CT safety
Patient Care 1. Define Ionicity: Ionic contrast agents are those that dissociate into charged particles in the presence of water and blood. Known to be more toxic to patients, in part due to its increased osmolarity. Most common contrast agents are non-ionic: do not form charged particles in the blood.
Patient Care 2. Define osmolarity: Measure of the osmotic pressure of a solution concerning the number of particles per volume of solution. Blood and human serum have an osmolarity of approximately 300 mOsm/L.
Patient Care 3. Are most IV contrast agents ionic or nonionic? Why? Most common contrast agents are non-ionic: do not form charged particles in the blood. Ionic contrast agents are known to be more toxic to patients, in part due to its increased osmolarity.
Patient Care 4. Why is water-soluble iodine the only contrast agent acceptable for IV injection? Water-soluble iodine is chemically bound to a benzene ring to make it biologically compatible with the body. It is the only contrast agent suitable for IV injection. The ideal water-soluble contrast agent is both non-ionic and low-osmolar.
Patient Care 5. How is barium used in CT different than the barium used for fluoroscopy exams? CT Barium is not the same as barium used in fluoroscopy. Barium sulfate has a high atomic number (Z=56), is a suspension (suspended in water, not dissolved), and is administered orally or rectally. The PowerPoint does not provide further differences.
Patient Care 6. What are the dietary restrictions for administering IV contrast? Pre-hydration; NPO 1 hour; post-hydration.
Patient Care 7. State the ânormal rangeâ for each of the following laboratory tests: a. BUN; b. Creatinine; c. GFR: BUN: 7â21 mg/dL. Creatinine: 0.7â1.5 mg/dL. GFR: 90â120 mL/min.
Patient Care 8. State the lab value that may indicate renal failure for the following laboratory tests: a. BUN; b. Creatinine; c. GFR: BUN: Above 50 may indicate renal failure. Creatinine: Above 2.0 may indicate renal failure. GFR: Below 30 may indicate renal failure.
Patient Care 9. What are the common medications used to premedicate a patient designed to decrease the chance of a contrast reaction? For known iodine allergy, two types are generally used: corticosteroidâprednisone; anti-histamineâdiphenhydramine (Benadryl).
Patient Care 10. Define cannulation: Process of placing a peripheral IV.
Patient Care 11. State the preferred sites for placing an IV in CT: Cephalic vein, Basilic vein, Median cubital vein.
Patient Care 12. Why should the small veins of the hands and wrist be avoided for IV contrast injection? NOT PROVIDED IN POWERPOINT. The PowerPoint identifies the preferred sites in CT as the cephalic vein, basilic vein, and median cubital vein, but does not explain why small veins of the hands and wrist should be avoided.
Patient Care 13. During IV cannulations, when should hand hygiene be performed? First step of aseptic technique: wash hands, then wear gloves, then clean the skin.
Patient Care 14. What kind of gloves are required when starting an IV? (choose one) The PowerPoint states to wear gloves but does not specify clean or sterile gloves.
Patient Care 15. Under what circumstances can the technologist use the patientâs venous catheter for pressure injections? Peripheral IV: Catheters of 20 gauge or higher in the antecubital or forearm are preferred for pressure injecting contrast in adults. PICC lines and central lines: Must be labeled for pressure injection; maximum injection rate is provided.
Patient Care 16. What type of venous access devices are not acceptable for pressure injections? Jugular lines: Not acceptable for pressure injection; extravasation in the neck too risky. Dialysis catheters: Not acceptable for pressure injection.
Patient Care 17. Define extravasation. The leakage of contrast (or any fluid) outside the blood vessel.
Patient Care 18. What are the initial signs and symptoms of extravasation? NOT PROVIDED IN POWERPOINT. The slide includes the heading âInitial Signs and symptomsâ but does not list them.
Patient Care 19. How is extravasation treated? NOT PROVIDED IN POWERPOINT. The slide includes the heading âTreatmentâ but does not provide treatment steps.
Patient Care 20. Describe the treatment of an air embolism. Administration of 100% oxygen; placing patient in left lateral decubitus. Air embolism is a potentially fatal but rare complication of IV contrast media injection.
Patient Care 21. List the mild contrast reactions: Urticaria (hives); Pruritus (itching); Rhinorrhea (runny nose); Nausea and Vomiting; Coughing; Dizziness.
