Computed Tomography and Its Application to Nuclear Medicine

Computed Tomography and Its Application to Nuclear Medicine

Chapters 18-19 (Cherry), Chapter 11-12 (Gilmore)

Lecture Schedule
  • Breakout Sessions

  • Scheduled Breaks:

    • 1:50

    • 2:50

    • 3:45

Objectives
  • Explain CT scanner imaging and scanner components

  • Describe CT quality control methods

  • Discuss SPECT/CT and PET/CT Hybrid Imaging

  • Explain additional quality control (QC) methods needed for hybrid scanners

  • Identify artifacts generated by hybrid scanning

  • Describe MRI imaging and the use of PET/MR technology

Basics of X-Ray and CT
  • Production of X-Rays:

    • X-ray tube involves a filament that releases electrons, which are accelerated toward a tungsten anode.

    • Electrons undergo ionization and Bremsstrahlung interactions in the anode.

    • Bremsstrahlung and Characteristic X-Rays:

    • These x-rays are directed toward the patient.

    • Operator Control:

    • Controls for current (mA) and potential (kVp).

X-Ray Tube (Figure 18-2)
  • Description:

    • An evacuated glass container housing a cathode (heated tungsten filament) that produces electrons.

    • Anode containing tungsten target where electrons interact via Bremsstrahlung and ionization interactions to generate x-rays.

    • Bremsstrahlung interactions yield many low-energy x-rays that do not contribute to imaging; filters are used to absorb these.

    • Collimators absorb wide-angle x-rays, creating a compact x-ray beam.

X-Ray Energy Spectrum
  • Effect of kVp on Spectrum:

    • A higher kVp generates a Bremsstrahlung spectrum that extends to higher energies.

  • X-Ray Production:

    • Tube current (mA) determines the number of electrons and thus the quantity of x-rays produced
      (Figure 18-3).

CT Gantry (Figure 18-5)
  • Description:

    • Houses the x-ray tube and detector arrays opposite each other.

    • In helix CT, the gantry rotates continuously around a table that moves.

X-Ray Detection
  • Types of Detectors:

    • Scintillating or solid-state detectors.

    • Photodiodes convert x-rays into current signal.

    • Small detector arrays (0.5 mm dimensions).

  • Tissue Attenuation:

    • Different tissues (bone, soft tissue, air) absorb varying amounts of x-rays, resulting in different intensities passing through the object.

Image Reconstruction
  • Based on the attenuation equation and Radon’s concept of reconstructing 3D objects from multiple 2D projections.

  • Each small cube of tissue has its unique attenuation value ($ ext{µ}$).

  • Filtered Back Projection (FBP):

    • Techniques include using ramp and Shepp-Logan or Hamming filters for reconstruction.

Multidetector Helical CT
  • Helical Scan Motion:

    • The table moves through the gantry while the gantry rotates.

    • One rotation is divided into approximately 1000 views akin to 2D projections in SPECT.

    • The x-ray beam is cone-shaped to illuminate all detectors constantly.

Pitch in Multidetector Helical CT
  • Definition:

    • Ratio of table speed to gantry rotation speed.

  • Interpretation of Pitch (>1):

    • Indicates that not all tissues are directly exposed to x-rays.

  • Pitch Equation:

    • extpitch=table movement per rotation (mm)(number of slices)×(slice width (mm))ext{pitch} = \frac{\text{table movement per rotation (mm)}}{\text{(number of slices)} \times \text{(slice width (mm))}}

Interpolation of Helical Acquisitions
  • Creates a transverse slice from interpolation data of helical passes on either side.

  • Each point in the helical sinogram contributes based on proximity to slice location (Figure 18-7c).

  • Multiple slices can be created through interpolation using appropriate weights.

  • Cited Reference: Bushberg et al., The Essential Physics of Medical Imaging.

CT Image Acquisition
  • Procedure:

    • An IV catheter is placed, and the patient consumes two liters of dilute barium oral contrast.

