10. MR Physics & Imaging

Introduction

  • MRI is non-ionizing
  • MRI can have both anatomical and functional imaging
  • MRI uses precession imaging
  • NMR is nuclear magnetic resonance, but is not related to radioactive decay (the nucleus is spinning, but there’s no radioactivity)
  • MRI offers images of the same object in different contrasts to bring attention to different areas

Physics

  • Protons and neutrons are spinning inside the atom, which act like magnets
  • Spinning nuclei in an atom ==must have an odd atomic number== or odd mass number to exhibit this behavior
      * If a nuclei has an even atomic number, the magnetic fields cancel out
      * Spinning nuclei possess angular momentum (J) called spin
      * Placed in a magnetic field, the spin of the protons and neutrons assume certain orientations
      * Each spin produces a microscopic magnetization vector (𝜇)
        * 𝜇 = 𝛾 * J where gamma is the gyromagnetic ratio (specific to each material)
          * Gyromagnetic ratio gamma is radians per second*Teslas
          * The gyromagnetic ratio can also be gamma divided by 2pi
  • The body is mostly water, which has a lot of hydrogen and is a great source for MRI signal
      * ==Hydrogen== has the highest gyromagnetic ratio
  • The net (macroscopic) magnetization of a magnetic field is zero in its natural state
      * A single particle will have its own nonzero magnetic field

Bulk Magnetization Vector

  • Once an external magnetic field (B0) is applied…
      * The spins of each particle align in parallel or anti-parallel (a semi majority in parallel)
      * The net (bulk) magnetization vector with an external magnetic field is nonzero
        * The bulk magnetization (𝜇 vector) vector is the source of the MR signal (also called M vector)
  • B0 is always applied along the z direction (up-down on the body)
      * Field strengths in MRI are commonly 1.5 T or 3 T (Tesla)
      * A 1.5T magnet is about 30,000 times stronger than Earth’s magnetic field
      * RF cage has to be in the MRI room to absorb any signals from outside, otherwise FM radio frequencies would be measured
  • The vector 𝜇 precesses (spins) around the z axis st an angular frequency 𝜔 (radians per second)
      * 𝜔0 = 𝛾 * B0
      * Larmor Frequency: The angular frequency at which the bulk magnetization factor precesses around the z-axis
      * f = 𝛾 / (2pi) * B0
      * Larmor frequency depends only on the material (𝛾) and the main magnetic field strength (B0)

