Medical Gases, Delivery Systems, Safety Standards, and Equipment Calculations

Physical and Chemical Properties of Medical Gases

  • Oxygen (O2\text{O}_2) Characteristics and Thermal Hazards

    • Pure liquid oxygen is extremely cold (less than 183C-183\,^\circ\text{C}); direct contact causes severe freezer burns or tissue freezing.

    • Oxygen is a diatomic molecule (O2\text{O}_2) delivered in medicinal practice as pure gas or liquid.

    • Flammability versus Combustion Support: Oxygen is not flammable on its own and does not spontaneously explode. However, it strongly supports and accelerates combustion. Combustible materials burn much more violently in an oxygen-enriched environment.

    • Media depictions of oxygen cylinders randomly exploding without an ignition source or underlying structural failure are inaccurate.

  • Regulatory and Clinical Classification

    • Medicinal oxygen is classified as a drug requiring a prescription from a licensed ordering provider (Physician, Nurse Practitioner, or Physician Assistant).

    • Therapeutic Gases: Gases administered for direct patient treatment (e.g., oxygen, Heliox, inhaled nitric oxide). They require medical orders and carry clinical risks such as oxygen toxicity.

    • Nontherapeutic Gases: Gases used primarily for equipment calibration, diagnostic testing, or laboratory procedures (e.g., nitrogen, pure carbon monoxide, ethylene oxide).

  • Physical Gas Behavior and Dynamics

    • Viscosity: Heating a gas increases its viscosity, whereas heating a liquid decreases its viscosity.

    • Density: Relative mass per unit volume. Helium and oxygen are the primary clinical gases where density is leveraged therapeuticly.

    • Specific Gravity: The ratio of a gas's density relative to a standard reference gas (typically pure oxygen for gas physics, analogous to urine specific gravity relative to water in clinical laboratory settings).

    • Solubility: The physical capacity of a gas to dissolve or dissipate within a liquid solution.

    • Fluid Dynamics in Airways:

    • Bulk Flow: Gas movement from the upper airway down to the terminal bronchioles occurs in bulk flow (organized jet stream/bolus moving through branching bifurcations).

    • Molecular Diffusion: At the alveoli, gas transport transitions from bulk flow to molecular diffusion, characterized by random, spread-out molecular collision across the alveolocapillary membrane.

  • Thermodynamic States and Transitions

    • Boiling Point: Boiling requires the entire liquid volume to reach its specific boiling temperature. Evaporation occurs during boiling, but evaporation can also occur at temperatures below the boiling point at the liquid surface.

    • Humidity vs. Aerosol: Steam (hot boiling water) is never administered to patients. Cold water vapor is delivered via humidifiers. Visible liquid water droplets suspended in a gas stream are defined as aerosols.

    • Critical Temperature: The maximum temperature at which a substance can exist as a liquid regardless of the applied pressure. Liquid oxygen violently boils almost instantaneously if removed from its vacuum-insulated storage environment and exposed to ambient conditions.

    • Critical Pressure: The minimum pressure required to liquefy a gas at its critical temperature.

    • Triple Point of Water: The unique temperature and pressure condition where water coexists simultaneously in solid, liquid, and gas phases. This occurs at exactly 0.01C0.01\,^\circ\text{C}.

Healthcare Gas Classifications and Clinical Applications

  • Scope of Practice for Respiratory Therapists

    • Respiratory Therapists (RTs) manage all medical gas delivery, central piping, storage, and bedside equipment outside of specialized anesthesia suites.

    • Clinical application environments include Intensive Care Units (ICU), progressive/floor care, Neonatal Intensive Care Units (NICU), outpatient clinics, pulmonary rehabilitation, hyperbaric chambers, and emergency transport.

  • Medical Air vs. Room Air

    • Room Air: Contains approximately 21%FiO221\%\,\text{FiO}_2, 78%78\%\text{--}79%N279\%\,\text{N}_2, and variable trace gases (water vapor, carbon dioxide, pollutants).

    • Medical Air: Standardized mixture of pure 21%FiO221\%\,\text{FiO}_2 and 79%FiN279\%\,\text{FiN}_2 (filtered, oil-free, dry). It contains zero trace contaminants.

    • Density Comparison: Pure oxygen is more dense than room air or medical air because nitrogen (N2\text{N}_2) has a lower molecular weight than oxygen (O2\text{O}_2).

