Lesson 1: Indirect BP
Lesson 1: Indirect Blood Pressure Measurement
Auscultation Method
Relies on Korotkoff sounds.
SBP: measured at the first sound.
DBP: measured when the last sound disappears.
MAP: must be calculated.
Oscillometric Method
Monitors pressure fluctuations as cuff pressure is released.
SBP: measured when oscillations first appear.
DBP: measured at minimum pressure of oscillations.
MAP: most accurate measurement.
Note: Requires pulsatile flow; unsuitable for patients on cardiopulmonary bypass or with left ventricular assist devices.
Factors Affecting Accuracy
Bladder Size:
Ideal length: 80% of extremity circumference.
Ideal width: 40% of extremity circumference.
Cuff Location:
SBP is lowest at aortic root, highest at dorsalis pedis.
Arm Positioning:
BP changes by 7.4 mmHg for every 10 cm change in height.
Complications of NIBP Measurement
Pain
Neuropathy
Measurement errors
Limb ischemia
Bruising
Avoid cuff placement over PICC lines, fractures, or AV fistulas.