RECD Quick Reference (Real Ear to Coupler Difference)
RECD (Real Ear to Coupler Difference) – Quick Notes
Definition: RECD quantifies the acoustic difference between SPL in a standard 2cc coupler and SPL in the ear canal for a given hearing aid/earmold.
Purpose: allows REUR conversion factor to program a hearing aid in a test box rather than on a person’s ear; vent effect not measured in this process.
Key relationships:
$$RECD(f) = SPL (canal) - SPL (coupler)
Ear canal = RECD + SPL (coupler)2cc coupler simulates ear canal resonance; measurements span frequency to create a conversion factor.
Not limited to pediatrics; RECD used across ages.
Why Measure RECD
Predict real-ear output from coupler-based measurements.
Convert an audiogram (HL) to dB SPL using personalized RECD conversion factors.
Enables preprogramming of devices in a test box without patient presence.
Note: vent effects are not included in RECD.
Accuracy and Personalization
Personal canal volume and impedance affect SPL; RECD customizes conversion for precise fitting.
Differences due to canal size, perforations/PE tubes, middle-ear fluid, surgical changes, etc.
With RECD, ear canal SPL targets can be achieved within ~1 dB accuracy.
ANSI Standards (Formal RECD measurement)
Use either an ear insert or a custom earmold.
Standard requires use of HA-1 coupler; same coupler used for both measurements.
EAR insert measures ear canal resonance and the coupler’s resonance, or the custom mold measures both.
Historical cross-contamination concerns with materials like fun-tak; ANSI provides conversion factors between HA-1 and HA-2.
RECD Measurement Overview
Signal: pink noise; identical signal for both measurements.
The difference between ear canal resonance and coupler resonance provides the conversion factor.
The 2cc coupler simulates ear canal resonance when the conversion factor is added to the coupler resonance.
Terminology: HA-1 coupler resonance vs. ear canal resonance.
Step-by-Step RECD Procedure
STEP 1: Measure the coupler response using RECD transducer
Attach transducer to HA-2 coupler
On-Ear measures → RECD → coupler calibration
STEP 2: Connect the transducer to the foam insert to measure ear canal resonance
Probe tube insertion
STEP 3: Print results as graph and table
Includes real ear resonance, coupler resonance, and their difference
Reference line represents average RECD for the entered age
Check and Confirm Accuracy
The difference (ear vs. coupler) should be positive.
Negative low-frequency (LF) values suggest a slit leak; fix foam tip seal, ensure full expansion, or adjust tip size.
Aquaphor can improve foam tip seal.
Negative RECD results can occur with perforations or PE tubes due to combined outer and middle ear volumes.
If RECD < -10 dB in 4–6 kHz, probe tube blockage is likely.
Verifit 2 and WRECD
Verifit 2 uses wideband RECD (WRECD) with a 0.4 cc coupler to better mimic ear canal size.
TRIC adapter is used to attach the RECD transducer.
Expect negative LF responses on adults; ear canal resonance above 1 kHz should align with the 0.4 cc coupler response.
If WRECD above 1 kHz falls below the 0.4 cc response, this may indicate blockage or a blocked/pinched probe tube.
Documenting and Verification
Confirm accuracy by selecting an equivalent HA-1 RECD in Verifit to ensure no slit leaks or probe blockage.
Document HA-1 RECD values even if measured with HA-2 for future adult fittings unless canal changes occur.
Practical Use: RECD in Fitting and Lab Tasks
Use RECD to program devices in a test box (e.g., NAL-NL2 targets) and verify audibility.
Lab tasks commonly include:
Measure right RECD for one student and left RECD for another using inserts.
Confirm probe depth and print results (graph/table).
Fit a device in the test box using RECD values and verify targets.
For Verifit/RECD workflows, choose the correct coupler (HA-1 vs HA-2) per software version and setup recommendations.
Choosing Couplers and Software Notes
If software version is current, select HA-1 even if HA-2 is chosen on the setup screen.
Ensure consistency between the coupler used and the software’s expected RECD workflow.
Critical Concepts to Comprehend
RECD justifies improved fitting accuracy for every adult ear canal.
ANSI standards define RECD measurement workflow (ear insert or custom mold; HA-1 coupler; same coupler for both measurements).
Protocols: measure ear canal and coupler resonances with identical signals; compute the difference to obtain the conversion factor.
Interpret RECD in the context of ear canal volume variations, PE tubes, perforations, etc.
Differentiate expectations for Verifit 1 RECD vs. Verifit 2 WRECD.
Slit leaks and blocked probes adversely affect RECD; understand how to detect and remedy.