    • Nitrogen Behavior: Nitrogen is an inert gas that does not readily cross the alveolocapillary membrane under normal physiological conditions.

  • Therapeutic Gas Overview

    • Heliox (Helium-Oxygen Mixture):

    • Low-density gas mixture used to decrease work of breathing in upper airway obstruction, severe croup, post-extubation stridor, and severe asthma exacerbations.

    • Standard clinical concentrations (Helium % / Oxygen %):

      1. 80/20 Mixture: 80%He80\%\,\text{He} / 20%O220\%\,\text{O}_2 (Note: Delivered FiO2\text{FiO}_2 is 0.200.20, which is below room air concentration).

      2. 70/30 Mixture: 70%He70\%\,\text{He} / 30%O230\%\,\text{O}_2 (FiO2=0.30\text{FiO}_2 = 0.30).

      3. 60/40 Mixture: 60%He60\%\,\text{He} / 40%O240\%\,\text{O}_2 (FiO2=0.40\text{FiO}_2 = 0.40).

    • Inhaled Nitric Oxide (iNO / INOmax):

    • Selective pulmonary vasodilator administered in parts per million (ppm).

    • Relaxes pulmonary vascular smooth muscle to treat severe pulmonary hypertension and right ventricular failure (cor pulmonale) by reducing right heart afterload.

    • Upon entering pulmonary capillary blood, iNO is rapidly bound and neutralized by hemoglobin, preventing systemic arterial vasodilation.

    • Toxic Byproduct: iNO reacts with oxygen to form Nitrogen Dioxide (NO2\text{NO}_2), a toxic gas that can cause methemoglobinemia and pulmonary tissue injury.

    • Nitrous Oxide (N2O\text{N}_2\text{O}):

    • Anesthetic and analgesic gas ("laughing gas") commonly used in dental procedures.

    • Inhalation of commercial canisters ("whips") causes severe neurological damage and irreversible brain injury.

    • Carbogen:

    • Mixture of oxygen and carbon dioxide (e.g., 95%O295\%\,\text{O}_2 / 5%CO25\%\,\text{CO}_2 or 97%O297\%\,\text{O}_2 / 3%CO23\%\,\text{CO}_2).

    • Used clinically to treat Singultus (intractable hiccups) by stimulating central respiratory drive, and in specific cerebrovascular/retinal diagnostic procedures.

    • Ethylene Oxide (ETO):

    • Non-medicinal, gaseous chemical sterilant used for heat-sensitive and moisture-sensitive equipment (e.g., flexible fiberoptic bronchoscopes, metal fittings).

    • Toxic if inhaled; requires strict aeration protocols. (Distinct from liquid Ethanol, ETOH).

    • Carbon Monoxide (CO):

    • Administered in minute trace amounts during diagnostic pulmonary function testing to measure the Diffusion Capacity of the Lung for Carbon Monoxide (DLCO).

    • Cyclopropane:

    • Historical volatile anesthetic gas, now obsolete in patient care and restricted to biomedical engineering contexts.

  • Clinical Indications and Differential Selection

    • Upper Airway Obstruction: Benefitted most by Heliox due to its low density promoting laminar flow past structural narrowings.

    • Cor Pulmonale / Right Heart Strain: Benefitted most by Inhaled Nitric Oxide (iNO) due to selective pulmonary capillary vasodilation.

    • Singultus: Benefitted by Carbogen administration.

  • Oxygen Toxicity and Dosing Principles

    • Toxicity Factors: Driven by high FiO2\text{FiO}_2 concentration, total exposure duration, and individual genetic predisposition.

    • Continuous exposure to 100%FiO2100\%\,\text{FiO}_2 presents the highest risk for acute pulmonary tissue damage, absorption atelectasis, and oxygen toxicity.

    • Exposure to 40%FiO240\%\,\text{FiO}_2 for 24 hours can present a higher cumulative tissue exposure risk than 100%FiO2100\%\,\text{FiO}_2 for 1 hour.

    • Obstructive Pulmonary Diseases (A-B-C-D-E-F Mnemonic):

    • A: Asthma

    • B: Bronchiectasis

    • C: Cystic Fibrosis

    • E: Emphysema

    • F: Foreign body airway obstruction

    • CO2 Narcosis Risk: In patients with chronic hypercapnia (e.g., severe COPD), excessive oxygen administration can abolish hypoxic respiratory drive and worsen ventilation-perfusion matching, causing hypoventilation, lethargy, severe hypercapnia, and respiratory failure.

