Clinical Biochemistry 1
Clinical Biochemistry – Purpose & Scope
- Supports diagnosis, monitoring, treatment decisions across NI’s hospital trusts (≈ million tests in )
- Core lab (NHSCT) operates ; TAT: h urgent, h inpatient, h routine; > samples/day, assays
- Point-of-Care Testing present in all hospitals
Specimen Pathway
- Pre-analytical: receipt → sorting (urgent / inpatient / routine) → accessioning in LIMS → centrifugation (serum/plasma; min) → automated aliquoting/tracking
- Analytical: automation lines (Roche Cobas c8000: ISE, photometry, ECL); specialist sections (Proteins, HbA1c, Osmolality, CSF)
- Post-analytical: instrument → middleware (Infinity) → LIMS (WinPath) → EPR (Encompass); technical validation applied before release
Major Analytical Techniques
Potentiometry (ISE)
• Ions , , ; indirect (dilution ) measurement
• Uses adapted Nernst equation (compares EMF of sample vs. internal standard)
• Electrolyte exclusion effect: high lipids/protein ↓ liquid phase → falsely low resultsPhotometry / Spectrophotometry
• Beer–Lambert law: (absorbance proportional to concentration)
• Endpoint & rate assays; bichromatic reading reduces interference
• Example: Mg²⁺ + xylidyl blue (purple) measured atElectrochemiluminescence (ECL) Immunoassay
• Voltage-triggered Ru/TPA reaction produces light; intensity ∝ analyte
• Biotin–streptavidin capture on magnetic beads
• Sandwich (e.g. TSH) → direct; Competitive (e.g. FT4) → inverse relationAdditional methods
• Freezing-point depression for osmolality (formula: ; gap = measured – calculated)
• Capillary/Gel electrophoresis: serum proteins, HbA1c
• CSF xanthochromia: spectral scan for bilirubin
Quality Management
- Internal QC (IQC)
• Third-party controls (low/normal/high) run at start-up & throughout day
• Acceptable if within of target; visualised on Levy-Jennings; Westgard rules flag shift/trend, imprecision, bias - External QA (EQA)
• Blind samples from schemes (e.g. NEQAS) assess accuracy, bias, consistency; retrospective traffic-light reports - Calibration
• Sets analyser–concentration relationship (typically -point: water standard)
Error Sources & Detection
- Biological: age, sex, posture, diet, drugs, circadian timing
- Pre-analytical: mis-labelling, wrong tube, IV contamination, delayed transport/storage
- Analytical: mechanical faults, reagent issues, calibration drift, method variation (seen on IQC/EQA)
- Post-analytical: transcription/IT errors; mitigated by middleware rules, delta checks
- Delta check: current vs. previous result; flags change > preset %
- Clinical & phone limits: middleware blocks / prompts urgent communication when outside defined thresholds
Key Concepts / Definitions
- Reference range: of healthy population (covers )
- Accuracy: closeness to true value; Precision: repeatability
- Osmolar gap: presence of unmeasured osmoles (ethanol, EG)
- Serum indices (H, I, L): detect haemolysis, icterus, lipaemia; middleware suppresses affected results
Automation Benefits
- High throughput, consistent quality, reduced TAT, minimized biohazard exposure, integral QC monitoring
Clinical Relevance
- Correct results critical: errors → misdiagnosis, treatment delay or harm
- Some tests diagnostic alone (BNP, PSA), others need context (e.g. hyponatraemia work-up)