MRI Systems - Fringe Fields, Shielding, Shimming and Gradients
Fringe Fields
- Fringe field is the static magnetic field around the scanner, measured in Gauss.
- It extends beyond the walls, floor, and ceiling.
- Different MRI systems have different fringe fields.
- Vertical open bore magnets have smaller fringe fields compared to closed bore horizontal magnets.
- Permanent magnets have relatively low fringe fields.
- Unshielded high field systems can have large fringe fields, extending several feet or meters.
- Fringe field must be contained, which is a consideration when siting a scanner.
Magnetic Shielding
- Shielding can decrease the fringe field.
- Consider potential projectiles when siting a magnet.
- Two methods of magnetic shielding:
- Passive shielding: No current or passivity. Achieved by surrounding the magnet or lining the magnet room with steel plates. More about the room around the magnet.
- Active shielding: Implies current and activity. Is within the magnet itself. More expensive and inconvenient. Can weigh up to 40 tons. Requires specially prepared foundations.
- Most superconducting scanners use active shielding.
- Uses an additional solenoid electromagnet around the main magnetic coils at each end of the bore.
- These are inside the cryostat and are superconducting coils.
- Exhibit equal but opposite effect to the main magnet which leads to a significant reduction in the size of the fringe field.
- For an MR system that is not shielded, the scan room would need to be the size of a doubles tennis court to contain the 5 Gauss line.
Shim Coils and Shim System
- When taking an image, such as of the abdomen, fat suppression is used.
- Fat suppression should appear uniform across the image.
- The magnetic field arrives from the factory at 1000 parts per million but needs to be within 4 parts per million for a good image.
- Spectroscopy studies require better than 1 part per million.
- Passive shimming: Achieved by placing small ferromagnetic plates in specially constructed nonferrous metal trays around the circumference of the warm bore of the magnet.
- Refers to the circumference of the inner wall of the cryostat.
- Anything close to the side of that magnet is not as homogenous.
- Passive shimming is used to get the magnetic field to a particular level of homogeneity, and then active shimming is used to optimize for each patient exam.
Gradient Coils and Gradient System
- Found in the warm bore of the magnet is the gradient set.
- It's a cylindrical structure containing three individual electromagnets.
- Also contains 18 individual solenoids that make up the active shim system.
- Gradient coils are supplied by one or two powerful amplifiers.
- Gradient set is at room temperature, not superconducting with three planes.
- Gradient coils are used to spatially locate where a signal is within that body and there used for spatial encoding.
- They are also used to rephase spins to produce echoes.
- A gradient is a linear slope in the magnetic field strength across the imaging volume in a particular direction.
- The strength of the magnetic field can be changed by:
- The current passing through the windings.
- The number of windings in the coil.
- The diameter of the wire used in the windings.
- The distance of spacing between the windings.
- Altering any of the first three factors would change the amplitude or strength from a low to a higher field.
- Altering spacing between the loops is possible to slope the magnetic field.
- The coil has 12 windings uniformly spaced that's attached to an electrical terminal at each end.
- The current flows in one direction through the coil, resulting in the magnetic field.
- Each time a gradient is switched on, this produces noise.
- By changing the field strength in a linear fashion using a gradient, the processional frequency and the phase of magnetic moments are also altered linearly.
- Faults in gradient coils or gradient amplifiers can result in geometric distortions in the MR images.
Gradient Characteristics
- Gradient characteristics need to be understood to accomplish the goals of spatial encoding, refocusing, and other tasks during image acquisition.
Definitions
- Gradient strength/amplitude: Defines how steep or strong a particular gradient is. Measured in milliteslas per meter or Gauss per centimeter.
- Gradient speed/rise time: Defines the time it takes for the given gradient to reach its maximum amplitude. Measured in microseconds.
- Slew rate: Defines the time it takes for a given gradient to reach its maximum amplitude and what the amplitude is.
- Rise time affects how fast a gradient can be switched on an off, affecting scan time.
- As slew rates increase, the potential for time-varied magnetic field effects increases.
- As the duty cycle increases, gradient heating can increase, and the number of attainable slices can be reduced.
Acoustic Noise
- Acoustic noise generated by the scanner is caused by vibrations of the gradient set.
- Higher amplitude gradient and rapid gradient activation increase acoustic noise.
- Manufacturers are modifying gradient systems in an attempt to reduce noise.