    • Flow Rate vs. FiO2 Differentiation:

    • Flow (LPM) and Concentration (FiO2\text{FiO}_2) are separate clinical variables.

    • An air-hungry, dyspneic patient with an SpO2\text{SpO}_2 of 99%99\% needs gas flow, not increased FiO2\text{FiO}_2.

    • Neonatal patients frequently require extremely low flow rates (e.g., 0.1L/min0.1\,\text{L/min}) but may need high concentrations (70%70\%\text{--}80%FiO280\%\,\text{FiO}_2) because their delicate tissues are highly sensitive to volumetric displacement and oxygen toxicity.

    • Life-Extending Interventions: The only two interventions clinically proven to increase overall lifespan (survival length) in severe COPD are Long-Term Oxygen Therapy (LTOT) and Smoking Cessation.

Gas Physics, Dynamics, and Measurement Standards

  • Manufacturing and Environmental Conditions

    • STPD (Standard Temperature and Pressure Dry):

    • Condition: 0C0\,^\circ\text{C}, 760mmHg760\,\text{mmHg} pressure, 0%0\% relative humidity.

    • Used in industrial manufacturing, gas volume standardization, and gas cylinder filling processes.

    • ATPD (Ambient Temperature and Pressure Dry):

    • Condition: Ambient room temperature (typically 68F68\,^\circ\text{F}\text{--}70F70\,^\circ\text{F} / 20C20\,^\circ\text{C}\text{--}21.1C21.1\,^\circ\text{C}), ambient barometric pressure, 0%0\% relative humidity.

    • Bedside clinical gas delivery directly from compressed outlets is ATPD. Wall connections explicitly state "Use No Oil" to prevent spontaneous combustion of organic lubricants under pressure.

  • Commercial Gas Production Methods

    • Fractional Distillation of Liquefied Air: Primary large-scale industrial method for producing medicinal oxygen and nitrogen. Air is cooled to a liquid state and gradually warmed; gases evaporate off at their distinct boiling points and are harvested sequentially.

    • Molecular Sieve (Physical Absorption): Air is pumped through synthetic zeolite material that physically adsorbs nitrogen, water vapor, and trace gases, passing an effluent gas containing up to 90%90\%\text{--}95%O295\%\,\text{O}_2

    • Semipermeable Membrane: Uses specialized porous plastic membranes that allow smaller oxygen molecules to diffuse through faster than nitrogen molecules, producing low-flow enriched oxygen.

  • Standards and Regulatory Agencies

    • CGA (Compressed Gas Association): Establishes safety standards, valve specifications, and cylinder color-coding systems.

    • NFPA (National Fire Protection Association): Sets fire safety regulations for bulk oxygen storage and hospital piping systems.

    • ANSI (American National Standards Institute): Oversees national engineering standards for medical devices.

    • Z79 Committee: Specialized standards committee dedicated to anesthesia and respiratory equipment specifications.

    • USP (United States Pharmacopeia): Sets purity standards for medicinal gases (e.g., USP Medical Air, USP Oxygen).

Bulk Gas Supply Systems, Storage, and Oxygen Concentrators

  • Central Piping and Working Pressure

    • Standard hospital pipeline distribution systems maintain a constant working pressure of 50PSIG50\,\text{PSIG} (pounds per square inch gauge).

    • Respiratory care equipment, bedside flowmeters, mechanical ventilators, and pneumatic devices are engineered to operate at 50PSIG50\,\text{PSIG}.

    • Zone Valves: Manual shut-off valves located throughout hospital units that allow specific floors or wards to be isolated during fires, emergency pipe breaches, or system maintenance.

  • Bulk Liquid Oxygen Systems

    • Used in medium-to-large medical centers due to spatial efficiency.

    • Liquid-to-Gas Expansion Ratio: 1Liter1\,\text{Liter} of liquid oxygen expands to equal 860Liters860\,\text{Liters} of gaseous oxygen at standard ambient conditions.

