RCP Section 1: Medical Gas Supply Equipment
Kinetic Theory of Gases
Gases Composition: Gases are comprised of discrete molecules in random motion.
Molecular Behavior:
Collisions between molecules are elastic - they bounce off one another.
Their activity is temperature-dependent.
No physical attraction exists between molecules.
Gas Pressure
Definition: Gas pressure is the force exerted by a gas per unit area on an object.
Influencing Factors:
Results from molecular collisions with each other and the walls of their container.
Temperature affects the velocity of gas molecules, with a rise in temperature leading to increased kinetic energy.
Pascal's Law
Principle: The pressure exerted by a confined fluid is undiminished and transmitted equally in all directions.
This principle explains that pressure against the wall of a container acts uniformly throughout.
Instruments Measuring Gas Pressure
cmH2O
Definition: Unit of measurement for gas pressure, equivalent to centimeters of water pressure.
Mercury Barometer
Description: Measures air pressure changes using a glass tube partially filled with mercury.
Function: The mercury column opposes atmospheric pressure to determine atmospheric pressure levels.
Aneroid Barometer
Description: Changes in air pressure are measured without liquids, utilizing a metal container, a spring, and a pointer.
Bourdon Gauge
Design: A dial that shows oxygen flow in liters per minute.
Construction: Consists of a hollow coiled metal tube with an elliptical cross-section, commonly found on medical gas cylinders.
Gas Flow Principles
Gas Flow
Definition: Gas flow occurs due to pressure differences between two points, moving from high to low pressure.
Factors Influencing Flow Velocity:
Pressure gradient
Size of the opening
Bernoulli's Principle
As gas flows through a tube, an increase in the velocity results in a decrease in lateral pressure, maintaining constant total energy.
Viscous Shearing
Definition: A high-velocity jet injected into a stationary gas creating shearing forces.
Vorticity
Refers to the decrease in jet gas velocity due to the swirling mixture of two gases.
Ducted Ejectors
Utilize a nozzle, viscous shearing, and vorticity to maintain total flow efficiently without changing tube diameter.
Used in nebulizers for high velocity and stable pressure, less affected by downstream pressure changes.
Venturi's Principle
An extension of Bernoulli's principle where a tube with a larger radius (less than 15°) draws ambient air into a high-velocity gas flow.
Oxygen Delivery Devices
HAFOE (High Air Flow with Oxygen Enrichment) Mask
Also known as the Venturi mask.
Increased downstream pressure decreases ambient air entrainments, resulting in higher fractional inspired oxygen (FiO2).
Venturi Mask
Designed for oxygen delivery, featuring escape holes for exhaled air and adjustable color-coded entrainment ports for oxygen concentration.
Flow Dynamics and Gas Behavior
Choked Flow
Occurs when actual pressure ratio is below the critical ratio, restricting flow.
Defined as when gas velocity stabilizes at a specific pressure, for example, at 26 psi driving pressure in nebulizers.
Reynolds Number
Determines gas flow type (laminar vs. turbulent).
Laminar Flow: Smooth and less energy intensive (Re < 2000).
Turbulent Flow: Erratic and more energy-consuming (Re > 2000).
Formula: R = (velocity x density x diameter) / viscosity.
Poiseuille's Law
Describes the relationship between flow rate, pressure gradient, vessel radius, and viscosity.
As the tube radius is halved, resistance to flow increases by a factor of 16.
STP (Standard Temperature and Pressure)
STP involves conditions of 0.00°C and 1 atm (760 mmHg).
Conversion from Celsius to Kelvin: K = C° + 273.
Gas Laws
Boyle's Law
Formula: P1V1 = P2V2.
Describes the relationship between pressure and volume at constant temperature.
Charles' Law
Formula: V1/T1 = V2/T2.
States that volume increases with temperature at constant pressure.
Gay-Lussac's Law
Formula: P1/T1 = P2/T2.
Describes pressure changes with temperature when volume is constant.
Combined Gas Law
Formula: P1V1/T1 = P2V2/T2.
Integrates pressure, volume, and temperature relationships for a fixed amount of gas.
Dalton's Law of Partial Pressures
The total pressure of a gas mixture equals the sum of the partial pressures of individual gases in the mixture.
Fick's Law of Diffusion
The net diffusion rate of a gas across a fluid membrane is proportional to the difference in partial pressure and inversely proportional to the membrane's thickness.
Henry's Law
States gas solubility in liquid is directly proportional to its partial pressure above the liquid.
Graham's Law
The rate of diffusion of a gas is inversely proportional to the square root of its molar mass.
Medical Equipment and Properties
Water Content in Gas
At 37°C (body temperature), gas can hold 44 mg of water per liter; at 760 mmHg (1 atm), 47 mmHg is attributed to water vapor pressure.
FiO2 (Fraction of Inspired Oxygen)
Represents the percentage of oxygen in inhaled gas.
Compressors
Provide compressed air for equipment while mixing with O2; must be oil-free to avoid fire hazards.
Operate under 50 psi for safe use.
Types:
Piston
Diaphragm
Centrifugal
Piston Compressor
Utilizes a motor-driven piston to compress atmospheric air.
