MRI Gradients, Coils, and Quality Assurance – Lecture Vocabulary
Gradient Fundamentals
- Gradient = any magnetic component that varies over space (and/or time) and is super-imposed on the static main field (B_0).
- MRI system contains three orthogonal, independently-driven gradient sets:
• Z (longitudinal/head–feet)
• Y (anterior–posterior)
• X (left–right) - Purpose: locally increase or decrease B_0 so that frequency (or phase) of proton precession becomes position-dependent → essential for slice selection, phase encoding, frequency/readout encoding.
- Gradients are separate from: main superconducting coils, passive/active shielding coils, and shim coils.
Gradient Coil Designs
- Z-gradient: Helmholtz pair – two identical circular windings mounted at each bore end; produces symmetric field that rises toward each coil.
- X & Y gradients: “Pear-saddle” coils – saddle-shaped windings creating transverse linear fields.
- All gradient sets are resistive; field exists only while current is applied (unlike persistent superconductive main coils).
- Gradient shield windings placed outside primary gradient conductors confine fringe gradient field inside bore → reduces eddy currents in cryostat & limits external RF noise.
Gradient Field Mathematics & Example
- Example: B0 = 3.0\,\text{T}. A z-gradient generates a linear variation Gz = \pm1.5\,\text{T} across bore:
• At + side: B{tot}=3+1.5=4.5\;\text{T}, 3+1=4.0\;\text{T}, 3+0.5=3.5\;\text{T}
• At isocenter (0 point): B{tot}=3.0\;\text{T}
• At − side: 3-0.5=2.5\;\text{T} … 3-1.5=1.5\;\text{T} - Reversing gradient current flips polarity; same magnitude, opposite sign.
- “Linear” property: equal distance left/right of isocenter gives equal magnitude but opposite sign change (e.g., \Delta B = \pm1.0\,\text{T} at ±10 cm).
- Gradient Amplitude: peak strength per unit distance
G_{max} in \text{mT/m} or \text{G/cm} (10 \text{mT/m}=1\,\text{G/cm}). - Rise Time: time from 0 to G_{max} (ms).
- Slew Rate: S = G_{max}/\text{rise time} → units \text{mT/m/ms} or \text{T/m/s}; combines power & speed; major spec quoted by vendors (e.g., 120\,\text{mT/m/ms}).
- Duty Cycle: % of each TR that gradient is energized.
- Balanced Gradient: second gradient of equal magnitude & opposite polarity applied to re-phase spins and cancel phase dispersion.
Biological & Safety Considerations
- Rapid switching induces electric fields in conductive tissue → peripheral nerve stimulation, muscle twitch, phosphenes (optic nerve), potential cardiac arrhythmia.
- Duty-cycle & slew-rate limits in software avert PNS and ventricular fibrillation.
- Acoustic noise: gradient vibration against cryostat; up to >100 dB; louder on high-field/higher-slew systems.
Pulse Sequence Context
- Typical spin-echo: 90° RF → TE/2 → 180° RF → echo; gradient “blocks” timed for
• Slice-select,
• Phase-encode,
• Frequency(read)-encode. - Four separate gradient events can be toggled within 20\,\text{ms}, causing audible “bang–bang”.
Hardware Requirements Synergy
- High B_0 alone is insufficient → must be paired with high-slew, high-amplitude gradients AND high-performance RF coils to obtain speed, resolution, SNR.
RF Coil Taxonomy
- By function:
• Transmit (Tx)
• Receive (Rx)
• Transceiver (Tx/Rx) - By geometry:
• Volume (encircle anatomy)
• Surface (flat/contoured, one side of anatomy)
• Intracavitary (endorectal, endovaginal, etc.) - By polarization:
• Linear (single loop)
• Quadrature / Birdcage (two orthogonal loops; circularly polarized) - By channel count:
• Single-channel
• Phased-array (multi-element, independently amplified)
Body vs Local Coils
- Body coil (built-in birdcage) usually acts as Tx; distance → poor SNR for Rx.
- Local coil placed directly on anatomy yields SNR increase ≈ distance² law (radio metaphor: farther station → noise).
- Tx/Rx local coils (if available) lower \text{SAR} because RF is confined to region instead of whole torso.
Polarization & Quadrature Advantage
- Quadrature coil delivers a 90° flip using 50 % less power vs. linear → \downarrow SAR.
- Provides ≈40 % SNR improvement (two orthogonal receive channels combined in quadrature).
Phased-Array & Parallel Imaging
- Phased-array = single housing with many decoupled Rx loops.
- Enables:
• Larger FOV without noise penalty (activate only sections under interest).
• Parallel imaging (SENSE, ASSET, iPAT) – under-sample k-space in multiple channels → scan-time cut ~40-60 %; cost = lower SNR by factor \sqrt{R} + g-factor. - Requires:
• Independent pre-amps,
• Coil decoupling & tuning each exam.