    • A topogram is acquired to set scan limits for CT acquisition, performed while the x-ray tube is static above the patient.

    • IV contrast is administered via a high-pressure injector, and CT acquisition occurs as the bolus reaches the target area, typically requiring breath-hold.

CT Image Display
  • Attenuation Measurement:

    • Expressed relative to water using Hounsfield units (HU), with a range from -1000 (air) to +1000 (metal). Water is defined as 0.

  • Window/Level Combinations:

    • Selection of HU levels to manipulate gray scale display, with equations for windowing where HU<em>y=μμ</em>waterμyμHU<em>y = \frac{\mu - \mu</em>{water}}{\mu_{y} - \mu}

Image Quality in CT
  • Image Contrast:

    • Can visualize attenuation differences as low as 0.5%.

    • Most affected by mA (number of x-rays).

  • Spatial Resolution:

    • Typically around 0.1 mm, mainly determined by detector size and focal spot.

  • Image Noise:

    • Dependent on mA, kVp, and pitch.

Radiation Dosimetry
  • Radiation dose metrics include:

    • CT Dose Index (CTDI100):

    • Accumulated dose from multi-rotation scan of dosimetry phantom.

    • Dose-Length Product (DLP):

    • Total radiation measure for an acquisition.

    • Radiation Protection for Technologists:

    • Safety measures against both scattered radiation and leakage radiation from the x-ray tube.

Contributors to Radiation Dose
  • Key factors affecting radiation dose include:

    1. Patient size and density

    2. Detector size

    3. Patient dose

    4. Slice thickness

    5. Image matrix size

    6. Display field

Daily Quality Control for CT
  • Procedures include:

    • Warm-up of x-ray tube, verification of tube output, and detector response across various kVp and mA settings.

    • Assessments for tomographic uniformity, CT number accuracy, image noise, and table increment accuracy.

    • Accuracy evaluations via CT dose index (CTDI), slice localization, gantry alignment, and slice thickness.

Image Quality Factors in CT
  • Include:

    • Contrast resolution

    • Linearity

    • Noise

    • Uniformity

    • Spatial resolution

    • Temporal resolution

Improving Contrast Resolution
  • Achieved through:

    1. Smaller field of view

    2. Thicker slices

    3. Smaller matrix size

    4. Higher mAs values

    5. Utilizing low pass filters

Factors Influencing Quality Control
  • Determined by technologist control:

    • Selectable scan parameters

    • Viewing conditions

    • Beam geometry

    • Image receptor type

Slice Thickness and Spacing
  • Thinner slices yield more noise and cover less anatomy but enhance image quality.

  • Overlapped slices provide the best image quality but increase patient radiation exposure.

Artifacts in CT
  • Common artifacts include:

    • Partial volume averaging

    • Motion artifacts

    • Metal artifacts

    • Equipment artifacts

    • Ring artifacts

    • Cone beam artifacts

    • Edge gradient artifacts

    • Out of field artifacts

    • Beam hardening artifacts

  • Streak Artifact:

    • Influenced by:

    1. mAs and kVp settings

    2. Patient size

    3. Use of CT contrast agents

Hybrid Imaging (Chapter 19, Cherry)
  • Defining Hybrid/Fusion Imaging:

    • A process where patients are scanned using CT, MRI, and either SPECT or PET modalities.

    • Spatial resolutions:

    • CT: 5-10 mm

    • MRI: Approximately 1 mm

    • SPECT: Greater than 10 mm

    • Dedicated PET: 5-10 mm

  • Accurate image registration is essential for avoiding artifacts in images combined from different modalities.

    • Fusion increases accuracy for surgeons and radiologists in image interpretation.

Steps for Image Fusion
  1. Transfer data to a single computer system.

  2. Ensure registration of 3D or tomographic datasets to a common coordinate system (e.g., DICOM compatibility).

  3. Transform each dataset into the shared coordinate system with consistent spatial resolution and alignment along X, Y, and Z axes.

