CT Acquisition and Parameters
Projections
Definition: A view from the anatomical cross-section along the path of the x-ray beam from only one vantage point is defined as a projection.
CT Scan Formation: A CT scan is created from numerous projections in order to provide a comprehensive view of the object.
Perspective Requirement: To ensure accurate representation of an object's structure, it is essential to observe it from all angles.
Topogram (Localizer) Scanning
Functionality: Provides a single projection and does not create a cross-sectional image.
Purpose: Used to position the Field of View (FOV) where cross-sectional CT and PET images will be captured.
Operation: Utilizes a stationary x-ray tube paired with a moving table.
Topogram Lengths for the Biograph Systems
Sequence Scanning
Description: A cross-sectional imaging method that allows for retrospective reconstruction.
Movement: Patient table advances incrementally after each x-ray tube rotation.
Rotation Details: The tube and detector assembly rotate 360° in one direction and back for each slice increment.
I.S.D: Inter Scan Delay must be noted.
Scanning Methods Details
Standard Sequence Scanning
Characteristics: Longer I.S.D. but instant data display.
Dynamic Sequence Scanning
Characteristics: Faster acquisition with delayed image display.
Downfalls of Sequence Scanning
Issues Identified:
Respiratory-induced slice misregistration.
Data reconstruction is limited to scanned areas; no retrospective slice overlap is available.
Potentially long inter-scan delays.
Higher volumes of contrast material may be necessary.
Single Slice vs. Multi-Slice System
Single Slice System: Only one slice per rotation can be acquired.
Multi-Slice System: Multiple slices can be acquired per rotation.
Spiral (Helical) Scanning
Structure: The x-ray beam follows a spiral or “corkscrew” path around the patient.
Path Determination: The trajectory of the x-ray beam is based on:
Thickness of the x-ray beam, correlated with pre-patient collimation and slice thickness.
The speed of the patient table movement.
Construction: X-ray tube and detectors rotate around the patient simultaneous with table movement, yielding complete volume data acquisition.
The 4 “C”s of Spiral CT Imaging
Continuous rotating tube/detector system.
Continuous radiation.
Continuous data acquisition.
Continuous table feed.
Advantages of Spiral Scanning
Benefits:
Faster acquisition times.
More anatomical coverage with each breath-hold, enhancing detection of small lesions.
Reduced contrast material is required.
Arbitrary slice positioning is possible, ideal for various post-processing techniques.
Preparation of Spiral Images for Reconstruction
Data Characteristics: Raw data collected does not correspond to specific slice positions since it is acquired in a “corkscrew” format. Mathematical operations are essential for data reconstruction, deriving attenuation information from planar single slice designs.
Factors that Affect CT Image Quality
Key Parameters:
mAs
kVp
Slice Thickness
Pitch
Isocentering
mAs (Milliampere-Seconds)
Components:
mA (tube current): Controls the photon quantity generated by the x-ray tube.
s (seconds): Rotation time, denoting the interval needed for a full 360° rotation.
Impact on Scan: Affects total scan time and coverage along the Z-axis.
mAs Definition and Effects
Definition: mAs is the product of tube current (mA) and rotation time (s). It dictates the amount of radiation the patient is subjected to.
Visual: Conceptualize mAs as a rainstorm of photons, where mA is the intensity of the rain, and s is the pace of movement through the downpour.
Effects of Increasing mAs:
Decreases image noise.
Enhances image detail.
Clinical Scenarios:
For soft tissue diagnostics, higher mAs is required due to closely similar CT values between tissues.
For structures like bones or lungs, lower mAs is appropriate since CT values differ significantly.
kVp (Kilovoltage Peak)
Definition: Voltage applied to the x-ray anode and cathode, necessary for facilitating tube current flow.
Energy Interpretation: Expressed in kilo-electron volts (keV), representing the maximal energy of x-ray photons.
Impact on Image: Controls image contrast—higher kVp enhances penetration while lower kVp reduces it.
Slice Thickness
Concept: Measure of millimeters of anatomical detail covered by the x-ray beam; affects image properties directly.
Trade-offs: Increasing slice thickness provides greater anatomical coverage but may reduce resolution and increase partial volume effects. Slice thickness represents a balance between edge definition and noise levels.
Table Increment (Sequence Mode)
Definition: The millimeter advancement of the patient table between successive sequence slices dictates the overlap and gaps.
Visualization Examples:
Increment equal to slice thickness = no gaps.
Increment less than the slice = overlap.
Increment greater than slice thickness = gaps.
Overlap Benefits: Enhances fine detail visibility but increases radiation dosage.
Pitch
Definition: Describes the stretch of x-ray data during a spiral scan.
Calculation for Multi-Slice Systems:
ext{Pitch} = rac{ ext{Table Feed}}{ ext{Rotation} imes ext{Collimator Size} imes ext{# of Collimators}}Changes due to Pitch Adjustment: Increasing pitch beyond 2 can lead to anatomical skip but allows faster scanning, minimizing patient doses and expediting emergency procedures.
Isocentering
Definition: Proper positioning to ensure the x-ray beam accurately displays anatomy across all tube positions.
Importance of Isocentering: Non-centering can lead to image artifacts, compromising interpretation accuracy.
Review Questions
Scannings Setting: The scanning mode used to set the field of view is:
1. Sequence
2. Spiral
3. Topogram
4. Single slice
Downfalls of Sequence Scanning: Which are downfalls?
I. Larger amounts of contrast required.
II. Respiratory induced mis-registration.
III. No inter-scan delay.
1. I and II
2. II and III
3. I and III
4. I, II and III
Spiral Scanning Characteristics: Which are true?
I. Longer acquisition times compared to sequence scanning.
II. Tube/detector system rotates around the patient in a cork-screw structure.
III. Complete volume of data is acquired.
1. I and II
2. II and III
3. I and III
4. I, II and III
Factors Affecting Photon Exposure: Which parameter affects photon exposure?
1. kVp
2. Pitch
3. mAs
Max Energy of Photons: What is the maximum energy called?
1. mAs
2. Pitch
3. kVp
Reasons to Increase Pitch: Reasons include:
I. Greater anatomical coverage in less time.
II. Faster scanning for trauma patients.
III. Increase patient dose.
a. I only
b. I and II
c. II and III
d. I, II and III
Final Note: All images are credited to Siemens Training and Development Center as per the transcript.