L3 Spectrophotometers

Introduction to Spectroscopy

  • Overview of spectrophotometers

  • Course: CH4303 Analytical Chemistry 1, Lecture 3

Key Definitions

Absorbance (A)

  • Definition: A = -log10 T = -log(P/P0) = log(P0/P)

  • A increases with attenuation of light

Beer's Law

  • Equation: A = ecb

    • A: absorbance

    • e: extinction coefficient (M^-1 cm^-1), wavelength-dependent

    • b: cuvette path length (cm)

    • c: concentration of absorbing species (M)

  • Relationship: Absorbance is linearly related to concentration and path length.

Transmittance (%T)

  • Relation to absorbance: %T = 100 - %A

  • Example Calculation:

    • If 99% of light is absorbed: A = -log(T), leading to A = log(100/1) = 2

Instrumental Components

Basic Components of Spectrophotometers

  1. Stable Light Source - Provides consistent radiation.

  2. Monochromator/Filter - Isolates a narrow wavelength region.

  3. Sample Container - Holds samples in the beam of light.

  4. Radiation Detector - Converts radiant energy to a measurable electrical signal.

  5. Signal Processor/Readout Device - Amplifies signal and displays magnitude.

Spectrophotometer Types

  • Most modern spectrophotometers are double-beam devices.

  • Require a baseline spectrum with solvent in both cuvettes.

Electromagnetic Spectrum

  • Spectroscopic methods categorized by region:

    • From radio waves (NMR) to X-rays and gamma rays.

Sources of Light

Visible Light

  • Tungsten filament lamp at ~2900 K, providing continuous spectrum (320-2500 nm).

UV Light

  • Deuterium lamps emit radiation (160-380 nm).

  • Use electric discharge to dissociate D2 into 2D+ and hν.

Infrared Light

  • Emitted by silicon carbide rod (Globar®) heated to ~1500 K.

  • Generates spectrum similar to blackbody at 1000 K.

Lasers and Monochromators

Lasers

  • Provide light of a single wavelength with narrow bandwidth.

    • Require population inversion to achieve lasing.

Monochromators

  • Utilize gratings and prisms to isolate narrow bands of radiation.

  • Benefit from stricter adherence to Beer's law and improved sensitivity.

Reflection Gratings

  • Polychromatic light focused onto grating, wavelengths diffracted at different angles.

  • Each groove on the grating acts as a partial radiation source.

Wavelength Selection and Interference

Bandwidth Selection

  • Narrower slit widths allow for greater resolution.

  • Effective bandwidth determined by slit width and grating.

Constructive and Destructive Interference

  • Constructive interference occurs when paths differ by integral multiples of the wavelength.

  • Destructive interference occurs when paths are out of phase by half-integral multiples.

Photodetectors

Phototubes

  • Emit electrons when struck by photons, generating current proportional to radiation intensity.

Photomultiplier Tubes

  • Similar to phototubes, but with additional electrodes (dynodes) amplifying the signal.

Photodiode Arrays

  • Measure all wavelengths simultaneously, enabling quicker spectrum acquisition.

Fourier Transform Spectroscopy

  • Acquires entire spectrum simultaneously and transforms into component frequencies.

  • Involves Fourier analysis to decompose curves into sine and cosine terms.

Michelson Interferometer

  • Uses beamsplitter to split light and create interferograms by measuring light intensity.

  • Constructive/destructive interference based on path length differences; termed as retardation factor (d).

  • Intensity at the detector is influenced by the path differences of light waves.

Summary of Interferograms

  • The sample will absorb specific wavelengths, altering the interferogram.

  • Analysis reveals the spectrum of the source minus absorption patterns of substances.