  4. Display and evaluate combined registered images using DICOM-compatible software.

Methods of Image Registration
  • Visual Method:

    • The viewer aligns images manually, typically within an error margin of 4 mm or 3 degrees.

  • Interactive Landmark Method:

    • At least three landmarks are identified for each modality to line up data sets.

  • Automated Registration:

    • Utilizes algorithms to improve accuracy by matching corresponding surfaces across datasets through matrix multiplications for transformation.

SPECT/CT Details
  • SPECT/CT systems consist of:

    • Acquisition computer

    • CT component

    • Patient bed

    • Detector heads, including:

    • Collimator

    • Scintillator

    • Photomultiplier tubes and associated electronics

Quality Control for SPECT/CT
  • Combined QC Procedures for SPECT/CT include:

    • Water phantom QA

    • Tube warm-up protocols

    • Air calibration for fast QA procedures

    • Water phantom checks for slice thickness, accuracy, and positioning.

    • Frequency of assessments:

    • Daily for most tests

    • Monthly for advanced verification processes

Advantages of Combined PET/CT and PET/MR
  • Benefits include:

    • Co-registered functional and anatomical information from PET combined with CT/MR scans.

    • Reduction in examination time by using available CT/MR data in place of lengthy separate PET transmission images.

    • This increases accuracy and limits noise propagation in attenuation correction.

Fusion of PET with CT/MRI
  • PET Technology:

    • Identifies early biochemical markers of disease, detecting metabolic changes such as increased glucose consumption in tumors even before structural changes manifest.

  • CT and MR Capabilities:

    • Reveal structural changes—fusion of PET images with CT and MR enables comprehensive diagnosis.

PET/CT Gantry and Issues
  • Gantry Design:

    • Incorporates both PET and CT tomographs, ensuring coregistration.

  • Acquisition Process:

    • Generally involves performing a PET scan following a CT scan, with the earlier CT providing an attenuation map for reconstructing the PET data and anatomical localization.

  • Challenges:

    • Balancing between diagnostic and low-dose CT scans while converting Hounsfield units to 511-keV attenuation during PET scans.

    • Addressing misregistration and CT truncation artifacts.

    • Ensuring proper certification of technologists and physicians for standard operation and interpretation of PET/CT studies.

Image Display in CT and PET
  • Maximum Intensity Projections (MIP):

    • Generated from stacked image slices showing the brightest pixel along a defined path, enhancing visualization.

  • Contrast in Imaging:

    • CT images offer a gray scale contrast of approximately 4000, whereas nuclear images provide a contrast scale of just a few hundred.

    • Allow for generation of transaxial slices that can be repositioned to create coronal and sagittal views.

Misregistration and Interleaved Sampling
  • Misregistration Causes:

    • Incorrect reconstruction arises when CT and PET images misalign due to patient movement, predominantly respiratory.

    • Artifacts may also occur from differences in CT-based attenuation correction and the handling of overlapping data sets.

Artifacts Caused by CT Attenuation Correction
  • Artifacts may arise due to:

    • Resolution mismatch when combining datasets from different modalities leading to tissue boundary issues.

    • Patient motion between CT and SPECT or PET scans.

    • High-Z materials' presence and use of iodinated IV contrast agents can exacerbate these artifacts.

Understanding PET and CT Scans
  • Distinctions between imaging modalities include:

    • PET scans reflect cellular activity, while CT scans show structural details in organs and bones.

    • The combination of PET/CT offers precise localization of increased cellular activity in the body.

Magnetic Resonance Imaging (MRI)
  • Concept of Magnetic Resonance:

    • Involves transferring electrical energy to atoms with unpaired protons under a magnetic field.

    • MRI employs hydrogen atoms and captures the signals emitted as they return to equilibrium post resonance interaction.

Net Magnetization in MRI
  • Magnetic Susceptibility:

    • Hydrogen protons align with or against a strong magnetic field, generating a net magnetic moment pointing in the direction of the field.

    • Precession of these protons results in a magnetic moment primarily aligned in the Z-axis.

