LC

Week 9 Lecture 2: More B-Mode Imaging Artifacts

Depth Artifacts

  • Occur due to the incorrect depth display of tissues.

  • Types:

    • Propagation speed artifact

    • Reverberation artifact

      • Ring down artifact

      • Comet tail artifact

    • Range ambiguity artifact

  • Violated assumptions:

    • The transmitted wave travels along a straight line path from the transducer to the object and back to the transducer.

    • The round trip time is 13 microseconds for every cm of depth.

Propagation Speed Artifact

  • The ultrasound system calculates depth based on the propagation speed of tissue, assuming it's a constant (1540 m/s).

  • If the actual propagation speed is slower, the returning echo takes longer to arrive, and the system displays the echoes deeper than they are.

  • (u = 0.0022)

Range Ambiguity Artifact

  • Occurs when all echoes are not received before the next pulse is sent.

  • The system assumes all echoes are from the most recently transmitted pulse and places objects more superficially than they are.

  • Often seen with superficial cystic structures.

  • Can be mitigated by altering depth or PRF (Pulse Repetition Frequency) or using a lower frequency transducer.

  • Superficial reflected sound energy arrives at the transducer before a second sound pulse is generated

  • Deeper (delayed) arrive at the transducer after the transmit of the new pulse

  • The system interprets the echoes returning from depth as being associated with the second pulse

  • The system places these echoes in the image near field, assuming these are early returning pulses from the second pulse

  • Alter depth or PRF/utilize lower frequency transducer

Reverberation Artifacts

  • Tissues are displayed multiple times at equally spaced depths due to repeated reflections.

  • Only the first reflection is spatially correct.

  • Characteristics:

    • Parallel to the sound beam’s main axis.

    • Decreasing in intensity.

    • Equidistant from each other.

  • Echoes can appear between the transducer and a strong reflector or between two strong reflectors.

  • Types:

    • Simple reverberation

    • Ring-down artifact

    • Comet-tail artifact

Simple Reverberation Artifact
  • Sound bounces between the transducer and a strong specular reflector, or between two strong specular reflectors.

  • Often called “reverb.”

  • More common in superficial tissues where the sound beam has not been attenuated.

  • Commonly noted in anechoic structures.

Ringdown Artifact
  • Reflectors are gas bubbles.

  • Multiple echoes are seen due to sound reverberating back and forth between air/soft tissue interfaces and within gas bubbles.

  • Strong, bright, repeating lines are observed deep to gas bubbles.

Comet Tail Artifacts
  • Multiple echoes from repeated reflections within a small but highly reflective object.

  • The artifact decreases in amplitude and width (tapers and fades) as it gets deeper.

  • Often seen with metal, crystalline materials.

Beam Dimension Artifacts

  • Types:

    • Beam width artifact

    • Slice thickness artifact

    • Side lobe/grating lobe artifact

  • Violated assumptions:

    • All detected echoes originate from the axis of the main beam only, and beam dimensions are small in both section thickness (elevational) and lateral directions.

Beam Width Artifact

  • A strong reflector located within the widened beam beyond the margin of the transducer will generate detectable echoes.

  • These echoes are assumed to have originated from the narrow imaging plane and are displayed as such.

  • Occurs because of beam width divergence beyond the focal zone.

  • Typically seen in anechoic structures, e.g., gas shadowing seen “within” the bladder.

Slice Thickness Artifact

  • Similar to the beam width artifact but occurs due to the thickness (elevational plane) of the beam.

  • Often seen within anechoic structures at depth when adjacent bowel and bladder are contained within the same “slice thickness.”

Side Lobe/Grating Lobe Artifact

  • A strong reflector in a side lobe (single crystal transducer) or grating lobe (array transducer) can appear as an echo as if it arises from the central beam.

  • Example: Bowel with reverberation artifacts is misregistered in an anechoic ovarian cyst.

Beam Path Artifacts

  • Types:

    • Refraction artifact

    • Mirror image artifact

  • Violated assumptions:

    • The transmitted wave travels along a straight line path from the transducer to the object and back to the transducer.

    • All detected echoes originate from the axis of the main beam only, and the depth of objects is proportional to the round-trip time.

Refraction Artifact

  • The ultrasound system assumes that the beam has traveled in a straight line, but refraction has “bent” the beam due to Snell’s Law.

  • Refraction artifacts cause lateral displacement of reflectors.

Mirror Image Artifact

  • Duplication of a structure on either side of a strong specular reflector.

  • The true reflector and the false object are equal distances from the “mirror,” with the false object placed deeper on the image.