Coil Selection Best Practices
- Choose coil that is:
- Smallest footprint completely covering anatomy.
- Physically closest to tissue (maximize filling factor).
- Highest channel count (more SNR/parallel options).
- Quadrature or multi-channel rather than linear.
- Tx/Rx if SAR or high-field constraints present.
- Example: neonatal brain often fits knee volume coil (smaller diameter, proximity) → may outperform 16-ch adult head coil; must empirically compare.
- Surface coil penetration ≈ radius of loop; use volume coil for deep organs (liver, pancreas).
Large vs Small Coil Trade-offs
- Large coil: uniform field, relaxed positioning, but ↓SNR, ↑risk of aliasing (wrap) when FOV < coil coverage.
- Small coil: sharp drop-off outside sensitive region → high SNR, low aliasing, but precise placement critical.
RF Shielding & Zipper Artifact
- Entire room enclosed in Faraday cage (copper/aluminum walls, floor, ceiling, penetrations, window with copper mesh).
- Stops external RF (radio, TV, cell) in 10–300 MHz band used by MRI (same band as Larmor frequencies).
- Breach → zipper artifact (vertical line noise) on image.
• Causes: door opened during scan, broken light bulb filament, faulty equipment, physical tear in shield. - All conduits enter via filtered penetration panel.
Patient Tables & Workflow
- Fixed table (attached) vs detachable dock-able table.
- Dock-able benefits:
• Rapid patient turnover (pre-load outside),
• Emergency egress without ferromagnetic stretcher,
• Redundancy (second top). - Cost/benefit analysis depends on daily volume (e.g., saving 2 h/day ~ 4–6 extra cases worth ).
Control & Computer Rooms
- Control room: technologist console, ECG/patient monitoring, communication, observation through RF window (mesh embedded glass).
- Computer/equipment room: pulse-sequence controller, gradient amplifiers, RF amplifiers & receivers, digitizers, cryogen & environmental monitors.
Quality Assurance (QA) & Quality Control (QC)
- Personnel requirements:
• MRI technologist ARRT or ARMRIT certified.
• Board-certified radiologist.
• Qualified MR physicist. - Accreditation bodies: ACR, Joint Commission, RadSite; renewal every 3 yr; include phantom & clinical submissions.
Daily QA
- Quick phantom scan (~10 min) to log: SNR, center frequency, Tx gain, geometric accuracy.
- Visual checks: coil integrity, table motion, room temp/humidity, cryogen level.
Weekly (QC) – ACR Protocol
- Visual tests of patient transport, laser alignment, gradient/door interlocks, viewing monitors.
- ACR 7 quantitative tests (done with ACR head phantom):
- Table position accuracy (\pm2\,\text{mm}).
- Center frequency (e.g., 63.8\,\text{MHz} for 1.5 T).
- Transmitter gain (attenuation calibration).
- Geometric accuracy (Slice 5 must read 190\pm2\,\text{mm} in both axes).
- High-contrast resolution (resolve 0.9–1.1 mm holes).
- Low-contrast resolution (≥ 9 spokes visible).
- Artifact evaluation (no zipper, ghost, distortion).
ACR Phantom Testing Steps
- Level phantom in head coil; align lasers to central fiducial; acquire sagittal localizer.
- Acquire 11 axial slices per ACR protocol.
- Measurements:
• Slice 1: high-contrast holes.
• Side localizer: table offset \le\pm2\,\text{mm}.
• Slice 5: geometric calipers top–bottom & left–right =190\pm2\,\text{mm}.
• Slice N (contrast cylinders): count for low-contrast index (need ≥9).
• Spectrum plot: verify center frequency & fat–water separation (e.g., 220\,\text{Hz} at 1.5 T). - Record in ACR QC form; call service if any parameter outside tolerance.
Key Equations & Numerical References
- Larmor (precession) frequency: \omega = \gamma B0 with \gamma{\text{H}}\approx42.58\,\text{MHz/T}.
• B0=1.5\,\text{T}\Rightarrow f\approx63.8\,\text{MHz}.
• B0=3.0\,\text{T}\Rightarrow f\approx127.7\,\text{MHz}. - Chemical shift (fat–water) separation: \Delta f = 3.5\,\text{ppm}\times B_0 → \approx220\,\text{Hz} @1.5 T.
- Gradient amplitude–unit conversion: 10\,\text{mT/m}=1\,\text{G/cm}$$.
Practical/Clinical Implications & Ethics
- Coil mis-selection lowers diagnostic accuracy; knowledge prevents repeat exams, lowers cost & patient burden.
- Ignoring duty-cycle / slew-rate limits risks patient PNS or arrhythmia → ethical obligation to respect manufacturer limits.
- Regular QA/QC ensures image reliability, avoids misdiagnosis and costly downtime – technologist’s professional duty.