XM

Electrocardiography (ECG) Notes

I. Introduction

  • The heart functions as a dual pump, circulating blood through two circuits:

    • Pulmonary circuit: oxygenates blood in the lungs and removes carbon dioxide.

    • Systemic circuit: delivers oxygen and nutrients to tissues and removes carbon dioxide.

  • Three types of cells are required for the heart to beat:

    • Rhythm generators: produce an electrical signal (SA node, the normal pacemaker).

    • Conductors: spread the pacemaker signal.

    • Contractile cells (myocardium): mechanically pump blood.

II. Electrical and Mechanical Sequence of a Heartbeat

  • Cardiac tissue possesses inherent rhythmicity or automaticity due to specialized pacemaker cells.

  • The sinoatrial node (SA node) generates an electrical signal that spreads through specific conducting pathways:

    • Internodal pathways and atrial fibers

    • Atrioventricular node (AV node)

    • Bundle of His

    • Right and left bundle branches

    • Purkinje fibers (Fig. 5.1)

  • Systole: the mechanical event of contraction of contractile cells upon the arrival of a depolarization signal.

  • Diastole: the mechanical event of relaxation of myocardial cells upon the arrival of a repolarization signal.

  • Electrical signals precede and cause the mechanical pumping action of the heart (Fig. 5.2).

  • The SA node initiates each electrical and mechanical cycle as the normal pacemaker.

  • SA node depolarization leads to atrial muscle contraction.

  • The AV node delays the electrical signal for approximately 0.20 seconds, allowing atrial contraction before relaying the signal to the ventricles.

  • The signal is then relayed to ventricles via the bundle of His, right and left bundle branches, and Purkinje fibers.

  • Purkinje fibers stimulate ventricular muscle contraction (ventricular systole).

  • Ventricles repolarize and enter diastole during ventricular systole (Fig. 5.2).

  • The autonomic nervous system modifies heart rate (beats per minute or BPM) and contraction strength through its sympathetic and parasympathetic divisions.

III. Sympathetic and Parasympathetic Influence

  • Sympathetic division:

    • Increases automaticity and excitability of the SA node, increasing heart rate.

    • Increases conductivity of electrical impulses through the atrioventricular conduction system.

    • Increases the force of atrioventricular contraction.

    • Sympathetic influence increases during inhalation.

  • Parasympathetic division:

    • Decreases automaticity and excitability of the SA node, decreasing heart rate.

    • Decreases conductivity of electrical impulses through the atrioventricular conduction system.

    • Decreases the force of atrioventricular contraction.

    • Parasympathetic influence increases during exhalation.

IV. Electrocardiogram (ECG)

  • Echoes of depolarization and repolarization are detectable via electrodes placed on the body, producing an electrocardiogram (ECG).

  • The ECG allows inference of the heart’s mechanical activity.

  • Electrical activity varies through the ECG cycle (Fig. 5.2).

  • The ECG represents electrical events (depolarization and repolarization), while ventricular systole and diastole represent mechanical events (contraction and relaxation).

  • Electrical events happen quickly; mechanical events occur slowly.

  • Mechanical events follow the electrical events that initiate them.

  • Ventricular repolarization typically begins before the completion of ventricular systole in a resting Lead II ECG.

  • ECG changes can diagnose interruptions in electrical signal generation or transmission.

  • During exercise, heart position changes, complicating voltage standardization or quantification.

V. Components of the ECG

  • The ECG pattern includes an isoelectric line (baseline), a P wave, a QRS complex, and a T wave.

  • Intervals and segments are also components of the ECG (Fig. 5.2).

  • Isoelectric Line:

    • It's the baseline, indicating periods when no electrical activity is detected by the ECG electrodes.

  • Interval:

    • Time measurement including waves and/or complexes.

  • Segment:

    • Time measurement not including waves and/or complexes.

VI. ECG Component Details and Lead II Values

  • Table 5.1 lists components, what they represent, typical durations, and amplitudes.

  • P wave:

    • Represents depolarization of the right and left atria.

    • Duration: 0.07 – 0.18 seconds.

    • Amplitude: < 0.25 millivolts.

