficient → must be paired with high-slew, high-amplitude gradients AND high-performance RF coils to obtain speed, resolution, SNR.

RF Coil Taxonomy

  • By function:
    • Transmit (Tx)
    • Receive (Rx)
    • Transceiver (Tx/Rx)
  • By geometry:
    • Volume (encircle anatomy)
    • Surface (flat/contoured, one side of anatomy)
    • Intracavitary (endorectal, endovaginal, etc.)
  • By polarization:
    • Linear (single loop)
    • Quadrature / Birdcage (two orthogonal loops; circularly polarized)
  • By channel count:
    • Single-channel
    • Phased-array (multi-element, independently amplified)

Body vs Local Coils

  • Body coil (built-in birdcage) usually acts as Tx; distance → poor SNR for Rx.
  • Local coil placed directly on anatomy yields SNR increase ≈ distance² law (radio metaphor: farther station → noise).
  • Tx/Rx local coils (if available) lower \text{SAR} because RF is confined to region instead of whole torso.

Polarization & Quadrature Advantage

  • Quadrature coil delivers a 90° flip using 50 % less power vs. linear → \downarrow SAR.
  • Provides ≈40 % SNR improvement (two orthogonal receive channels combined in quadrature).

Phased-Array & Parallel Imaging

  • Phased-array = single housing with many decoupled Rx loops.
  • Enables:
    • Larger FOV without noise penalty (activate only sections under interest).
    • Parallel imaging (SENSE, ASSET, iPAT) – under-sample k-space in multiple channels → scan-time cut ~40-60 %; cost = lower SNR by factor \sqrt{R} + g-factor.
  • Requires:
    • Independent pre-amps,
    • Coil decoupling & tuning each exam.

Coil Selection Best Practices

  • Choose coil that is:
    1. Smallest footprint completely covering anatomy.
    2. Physically closest to tissue (maximize filling factor).
    3. Highest channel count (more SNR/parallel options).
    4. Quadrature or multi-channel rather than linear.
    5. Tx/Rx if SAR or high-field constraints present.
  • Example: neonatal brain often fits knee volume coil (smaller diameter, proximity) → may outperform 16-ch adult head coil; must empirically compare.
  • Surface coil penetration ≈ radius of loop; use volume coil for deep organs (liver, pancreas).

Large vs Small Coil Trade-offs

  • Large coil: uniform field, relaxed positioning, but ↓SNR, ↑risk of aliasing (wrap) when FOV < coil coverage.
  • Small coil: sharp drop-off outside sensitive region → high SNR, low aliasing, but precise placement critical.

RF Shielding & Zipper Artifact

  • Entire room enclosed in Faraday cage (copper/aluminum walls, floor, ceiling, penetrations, window with copper mesh).
  • Stops external RF (radio, TV, cell) in 10–300 MHz band used by MRI (same band as Larmor frequencies).
  • Breach → zipper artifact (vertical line noise) on image.
    • Causes: door opened during scan, broken light bulb filament, faulty equipment, physical tear in shield.
  • All conduits enter via filtered penetration panel.

Patient Tables & Workflow

  • Fixed table (attached) vs detachable dock-able table.
  • Dock-able benefits:
    • Rapid patient turnover (pre-load outside),
    • Emergency egress without ferromagnetic stretcher,
    • Redundancy (second top).
  • Cost/benefit analysis depends on daily volume (e.g., saving 2 h/day ~ 4–6 extra cases worth ).

Control & Computer Rooms

  • Control room: technologist console, ECG/patient monitoring, communication, observation through RF window (mesh embedded glass).
  • Computer/equipment room: pulse-sequence controller, gradient amplifiers, RF amplifiers & receivers, digitizers, cryogen & environmental monitors.

Quality Assurance (QA) & Quality Control (QC)

  • Personnel requirements:
    • MRI technologist ARRT or ARMRIT certified.
    • Board-certified radiologist.
    • Qualified MR physicist.
  • Accreditation bodies: ACR, Joint Commission, RadSite; renewal every 3 yr; include phantom & clinical submissions.

Daily QA

  • Quick phantom scan (~10 min) to log: SNR, center frequency, Tx gain, geometric accuracy.
  • Visual checks: coil integrity, table motion, room temp/humidity, cryogen level.

Weekly (QC) – ACR Protocol

  • Visual tests of patient transport, laser alignment, gradient/door interlocks, viewing monitors.
  • ACR 7 quantitative tests (done with ACR head phantom):
    1. Table position accuracy (\pm2\,\text{mm}).
    2. Center frequency (e.g., 63.8\,\text{MHz} for 1.5 T).
    3. Transmitter gain (attenuation calibration).
    4. Geometric accuracy (Slice 5 must read 190\pm2\,\text{mm} in both axes).
    5. High-contrast resolution (resolve 0.9–1.1 mm holes).
    6. Low-contrast resolution (≥ 9 spokes visible).
    7. Artifact evaluation (no zipper, ghost, distortion).

ACR Phantom Testing Steps

  1. Level phantom in head coil; align lasers to central fiducial; acquire sagittal localizer.
  2. Acquire 11 axial slices per ACR protocol.
  3. Measurements:
    • Slice 1: high-contrast holes.
    • Side localizer: table offset \le\pm2\,\text{mm}.
    • Slice 5: geometric calipers top–bottom & left–right =190\pm2\,\text{mm}.
    • Slice N (contrast cylinders): count for low-contrast index (need ≥9).
    • Spectrum plot: verify center frequency & fat–water separation (e.g., 220\,\text{Hz} at 1.5 T).
  4. Record in ACR QC form; call service if any parameter outside tolerance.

Key Equations & Numerical References

  • Larmor (precession) frequency: \omega = \gamma B0 with \gamma{\text{H}}\approx42.58\,\text{MHz/T}.
    • B0=1.5\,\text{T}\Rightarrow f\approx63.8\,\text{MHz}. • B0=3.0\,\text{T}\Rightarrow f\approx127.7\,\text{MHz}.
  • Chemical shift (fat–water) separation: \Delta f = 3.5\,\text{ppm}\times B_0 → \approx220\,\text{Hz} @1.5 T.
  • Gradient amplitude–unit conversion: 10\,\text{mT/m}=1\,\text{G/cm}$$.

Practical/Clinical Implications & Ethics

  • Coil mis-selection lowers diagnostic accuracy; knowledge prevents repeat exams, lowers cost & patient burden.
  • Ignoring duty-cycle / slew-rate limits risks patient PNS or arrhythmia → ethical obligation to respect manufacturer limits.
  • Regular QA/QC ensures image reliability, avoids misdiagnosis and costly downtime – technologist’s professional duty.