Clinical Biochemistry 1

Clinical Biochemistry – Purpose & Scope

  • Supports diagnosis, monitoring, treatment decisions across NI’s 55 hospital trusts (≈ 4040 million tests in 21/2221/22)
  • Core lab (NHSCT) operates 24/724/7; TAT: 11 h urgent, 44 h inpatient, 2424 h routine; >3,0003{,}000 samples/day, 65+65+ assays
  • Point-of-Care Testing present in all hospitals

Specimen Pathway

  • Pre-analytical: receipt → sorting (urgent / inpatient / routine) → accessioning in LIMS → centrifugation (serum/plasma; 77 min) → automated aliquoting/tracking
  • Analytical: automation lines (Roche Cobas c8000: ISE, photometry, ECL); specialist sections (Proteins, HbA1c, Osmolality, CSF)
  • Post-analytical: instrument → middleware (Infinity) → LIMS (WinPath) → EPR (Encompass); technical validation applied before release

Major Analytical Techniques

  • Potentiometry (ISE)
    • Ions Na+Na^+, K+K^+, ClCl^-; indirect (dilution 1:311:31) measurement
    • Uses adapted Nernst equation (compares EMF of sample vs. internal standard)
    • Electrolyte exclusion effect: high lipids/protein ↓ liquid phase → falsely low results

  • Photometry / Spectrophotometry
    • Beer–Lambert law: AcA \propto c (absorbance proportional to concentration)
    • Endpoint & rate assays; bichromatic reading reduces interference
    • Example: Mg²⁺ + xylidyl blue (purple) measured at 600nm600\,\text{nm}

  • Electrochemiluminescence (ECL) Immunoassay
    • Voltage-triggered Ru/TPA reaction produces light; intensity ∝ analyte
    • Biotin–streptavidin capture on magnetic beads
    • Sandwich (e.g. TSH) → direct; Competitive (e.g. FT4) → inverse relation

  • Additional methods
    • Freezing-point depression for osmolality (formula: 2[Na]+urea+glucose2\,[Na] + \text{urea} + \text{glucose}; gap = measured – calculated)
    • Capillary/Gel electrophoresis: serum proteins, HbA1c
    • CSF xanthochromia: spectral scan for bilirubin

Quality Management

  • Internal QC (IQC)
    • Third-party controls (low/normal/high) run at start-up & throughout day
    • Acceptable if within ±2SD\pm 2\,SD of target; visualised on Levy-Jennings; Westgard rules flag shift/trend, imprecision, bias
  • External QA (EQA)
    • Blind samples from schemes (e.g. NEQAS) assess accuracy, bias, consistency; retrospective traffic-light reports
  • Calibration
    • Sets analyser–concentration relationship (typically 22-point: water ++ standard)

Error Sources & Detection

  • Biological: age, sex, posture, diet, drugs, circadian timing
  • Pre-analytical: mis-labelling, wrong tube, IV contamination, delayed transport/storage
  • Analytical: mechanical faults, reagent issues, calibration drift, method variation (seen on IQC/EQA)
  • Post-analytical: transcription/IT errors; mitigated by middleware rules, delta checks
  • Delta check: current vs. previous result; flags change > preset %
  • Clinical & phone limits: middleware blocks / prompts urgent communication when outside defined thresholds

Key Concepts / Definitions

  • Reference range: μ±2SD\mu \pm 2\,SD of healthy population (covers 95%95\%)
  • Accuracy: closeness to true value; Precision: repeatability
  • Osmolar gap: presence of unmeasured osmoles (ethanol, EG)
  • Serum indices (H, I, L): detect haemolysis, icterus, lipaemia; middleware suppresses affected results

Automation Benefits

  • High throughput, consistent quality, reduced TAT, minimized biohazard exposure, integral QC monitoring

Clinical Relevance

  • Correct results critical: errors → misdiagnosis, treatment delay or harm
  • Some tests diagnostic alone (BNP, PSA), others need context (e.g. hyponatraemia work-up)