  • QRS complex:

    • Represents depolarization of the right and left ventricles.

    • Atrial repolarization is masked by the larger QRS complex.

    • Duration: 0.06 – 0.12 seconds.

    • Amplitude: 0.10 – 1.50 millivolts.

  • T wave:

    • Represents repolarization of the right and left ventricles.

    • Duration: 0.10 – 0.25 seconds.

    • Amplitude: < 0.5 millivolts.

  • P-R interval:

    • Time from onset of atrial depolarization to onset of ventricular depolarization.

    • 0.12-0.20 seconds.

  • Q-T interval:

    • Time from the onset of ventricular depolarization to the end of ventricular repolarization, representing the refractory period of the ventricles.

    • 0.32-0.36 seconds.

  • R-R interval:

    • Time between two successive ventricular depolarizations.

    • 0.80 seconds.

  • P-R segment:

    • Time of impulse conduction from the AV node to the ventricular myocardium.

    • 0.02 – 0.10 seconds.

  • S-T segment:

    • Period representing the early part of ventricular repolarization.

    • < 0.20 seconds.

  • T-P segment:

    • Time from the end of ventricular repolarization to the onset of atrial depolarization.

    • 0.0 – 0.40 seconds.

  • Tabled values are from a typical Lead II setup (wrist and ankle electrode placement) with a subject heart rate of ~75 BPM and are influenced by heart rate and placement.

VII. Leads

  • A lead is an arrangement of two electrodes (one positive, one negative) with respect to a third (ground) electrode.

  • Electrode positions are standardized for different leads.

  • This lesson records from Lead II: positive electrode on the left ankle, negative electrode on the right wrist, and ground electrode on the right ankle.

  • The QRS complex is the dominant ECG component in a normal standard lead record.

  • In Lead II, Q and S waves are typically small and negative, while the R wave is large and positive (Fig. 5.2).

  • Normal and abnormal factors influence QRS complex duration, form, rate, and rhythm.

  • Normal factors: body size (BSA), body fat distribution, heart size (ventricular mass), heart position, metabolic rate, etc.

  • In a person with a high diaphragm, the heart's apex shifts upward and to the left, altering the "electrical picture" and decreasing R wave positivity with increased S wave negativity.

  • Athletes with larger left ventricular mass may show similar Lead II QRS complex changes without cardiac disease.

  • An abnormal inverted QRS complex results when Lead II Q, R, and S wave amplitudes are all negative.

  • Abnormal factors include hyper- and hypothyroidism, ventricular hypertrophy, morbid obesity, essential hypertension, and other pathologic states.

VIII. Effects of the Resting Respiratory Cycle on Heart Rate

  • Minor heart rate increases/decreases during the resting respiratory cycle reflect adjustments by baroreceptor reflexes to intrathoracic pressure changes (Fig. 5.4).

  • Inspiration:

    • Inspiratory muscles contract, reducing intrathoracic pressure, causing thoracic veins to expand slightly.

    • Venous pressure, venous return, cardiac output, and systemic arterial blood pressure drop.

    • The carotid sinus reflex causes a momentary increase in heart rate due to reduced baroreceptor firing.

  • Expiration:

    • Inspiratory muscles relax, and early resting expiration increases intrathoracic pressure, compressing thoracic veins.

    • Venous pressure and venous return increase momentarily, reflexively increasing heart rate.

    • Increased cardiac output and systemic arterial blood pressure increase carotid baroreceptor firing, causing heart rate to decrease.

IX. Standard ECG Grid and Values

  • Average resting heart rate for adults: 60-80 beats/min (70 bpm for males, 75 bpm for females).

  • Athletes may have slower heart rates (as low as 50 beats/min) due to larger, more efficient hearts (“left ventricular hypertrophy”).

  • Low heart rate and hypertrophy in sedentary individuals may indicate failing hearts.

  • ECGs use standardized grids (Fig. 5.4):

    • Smaller squares: 0.04 seconds on the x-axis.

    • Darker vertical lines: 0.2 seconds apart.

    • Lighter horizontal lines: 0.1 mV apart (amplitude).

    • Darker grid lines: 0.5 mV.