Discuss primary methods to control scatter radiation reaching the image receptor (IR).
Explain the purpose and construction of beam-restricting devices.
Compare advantages and disadvantages of various beam-restricting devices.
Describe the effect of beam restriction on image quality and patient dose.
X-ray Interactions
Interaction Mechanisms
Transmission: An X-ray may penetrate the patient without interacting and reach the IR.
Absorption: An X-ray may completely interact with patient matter and deposit all energy (Photoelectric interaction).
Scatter: An X-ray may interact with the patient and deflect, depositing part of its energy.
Scatter Radiation
Definition
Scatter: Radiation that is deflected and travels in different directions after interacting with matter.
Creates secondary radiation via Compton interactions.
Impacts image quality by causing radiographic fog on the IR.
Contributes to dose received by the radiographer.
Production of Scatter
Origin:
Comes from large body parts, expansive field sizes, and high kilovolt peak (kVp) levels.
Generated when primary X-ray photons interact with tissue and objects when the light field is improperly collimated (should be 0.5 to 1 inch from skin).
Effects of Scatter:
Scattered photons create an evenly distributed exposure (fog).
Decreases visibility of subject contrast, leading to loss of detail in images.
Effects of Scatter
Primary Contributor: Personnel exposure is primarily due to scatter.
Image Formation: X-rays that scatter in a forward direction may help in image formation, while those scattered at angles can degrade image quality and increase receptor exposure.
Factors Affecting Scatter Production
Kilovoltage (kVp)
Selection: Based on the size of the body part.
Role:
Affects beam penetrability.
Higher kVp reduces interactions, allowing more photons to pass through to the IR.
Implications:
Increased kVp results in higher Compton scatter but lower photoelectric absorption, hence reducing patient dose.
Irradiated Material
Volume & Atomic Number: Affects the amount of scatter generated during interactions.
Volume controlled by field size and patient thickness.
Effects:
Higher atomic number materials (like bone) absorb more radiation via photoelectric interactions, resulting in reduced scatter compared to soft tissues.
Patient Thickness
Larger patient thickness necessitates higher kVp for adequate penetration.
Increasing mAs can lead to increased radiation dose.
A minimal increase in kVp is often sufficient, yielding lower patient doses, but is associated with increased scatter and reduced image contrast.
Controlling Scatter
Purpose: Minimize scatter production during Compton interactions to enhance image quality.
Methods:
Beam Restriction: Reduces the number of primary and scattered photons.
Use of Grids: Absorbs scatter radiation and improves image quality.
Patient Dose Management: Number of primary photons directly affects patient dose.
Beam Restrictors/Cataloging Devices
Types of Beam Restrictors
Collimators:
Restrict primary beam to various field sizes.
Contains lead shutters at right angles, moving in opposing pairs.
Bottom Shutters: Reduce penumbra (geometric unsharpness) at beam periphery through greater distance from focal spot.
Upper Shutters: Absorb off-focus radiation to prevent it from reaching the IR.
Collimator Structure
Mirror Assembly: Mounted at a 45-degree angle within the x-ray beam path.
Ensures alignment of the light field and x-ray field due to equal distance from the source.
Includes a light source that projects through the collimator to define the field size.
Positive Beam Limitation (PBL)
Device Function: Automatically adjusts collimation to match the IR size.
Rules:
The field size should never exceed the IR size.
Must include added filtration of 1mm aluminum.
Grids
Devices positioned between patient and IR to minimize scatter radiation.
Function: Only effective means to enhance image quality post scatter photon creation.
Results in a decrease in photons interacting with the IR, enhancing overall image quality.
Alternative Beam Restrictors
Aperture Diaphragm & Cones/Cylinders
Aperture Diaphragm: Simple shape to restrict beam size.
Cones/Cylinders: Shaped devices that limit beam field to specific geometries.
Ancillary Devices
Lead Blockers/Lead Masks: Restricted beam shapes tailored for specific examinations.
Lead Blocker: A flexible sheet of lead-impregnated material that can be molded.
Lead Mask: Custom-shaped masks secured to collimators for precise beam restriction.
Overview of Beam Restriction Impact
Decoupling Scatter Production: Smaller field sizes produce fewer scatter photons.
Principal Factors of Scatter Production: kVp and type of irradiated material.
Atomic Number's Effect: Increasing atomic number correlates with reduced scatter production.
Collimator Functionality: Manages primary beam field size effectively.
Positive Beam Limitation: Standardizes field size to the IR, reducing excess exposure.