Patient Care 22. List the moderate contrast reactions: Persistent Vomiting; Diffuse Urticaria; Headache; Facial Edema; Moderate laryngeal edema (throat swelling); Moderate Bronchospasm or dyspnea; Hypertension; Abdominal Cramps.
Patient Care 23. List the severe contrast reactions: Life Threatening Arrhythmias (ventricular tachycardia); Altered Mental Status; Severe Hypotension; Severe Bronchospasm; Laryngeal edema; Pulmonary edema; Seizures, syncope, and death.
Patient Care 24. What is the technologistâs first responsibility during an adverse contrast reaction? Immediately discontinue contrast administration.
Patient Care 25. What medication is typically administered for hives and itching? Antihistamine (Benadryl).
Patient Care 26. List the contraindications for iodinated contrast: Iodine allergy; Renal Failure; Recent contrast administration.
Patient Care 27. Is there evidence that iodinated contrast poses a risk to a developing fetus? No evidence that iodinated contrast poses a risk to a developing fetus. Not enough information to conclude there is no risk at all. Recommendation: Do not use for routine screening exams; okay to use when needed for diagnostic purposes.
CT Radiation Safety 1. The Computed Tomography Dose Index (CTDI) is calculated using: A medical physicist scans two CTDI phantoms (16 cm and 32 cm) embedded with ionization chambers to measure absorbed dose.
CT Radiation Safety 2. CTDI is measured in units of: (choose one) Milligray (mGy).
CT Radiation Safety 3. Explain the Dose Length Product (DLP). The product of the dose multiplied by the scan length. DLP = CTDI x scan length. CTDI: milligray (mGy); scan length: cm; DLP: mGy-cm. Represents both dose to tissues and total volume of tissues exposed.
CT Radiation Safety 4. If the CTDI for a can is 21 mGy and the scan length is 42 cm, what is the DLP (use the correct unit)? DLP = CTDI x scan length = 21 mGy x 42 cm = 882 mGy-cm.
CT Radiation Safety 5. Is the CTDI or the DLP the best for evaluating the total amount of radiation received by the patient and estimating the biologic effects? NOTE â added after document was sent to duplicating 2026. DLP. Best for evaluating the total amount of radiation received by the patient and estimating biologic effects. Conversions can be used to convert DLP into effective dose (EfD).
CT Radiation Safety 6. Explain the concept of optimal dose in CT. Dose that produces quality images with lowest possible radiation dose. Dose below optimal will be less biologic risk to patient but images may not be of diagnostic quality. Dose above optimal will produce good images but with excessive dose to patient.
CT Radiation Safety 7. Indicate how each variable change listed below affects patient dose. Assume each change is occurring independent of the other variables. a. Increase mA; b. Increased kVp; c. Increased pitch; d. Increased beam width: Increase mA: patient dose increases (directly proportional). Increased kVp: patient dose increases. Increased pitch: patient dose decreases (inversely proportional). Increased beam width: patient dose decreases.
CT Radiation Safety 8. Why is it important to position the patient directly in the center of the gantry? Scans using automatic modulation adjust dose based on scout images. The patient should be positioned directly in the center of the gantry.
CT Radiation Safety 9. How does automatic dose modulation work in CT? Ability of the CT scanner to change the mA during the scan based on the density of the tissues being imaged. Scanner can use the localizer images to identify the densest portion of the patient.
CT Radiation Safety 10. Explain the difference between a dose notification and a dose alert. Dose notification: Displayed on control panel when the CTDI for the scan will exceed the predetermined dose notification values; gives technologist opportunity to review settings prior to exposing patient. Dose alert: Warning displayed when cumulative CTDI for a series of scans will exceed 1,000 mGy; requires technologist to enter their name and state reason for using a cumulative CTDI above 1,000 mGy.
CT Radiation Safety 11. State the main differences between lead shielding and bismuth shielding. Lead shielding can never be placed in the scan field due to unacceptable beam hardening artifact. Lead shielding also causes increased tube output and patient dose when using automatic dose modulation/automatic exposure control. ACR and AAPM no longer support all forms of lead shielding. Bismuth shields attenuate only low-energy photons and reduce entrance dose while maintaining image quality; can be placed after the localizer image is obtained; may still cause a small amount of image artifact.