L3 Spectrophotometers
Introduction to Spectroscopy
Overview of spectrophotometers
Course: CH4303 Analytical Chemistry 1, Lecture 3
Key Definitions
Absorbance (A)
Definition: A = -log10 T = -log(P/P0) = log(P0/P)
A increases with attenuation of light
Beer's Law
Equation: A = ecb
A: absorbance
e: extinction coefficient (M^-1 cm^-1), wavelength-dependent
b: cuvette path length (cm)
c: concentration of absorbing species (M)
Relationship: Absorbance is linearly related to concentration and path length.
Transmittance (%T)
Relation to absorbance: %T = 100 - %A
Example Calculation:
If 99% of light is absorbed: A = -log(T), leading to A = log(100/1) = 2
Instrumental Components
Basic Components of Spectrophotometers
Stable Light Source - Provides consistent radiation.
Monochromator/Filter - Isolates a narrow wavelength region.
Sample Container - Holds samples in the beam of light.
Radiation Detector - Converts radiant energy to a measurable electrical signal.
Signal Processor/Readout Device - Amplifies signal and displays magnitude.
Spectrophotometer Types
Most modern spectrophotometers are double-beam devices.
Require a baseline spectrum with solvent in both cuvettes.
Electromagnetic Spectrum
Spectroscopic methods categorized by region:
From radio waves (NMR) to X-rays and gamma rays.
Sources of Light
Visible Light
Tungsten filament lamp at ~2900 K, providing continuous spectrum (320-2500 nm).
UV Light
Deuterium lamps emit radiation (160-380 nm).
Use electric discharge to dissociate D2 into 2D+ and hν.
Infrared Light
Emitted by silicon carbide rod (Globar®) heated to ~1500 K.
Generates spectrum similar to blackbody at 1000 K.
Lasers and Monochromators
Lasers
Provide light of a single wavelength with narrow bandwidth.
Require population inversion to achieve lasing.
Monochromators
Utilize gratings and prisms to isolate narrow bands of radiation.
Benefit from stricter adherence to Beer's law and improved sensitivity.
Reflection Gratings
Polychromatic light focused onto grating, wavelengths diffracted at different angles.
Each groove on the grating acts as a partial radiation source.
Wavelength Selection and Interference
Bandwidth Selection
Narrower slit widths allow for greater resolution.
Effective bandwidth determined by slit width and grating.
Constructive and Destructive Interference
Constructive interference occurs when paths differ by integral multiples of the wavelength.
Destructive interference occurs when paths are out of phase by half-integral multiples.
Photodetectors
Phototubes
Emit electrons when struck by photons, generating current proportional to radiation intensity.
Photomultiplier Tubes
Similar to phototubes, but with additional electrodes (dynodes) amplifying the signal.
Photodiode Arrays
Measure all wavelengths simultaneously, enabling quicker spectrum acquisition.
Fourier Transform Spectroscopy
Acquires entire spectrum simultaneously and transforms into component frequencies.
Involves Fourier analysis to decompose curves into sine and cosine terms.
Michelson Interferometer
Uses beamsplitter to split light and create interferograms by measuring light intensity.
Constructive/destructive interference based on path length differences; termed as retardation factor (d).
Intensity at the detector is influenced by the path differences of light waves.
Summary of Interferograms
The sample will absorb specific wavelengths, altering the interferogram.
Analysis reveals the spectrum of the source minus absorption patterns of substances.