Axes and Vectors in MRI (Figure 19-2)
  • Expressing Magnetization:

    • Net magnetization is illustrated by vectors, where MZ represents longitudinal magnetization, and MXY represents transverse magnetic vector.

    • Application of a second magnetic field causes the vector to move into the transverse plane.

Precession and the Larmor Equation
  • H protons precess in phase when exposed to an RF pulse at their Larmor or resonance frequency.

  • The frequency of precession is proportional to the magnetic field strength, aiding in creating a magnetic moment in the transverse plane.

Tissue Disturbance and Relaxation
  • Upon removal of the RF pulse, hydrogen atoms relax to random precession (relaxation process).

  • Relaxation components include:

    • T1 Relaxation: Growth of MZ vector

    • T2 Relaxation: Loss of MXY vector

  • The MRI tomograph measures the changing magnetic moment in the transverse plane as the free induction decay signal.

The Pulse Sequence
  • A series of RF pulses combined with magnetic field gradients facilitate the collection of signals that reflect differing relaxation properties.

  • Spin Echo Sequence:

    • An acquisition technique using 90° and 180° RF pulses, generating detectable signals as an 'echo' of the initial magnetic responses.

Forming the Magnetic Resonance Image
  • Susceptible MR signals are encoded by applying magnetic gradients reflecting their 3D body location.

  • Gradient Types include:

    • Slice Selection Gradient: In the Z direction selects a specific transverse slice.

    • Frequency Encoding Gradient: Determines resonance frequency per slice column in the X direction.

    • Phase Encoding Gradient: Creates phase shifts across different tissue rows along the Y direction, altering the gradient with each desired pixel row.

Data Storage and Image Reconstruction
  • Data undergoes collection in a k-space map; dimensions can be specified and may vary.

  • Each slice corresponds to its own k-space matrix:

    • One axis for frequency information and another for phase information, reconstructed using Fourier transformation.

MR Instrumentation
  • Key components include:

    • Superconducting magnet (requires liquid helium cooling),

    • Gradient coils constructed from loops of current-carrying wire,

    • RF coils for pulse transmission/receipt,

    • Computers for master control, pulse programming, and processing.

Tissue Characteristics in MRI
  • Adjustments made to repetition and echo times (TR and TE):

    • For T1-weighted images (short TE, short TR): Fat appears bright.

    • For T2-weighted images (long TE, long TR): Water/CSF appears bright, with fat remaining dark.

  • Proton Density-Weighted Images:

    • Achieved with long TR and short TE, showing hydrogen atom density with less contrast than T1 or T2 images.

Safety Issues in MRI
  • Potential Hazards:

    • Magnetic fields pose risks of localized heating and nerve stimulation.

    • Screening for metal implants is imperative to avoid accidents.

    • Objects susceptible to magnetic fields can become projectiles.

    • Noise from gradient coils can provoke discomfort among patients.

  • Use of Gadolinium:

    • Administered for contrast with a safety profile akin to CT contrast agents.

PET/MRI Technologies
  • Commercial Approaches:

    • Systems like GE Healthcare offer PET/CT and MRI tomographs connected via table tracks.

    • Alternative companies (Philips, Siemens) develop integrated systems optimizing PET-MRI synergy.

  • Technical Challenges:

    • PMTs need placement outside the MR tomograph or using fiber optics/avalanche photodiodes to protect against interference.

    • Ensure PET tomographs do not disturb MR signal for accurate imaging.

  • Clinical Benefits:

    • MRI's superior tissue contrast enriches anatomical correlation and clinical effectiveness.

    • True simultaneous imaging through combined modalities allows advanced evaluations for diagnostics.

Summary of MRI Techniques
  • MRI relies on the magnetic properties of protons in hydrogen, aligning protons in a magnetic field, and modifying that alignment with RF pulses.

  • Signals correspond to protons’ return to equilibrium and are measured to create MR images.

  • The integration into three-dimensional imaging harnesses precise control and technical nuances reflected in instrumentation, safety considerations, and procedural steps.