Diaphragm Compressor
Employs a flexible diaphragm driven by an electric motor suitable for powering small nebulizers.
Centrifugal Compressor
A dynamic compressor that uses an electrically powered impeller to generate large volumes of air suitable for medical facilities.
Oxygen Production Methods
Fractional Distillation (O2)
Air is liquefied, cooled, and subsequently heated.
Nitrogen and trace gases are separated due to lower boiling points, resulting in 99.5% pure oxygen.
Physical Separation (O2)
Involves molecular sieves and semi-permeable plastic membranes for oxygen extraction.
Concentrators
Medical devices that separate nitrogen from ambient air, allowing for breathability of up to 95% pure oxygen.
Molecular Sieve
A material featuring small, uniform holes that selective allow gas molecules of a certain size to pass while blocking larger molecules.
Membrane Enricher
Uses a semipermeable polymer to allow O2 molecules to pass based on size differences, designed for oxygen enrichment.
Medical Gas Storage and Supply
Bulk Supply Systems
Liquid reservoir systems containing large amounts of gas; 1 ft³ of liquid O2 is equivalent to 861 ft³ gas.
Portable Reservoirs
Liquid reservoir systems with liquid duration based on weight measurement, accommodating both stationary home models and ambulatory versions.
Supply (Piping Systems)
Includes manifolds, bulk liquid systems, or both, with a required reserve supply tested to 1.5 times working pressure.
Safety Features and Cylinder Information
Safety Features (Piping Systems)
Alarms to alert drops in system pressure.
Zone valves provided for emergency shutoff (e.g., fire).
Pressure sensors ensure maintained levels at 50 psi.
Cylinder Designations
E Cylinder: Holds 625L of oxygen, with a full pressure of 2200 psi.
H Cylinder: Large stationary oxygen tank holding 6900L of oxygen.
Cylinder Construction
Regulated by DOT, using chrome-molybdenum steel, aluminum alloy, and lightweight aluminum with carbon fiber wrap.
Fill to 3,000 psi with cylinder markings indicating important manufacturing and test information.
Color-Coded Gas Cylinders
Green: Oxygen (O2)
Gray: Carbon Dioxide (CO2)
Light Blue: Nitrous Oxide (N2O)
Orange: Cyclopropane (C3H6), very reactive.
Brown: Helium (He)
Yellow: Air
Brown and Green: Heliox
Gray and Green: CO2 + O2
Cylinder and Valve Specifications
Tank Factors
Calculation: Size (ft³) x 28.3L/ft³ / Pressure when full (psi).
Table for E and H cylinders detailing tank transport safety factors.
Cylinder Valves Overview
Direct Acting Cylinder Valve: Needle valve with two washers; opens directly for gases like oxygen and helium.
Diaphragm Cylinder Valve: An indirect acting system utilized in larger tanks (H cylinder).
Safety Systems
American Standard Safety System (ASSS): Prevents the attachment of incorrect equipment through unique threading.
Pin Index Safety System (PISS): Sorts connections using different pin placements to avoid incorrect gases.
Safety Mechanisms and Reducing Valves
Safety Mechanisms
Frangible Disk: A thin metal disk that ruptures excessively high pressure.
Fusible Plug: Melts at temperatures exceeding 208-220°F.
Reducing Valves
Reduce pressure to a working pressure of 50 psi, with various types suited for different applications, particularly medical ventilators.
Single Stage Reducing Valve: Operates with internal spring tension and is responsive to gas pressure variations.
Multistage Reducing Valve: Provides more precision and flow by combining multiple stages.
Regulators and Flow Meters
Regulators
A combination of a reducing valve and a flowmeter.
Types of flowmeters include Bourdon types and Thorpe tubes.
Oxygen Flowmeters
Include fixed orifice flowmeters that manage gas flow based on the orifice size, common in E cylinder applications.
Bourdon-Gauge Flowmeter: Combines pressure reading with flow measurement but can misrepresent flow under back pressure.
Main Supply Shut-Off Valve
Controls gas flow from the main source to all dispensing locations, centrally shutting off gas supply when needed.
Flowmeter Types
Uncompensated Thorpe Tube Flowmeter: Features needle valves proximal to the flow tube and isn't compensated for back pressures.
Compensated Thorpe Tube Flowmeter: Needle valves distal to the flow tube allow the ball to jump without affecting pressure.
Flowmeter Ranges and Proportioners
Flowmeter Ranges
Typical ranges from 0-15 LPM; lower ranges (0-3 LPM) are for pediatric/neonatal use or COPD patients to match smaller airways.
High flowmeter ranges can reach up to 75 LPM, suitable for CPAP or Vapotherm.
Proportioners (Air-Oxygen Blenders)
Facilitate precise mixing of air and oxygen, providing 50 psi with concentration control via proportioning valves and built-in alarms for pressure drops.
Demand Pulse Flow Delivery
Demand Pulse Flow O2 Delivery Devices
Deliver oxygen only during inspiration, minimizing the unwanted oxygenation of anatomical dead space.
Known as Oxygen Conserving Devices (OCD), these often require exercise oximetry studies for appropriate use.
Diameter Index Safety System (DISS)
Ensures non-interchangeable, gas-specific connections to the pipeline, preventing incorrect gas delivery to patients.