    • Storage Design: Held in large, double-walled vacuum-insulated vessels at temperatures below 183C-183\,^\circ\text{C}. Liquid oxygen is siphoned into external vaporizer coils, where ambient heat causes it to rapidly boil into gas, which is then regulated down to the 50PSIG50\,\text{PSIG} line working pressure.

    • Pressure Behavior Comparison:

    • Compressed Gas Cylinder: Pressure drops linearly in direct proportion to remaining volume.

    • Liquid Oxygen Vessel: Gauge pressure remains constant (in vapor-liquid equilibrium) until the liquid volume is completely exhausted.

  • Manifold Systems

    • A bank of multiple high-pressure cylinders (typically H/K size) linked together through a central pressure-reducing assembly.

    • Serves as a primary supply for smaller facilities or as an emergency automated backup system if the primary bulk liquid container fails.

    • Transfilling Principle: Gas flows from a higher-pressure cylinder to a lower-pressure cylinder until both reach equal pressure (equilibrium). A small cylinder at 2200PSIG2200\,\text{PSIG} can transfill a large empty cylinder until pressures equilibrate.

  • Oxygen Concentrators

    • Electrically powered home care or portable devices that generate enriched oxygen from room air without compressed gas tanks.

    • Operational Mechanism: Ambient air is drawn in via an internal compressor and passed through dual molecular sieve beds containing zeolite pellets, separating nitrogen from oxygen.

    • Oxygen Purity vs. Flow Relationship:

    • At low flow rates (11\text{--}2L/min2\,\text{L/min}), concentrators deliver 92%92\%\text{--}95%FiO295\%\,\text{FiO}_2

    • As flow rate increases (44\text{--}5L/min5\,\text{L/min} or higher), residence time in the sieve bed decreases, causing FiO2\text{FiO}_2 purity to drop (85%85\%\text{--}90%90\% or lower).

    • Requires routine intake filter cleaning. High-output models incorporate dual flowmeters to support cohabitating patients or high-flow needs. Battery-powered portable oxygen concentrators (POCs) utilize pulse-dose delivery for ambulatory home care.

High-Pressure Gas Cylinders, Color Coding, and Safety Testing

  • Cylinder Sizes and Physical Profiles

    • E Cylinder: Small, portable cylinder (knee-high); standard for patient transport, emergency carts, and ambulatory home use.

    • H/K Cylinder: Large, stationary cylinder (shoulder-high, weighing 150lbs\sim 150\,\text{lbs}); used for primary bulk manifold supply, heavy institutional delivery, or backup therapy. (Designated interchangeably as H, K, CH, or K cylinders).

  • Medical Gas Cylinder Color Coding (US Standard)

    • Oxygen (O2\text{O}_2): Green

    • Medical Air: Yellow

    • Helium (He\text{He}): Brown

    • Heliox (He/O2\text{He/O}_2): Brown body with Green shoulder

    • Nitrous Oxide (N2O\text{N}_2\text{O}): Teal / Light Blue

    • Carbon Dioxide (CO2\text{CO}_2): Gray

    • Nitrogen (N2\text{N}_2): Black

    • International Standard Note: Internationally, oxygen cylinders are frequently color-coded White.

  • Manufacturing, Specifications, and Markings

    • Manufactured under Department of Transportation (DOT) regulations:

    • DOT-3AA: Seamless alloy heat-treated steel cylinders (chrome-molybdenum steel).

    • DOT-3AL: Seamless aluminum alloy cylinders (lightweight, non-magnetic, safe for MRI suites).

    • Stamped Markings:

    • Stamped on the cylinder shoulder: DOT specification, service working pressure (e.g., 2015PSIG2015\,\text{PSIG}), manufacturer symbol, serial number, and original hydrostatic test date.

    • Special Testing Symbols:

    • Plus Sign (+): Indicates the cylinder is approved for fill pressures up to 10%10\% above its rated working pressure.       Maximum Authorized Fill Pressure=2015PSIG×1.10=2216.5PSIG2200PSIG\text{Maximum Authorized Fill Pressure} = 2015\,\text{PSIG} \times 1.10 = 2216.5\,\text{PSIG} \approx 2200\,\text{PSIG}

    • Star Symbol (* / Asterisk): Indicates the cylinder qualifies for a 10-year hydrostatic retest interval instead of the standard 5-year requirement.