Generating Data

  • Spin Equilibrium & spin excitation
      * At equilibrium, the net magnetization vector M precesses about the z-axis
      * When M is entirely along the z-axis, only the Mz (longitudinal) component exists, and the transverse component M(xy) = 0
      * If M tips away from the z-axis (during excitation), a component in the x-y plane is generated
      * The signal is always measured in the x-y plane
      * The signal component is always in the transverse x-y plane
  • RF excitation
      * An RF pulse moves M away from the equilibrium state (tips away from z-axis), by using energy from the RF pulse onto the magnetic field B1
        * B0 is the magnet by itself, and the magnet is always on
          * The coil with current running through it acts as the magnet, and is surrounded by liquid Helium to keep the coil cool (decreasing resistance and energy lost to heat)
          * Why do we want the maximum signal? To decrease noise
        * B1 is applied from external antennas
          * B1 is along on the x-axis, so B1x is the only non-zero value for B1
      * Resonance comes from B1 (the additional force) being at the same frequency as B0 (𝜔0) to make sure there is maximum energy being delivery
      * As M returns to equilibrium, an RF signal is produced by energy releasing from the absorbed energy that was given to the material from B1
  • RF Pulse Shape
      * While oscilating at 𝜔0, the RF pulse can have three shapes, rectangular, sinc, and triangular
  • Rotating frame of reference
      * Since the spins and B1 field are both rotating, you can image from a spinning reference
      * The transverse plane (xy) is rotating at 𝜔, which is equal to 𝜔0, so it can be considered a rotating plane (x’-y’)
  • Flip angle/tip angle
      * The flip angle depends on the (1) shape of the RF pulse, (2) field strength B1 and (3) duration of the RF pulse 𝜏p (tau p)
      * For a rectangular pulse after 𝜏 seconds, the vector M has rotated at an angle α
        * α = 𝛾 * *B1 ** 𝜏
      * For any B1(t)
        * α (always in radians) is the integration from 0 to 𝜏 over B1(t) with respect to t (in seconds)
        * For the max transverse component Mxy, α is set to pi/2 radians
          * α = 𝛾 * *B1 ** 𝜏 = pi/2
  • Relaxation: The process by which M returns to steady state configuration after a B1 is done being applied
      * The longitudinal Mz and transverse Mxy components vary as time changes, and are descibed by Bloch Equations
      *
      * Where T1 is the longitudinal relaxation time (spin minus lattice relaxation)
      * T2 is the transvese relaxation time (spin minus spin relaxation)
      *
      * The solutions become
        *
        * Mz^0 is the steady state of Mz
        * Mz(0+) is transverse magnetization Mx’y’ immediately after the RF pulse
        * We always measure the x-y plane and do everything else after
        * T1 is usually a LOT larger that T2
          * The longitudinal time is usually a lot longer than the transverse time (around 10 times more)
      *
      *
      * Mx’y’ decays at a faster rate (T2* instead of T2) because of signal desync
  • Spin Echo
      * To correct for dephasing and loss of Mx’y’ signal, another RF pulse (180 degree pulse) is applied to re-phase the spins
        * Spin echo makes the slower spins catch up with faster oens to sync the signal
  • Relaxation times T1 and T2 are dependant on materials (very high for water, really low for muscle, fat, tendons)

Acquisition & Contrast

  • Contrast in MRI is the difference between signal intensities generated by different tissues (B0, which is constant, vs gamma, which is material specific)
  • The difference in signal intensity is used to discriminate different tissues by representing them as brightness
  • Pulse sequence: Carefully timed set of scanner operations used to generate images.
      * Gradient-echo based pulse sequences are
      * Spin-echo based pulse sequences
        * 90 degree pulse
        * 180 degree pulse
        * Wait for relaxation echo
        * Record relaxation echo
        * Repeat

MR Signal Intensity

  • MR signal is always recorded in the xy plane
  • Signal intensity S of a SE sequence is
      * S = K x [H] x (1 - exp(-1 x TR / T1) x exp(-1 x TE/T2)
      * K is the scaling factor
      * [H] is the proton/spin density
        * Twice the number of spins means you have twice as large a signal
      * TR is the repetition time
      * TE is the echo time
      * T1 is the longitudinal relaxation time
      * T2 is the transverse relaxation time
  • Weighting
      * Proton density weighting: We want a long TR and a short TE
      * T1 weighting: We want an intermediate TR and a short TE (becomes a constant scaling factor, less noise)
      * T2 weighting: We want ta long TR (minimize T1 differences) and an intermediate TE
      *
      *
  • Gradients
      * Adding a small gradient field (extra magnetic field) allows the same material to be differentiated if it is at different distances
      * Gradient affects Larmor frequency
      * Gradients added using more coils
      * Gradients are localized in x, y, and z directions
        * Gz is a slice selection
          * z-directional coils are called Maxwell pairs
        * Gx is frequency encoding
          * x-directional coils are called Golay coils
        * Gy is phase encoding
          * y-directional coils are called Golay-type coils
      * Gradients vary the magnetic field strength along the z-direction, which affects the frequency and phase of the spins
  • Fourier Transform is used to break a signal into its freuqency components (sum of sinusoids) based on their amplitude and phase
  • Slice thickness
      * Centered at z=0 , B(z) = B0 + Gz(z)
      * Gz is gradient strength
      *
      *

Overview

  • The magnet is used to create polarization (creates the bulk of the ==magnetic== field)
  • The RF oil is used to promote excitation (sends RF energy at ==resonance== conditions)
  • The gradients are used to generate spatial localization (used to form the ==images==)
  • The receiver RF coils are there to target specific organs
  • \