Ch. 8 X-Ray Unit Circuitry Lecture Notes 10/14/25
Types of Current
Direct Current (DC)
Definition: DC is defined as current that flows in only one direction at all times.
Production: Produced by capacitors (batteries) and rectifiers (devices that convert AC to DC).
Application: Primarily used by x-ray tubes.
Alternating Current (AC)
Definition: AC is current that flows in a positive direction for half of the time and in a negative direction for the other half.
Production: Generated by generators.
Application: Used by transformers in x-ray circuits; transformers require alternating current to function properly.
Basic X-ray Machine Circuit
Source of Voltage: The incoming voltage is supplied by the local power plant.
Line Voltage Compensator: Ensures consistent voltage delivery to the x-ray tube and circuit.
Analogy: Functions similarly to a surge protector, stabilizing the incoming voltage.
Three Circuits Making Up the X-ray Circuit
Primary (Low Voltage) Circuit
Secondary (High Voltage) Circuit
Filament Circuit: Operates at low voltage and high current.
Transformers
Purpose of Transformers
To either increase or decrease voltage levels within a circuit.
Basic Components of a Transformer
1st Coil
2nd Coil
Ferromagnetic Core
Operation Principles
Mutual Induction: Transformers operate based on the principle of mutual induction and require alternating current exclusively.
Voltage and Current Relations: The primary voltage and current do not equal the secondary voltage or current unless the number of turns in the coils is equal.
Autotransformer
Controls the kilovolt peak (kVp) output of the circuit.
Transformer Law for Voltage and Current
Voltage Relation:
Formula:
Where:
= Voltage in secondary winding
= Voltage in primary winding
= Number of turns in secondary winding
= Number of turns in primary winding
Current Relation:
Formula:
Important Note: The power output cannot exceed the power input, thus emboldening the importance of the relationship defined by:
Types of Transformers
Autotransformer: Can increase or decrease both voltage and amperage.
Step-up Transformer: Provides high-voltage (low current, with fewer electrons) to the tube.
Step-down Transformer: Provides high current (low voltage) to the filament.
Necessitated for thermionic emission to occur, allowing electrons to be boiled off.
Mutual Induction: Maximum mutual induction occurs in shell core transformers.
Basic components include iron core, primary input, and secondary output.
Primary Circuit
Contains the Autotransformer.
Requires a substantial amount of electrons in motion.
Only consists of the primary coil of a step-up transformer.
Controls located on the low voltage side of the circuit.
Rectifiers
Purpose of Rectifiers
Designed to convert AC to DC.
Types of Rectifiers
Solid State Devices: Allow current flow in only one direction.
Circuit Positioning: Rectification occurs post step-up transformer and prior to the x-ray tube within the x-ray circuit.
Essential for pulsing the magnetic field from the primary to secondary coil.
Direction of Current and Electrons
Current and electrons flow in opposite directions:
Electrons Flow: From cathode to anode.
Types of Rectification
Half-wave Rectification: Stops backflow of negative current.
Full-wave Rectification: Converts backflow (negative) to forward flow (positive current).
Note: Exact number of rectifiers in a circuit is not necessary to memorize.
Phase Types of Current
Single Phase: Can be denoted as 1°.
Three Phase, Six Pulse: Denoted as 3°12P.
Three Phase, Twelve Pulse: Denoted as 3°12P.
Ripple Characteristics: Maximum and minimum ripple values to be monitored to maintain voltage stability.
Rectification and Voltage Ripple
The amount of voltage corresponds with variations in x-ray circuits:
Single Phase: 1/3% of set kV with a 100% ripple.
Three Phase, Six Pulse: 91% of set kV with a 14% ripple.
Three Phase, Twelve Pulse: 97% of set kV with a 4% ripple.
High-Frequency Generator: 100% of set kV with <1% ripple.
Filament Circuit
Necessitates a step-down process to elevate the current, ensuring that electrons are boiled off through thermionic emission.
Houses the step-down transformer which decreases voltage while increasing current.
Rheostat: A resistor that changes its resistance to allow current flow to decrease.
Control Panel on Low Voltage (Primary) Side
kVp and mA selectors: Connected to the autotransformer.
Timer: Mechanism to terminate exposure.
Types of timing mechanisms include:
Synchronous (less precise)
Electronic
AEC (Automatic Exposure Control)
Automatic Exposure Control (AEC)
Purpose of AEC
Ensures correct exposure settings regardless of patient size by controlling the duration of exposure.
Tech Technologist Role: Selection of mA and kVp is done by the technologist.
Critical Factors: Positioning and centering of the patient is vital for optimal imaging.
Components of AEC
Ion Chambers: 3-5 independent chambers positioned between the patient and the image receptor.
Backup Timer: Generally set to automatically terminate exposure after 5 seconds to prevent tube overload.
Re-calibration Necessity: Required when new image receptors or grids are introduced into the system.
Rectifiers Used in X-ray Circuits (Non-examination Material)
Single Phase, Full Wave: 2 or 4 rectifiers.
Three Phase, Six Pulse: 6 rectifiers.
Three Phase, 12 Pulse: 12 rectifiers.