    • Hydrostatic Re-testing: Cylinders are pressurized to 5/35/3 (166%166\%) of their normal service pressure (e.g., tested at 3360PSIG3360\,\text{PSIG}) inside a water jacket to measure volumetric expansion and metal fatigue.

  • Pressure Relief Safety Devices

    • Mounted on cylinder valves to prevent explosive rupture during pressure spikes or fires:

    1. Frangible Disc (Rupture Disc): A thin metal membrane that bursts at a specific high pressure limit.

    2. Spring-Loaded Valve: A valve held closed by a spring that opens to vent gas when excess pressure builds up, then reseats when pressure normalizes.

    3. Fusible Plug: A plug constructed of a low-melting-point metal alloy (e.g., Wood's metal) that melts at high temperatures (e.g., during a structure fire), venting gas safely before cylinder pressure reaches explosive limits.

Gas Safety Connection Systems (ASSS, PISS, DISS, Quick-Connects)

  • American Standard Safety System (ASSS)

    • Designed for large high-pressure cylinders (Sizes F, G, H, K, M).

    • Features large, threaded valve connections with specific thread dimensions, pitch, hand (left-handed vs. right-handed), and internal/external threading combinations (threaded nut-and-nipple design).

    • Prevents accidental attachment of incorrect equipment to high-pressure large tanks. Requires pipe tape or mechanical seating; never use grease or oil.

  • Pin Index Safety System (PISS)

    • Designed for small high-pressure cylinders (Sizes A through E).

    • Utilizes a yoke assembly with two pins that fit into specific matching holes on the cylinder valve stem.

    • Critical Pin Index Combinations:

    • Oxygen (O2\text{O}_2): Pin positions 2 and 5

    • Medical Air: Pin positions 1 and 5

    • Requires a rubber/plastic washer (gasket) between the valve stem and yoke to form an airtight seal, and an oxygen wrench/key to open the valve stem (turn counter-clockwise / "lefty-loosey" to open; clockwise / "righty-tighty" to close).

  • Diameter Index Safety System (DISS)

    • Designed for low-pressure connections (operating at pressures of 200PSIG200\,\text{PSIG} or less, typically 50PSIG50\,\text{PSIG} system outlets).

    • Found on wall station outlets, flowmeter inlets, ventilator gas supply hoses, and pressure regulators.

    • Features unique non-interchangeable thread diameters and bore sizes for each gas type to prevent cross-connection.

    • Christmas Tree Adapter: Technically termed a swivel nut nipple adapter, which threads onto a DISS male fitting on a flowmeter to attach smooth-bore oxygen tubing or nasal cannulas.

  • Quick-Connect Systems

    • Station-specific quick-latching wall station outlets (e.g., Ohmeda, Chemetron, Puritan-Bennett).

    • Utilizes gas-specific mechanical keying and color-coding to allow rapid one-handed connection and disconnection of flowmeters and equipment.

Pressure Regulators, Flowmeters, and Delivery Equipment

  • Functional Definitions

    • Reducing Valve: A mechanical device that reduces high variable cylinder pressure (2200PSIG2200\,\text{PSIG}) down to a constant, lower working pressure (50PSIG50\,\text{PSIG}).

    • Flowmeter: A device that controls and indicates gas flow rate in Liters per Minute (LPM).

    • Regulator: A combined single unit containing both a pressure-reducing valve and a flowmeter.

    • Multi-Stage Regulator: Features two or more pressure-reducing stages in series (indicated by two pressure gauges). Reduces high pressure down to an intermediate pressure, then down to working pressure (50PSIG50\,\text{PSIG}), ensuring smooth, uniform flow control.

  • Thorpe Tube Flowmeters

    • Standard vertical tapered glass or plastic tube containing a floating indicator ball.

    • The internal diameter of the tube is narrow at the bottom and gradually widens toward the top.

    • Reading Measurement: Flow rate must always be read at the CENTER of the float ball.

    • Pressure Compensated vs. Non-Pressure Compensated:

    • Pressure Compensated: The needle valve control is located downstream of the Thorpe tube. The tube remains under full 50PSIG50\,\text{PSIG} line pressure. When plugged into a wall outlet, the ball float will jump or "pop" up briefly and drop, indicating pressure compensation.

    • Non-Pressure Compensated: The needle valve is located upstream of the tube. Backpressure from downstream restrictions (e.g., kinked tubing or nebulizers) causes the float to indicate a flow rate lower than what is actually being delivered to the patient.

    • Flush Flow Setting: Opening a Thorpe tube needle valve completely past its highest marked graduation enters the flush setting, delivering 5050\text{--}80L/min80\,\text{L/min} (or max flush at 60L/min60\,\text{L/min}).

  • Bourdon Gauge Flowmeter

    • A fixed-orifice flow-indicator device incorporating a circular spring-loaded Bourdon tube pressure gauge calibrated in Liters per Minute.

    • Not affected by gravity; ideal for transport, patient positioning, or un-upright operation.

    • Disadvantage: Any downstream restriction/occlusion creates backpressure that causes the gauge to read falsely high while actual flow to the patient drops.

  • Integrated Flow Restrictor Dial Valves

    • Modern hospital "grab-and-go" transport e-cylinders incorporate built-in regulators, pressure gauges, and rotating dial-orifice flow restrictors directly onto the cylinder top, eliminating separate regulator assembly.

Clinical Cylinder Duration Calculations and Transport Safety

  • Cylinder Factor Derivations

    • Cylinder factors represent the usable volume of gas per pound per square inch (PSI) of pressure:     Tank Factor=Full Usable Volume (Liters)Full Cylinder Pressure (PSIG)\text{Tank Factor} = \frac{\text{Full Usable Volume (Liters)}}{\text{Full Cylinder Pressure (PSIG)}}

    • H/K Cylinder Factor: 3.14L/PSIG3.14\,\text{L/PSIG} (Memory aid: π=3.14\pi = 3.14, associated with high/large cylinders).

    • E Cylinder Factor: 0.283L/PSIG0.283\,\text{L/PSIG} (commonly rounded to 0.28L/PSIG0.28\,\text{L/PSIG}).

  • Cylinder Duration Calculation Formula   Duration of Flow (Minutes)=Cylinder Pressure (PSIG)×Tank FactorSet Flow Rate (L/min)\text{Duration of Flow (Minutes)} = \frac{\text{Cylinder Pressure (PSIG)} \times \text{Tank Factor}}{\text{Set Flow Rate (L/min)}}

    • Safety Margin Rule: Always ROUND DOWN to the nearest whole minute to preserve a conservative safety margin; never round up.

    • Low Pressure Limit: Cylinders at or below 500PSIG500\,\text{PSIG} are considered empty/depleted for transport purposes and must be removed from service (gauge red zone).

  • Step-by-Step Calculation Example

    • Scenario: A patient is being transported on a nasal cannula at 3L/min3\,\text{L/min} using a full E cylinder pressurized to 2200PSIG2200\,\text{PSIG}.

    • Formula Application:     Duration=2200PSIG×0.283L/PSIG3L/min\text{Duration} = \frac{2200\,\text{PSIG} \times 0.283\,\text{L/PSIG}}{3\,\text{L/min}}     Usable Gas Volume=2200×0.283=622.6Liters\text{Usable Gas Volume} = 2200 \times 0.283 = 622.6\,\text{Liters}     Duration=622.6Liters3L/min=207.53minutes\text{Duration} = \frac{622.6\,\text{Liters}}{3\,\text{L/min}} = 207.53\,\text{minutes}

    • Rounding and Conversion:     Rounded Duration=207minutes\text{Rounded Duration} = 207\,\text{minutes}     Duration in Hours=207minutes60min/hr=3.45hours3.4hours\text{Duration in Hours} = \frac{207\,\text{minutes}}{60\,\text{min/hr}} = 3.45\,\text{hours} \approx 3.4\,\text{hours}

  • Critical Clinical Transport Hazard

    • Closed Valve Stem Trapped Pressure: If a cylinder regulator valve stem is turned OFF while high-pressure gas remains trapped inside the regulator manifold, the pressure gauge will continue to read full (e.g., 2200PSIG2200\,\text{PSIG}).

    • If an RT or nurse connects a patient and turns on the flowmeter without opening the main cylinder valve stem, trapped gas will produce a brief 2-second burst of flow and then stop completely as pressure bleeds out.

    • If unobserved during transport, the patient will receive zero gas flow, leading to asphyxiation and death.

    • Safety Protocol: Always verify the main valve stem is physically turned counter-clockwise to the OPEN position, observe continuous flow, and confirm the pressure gauge actively responds before initiating patient transport.