Study Notes for Chapter 11: Fundamentals of CT in PET/CT

Chapter 11: Fundamentals of CT in PET/CT

Objectives

  • Describe the physics processes involved in the production of x-rays

  • Describe the role of each component in the x-ray tube and its operation

  • Discuss the proper adjustment of x-ray tube voltage and current in CT

  • Name the principal parts of a CT scanner

  • Discuss the function of each CT scanner component

  • Describe how a helical CT scanner operates and the component changes that made this technology possible

  • Discuss how CT image data are acquired and processed

  • Describe the calculation process of Hounsfield units

  • Discuss the advantages of multislice CT

  • Describe CT number values assigned to various tissues and how these values are assigned into meaningful display windowing

Physics of X-Rays

  • X-Ray Production

    • Produced after the ionization of an atom leaves an inner shell (typically the K shell) with an electron vacancy.

    • Outer-shell electrons fill the inner-shell vacancy and release excess energy as an x-ray photon with specific or characteristic energy equal to the energy difference between the two shells.

  • Characteristic X-Rays

  • Bremsstrahlung Radiation

    • German word meaning "braking radiation".

    • Occurs when electrons pass very near the nucleus of an atom.

X-Ray Tube Design

  • Components of the X-Ray Tube:

    • Glass envelope provides a vacuum environment.

    • Cathode (filament within a focusing cup).

    • Anode (rotating target).

    • Window (exit point for x-rays).

X-Ray Production

  • Operation:

    • Anode is positively charged.

    • Electrons are accelerated across the vacuum tube.

    • X-rays are produced when the electrons hit the anode, mainly through Bremsstrahlung (1%) and Characteristic processes (99% produces heat).

    • CT x-ray tubes feature high-speed rotating anodes to dissipate higher heat loads.

    • Current: Three-phase rectified current heats the cathode filament, boiling off electrons.

Voltage Variation

  • kVp and Energy:

    • kVp selects the maximum energy of the x-rays produced.

    • Higher energy x-rays penetrate tissues more deeply.

    • Typical kVp range for PET/CT: 80 to 140 kVp.

    • Advantages of higher kVp:

    • Greater penetration through the body.

    • Decrease in quantum noise.

    • Reduction in beam-hardening artifacts.

    • Compromises:

    • Higher dose to patient.

    • Reduces differences in tissue densities.

Current Variation

  • Current Settings:

    • The primary setting that can be adjusted in the x-ray tube.

    • Increased current (in milliamperes, mA) results in a proportional increase in the number of x-rays at all energies.

    • Typical CT range: 50 to 400 mA.

    • A wide range is required to adjust scanning parameters for the patient, clinical application, and anatomical area.

  • Image Quality:

    • More x-rays reaching the detector results in lower quantum noise.

    • Higher mAs increases patient dose but decreases image noise and increases contrast resolution.

    • Compromises: Higher mA increases patient dose.

X-Ray Filter

  • Purpose:

    • Absorbs low-energy x-rays that do not contribute to imaging and only increase patient exposure.

    • Results in "beam hardening," providing a more uniform average energy of the beam.

    • Can reduce patient radiation exposure up to 50%.

    • A bow-tie shaped filter may be used to ensure uniform energy distribution across the detector field.

Principles of Computed Tomography

  • CT History:

    • Anatomical cross sections referred to as transverse axial tomography.

    • CT offers minimal tissue superimposition and improved contrast differences compared to planar x-ray techniques.

    • Developed by Godfrey Hounsfield in the late 1960s with the system creating projection information from multiple angles (projections).

    • Early studies in humans conducted by Hounsfield and James Ambrose in 1971; now able to differentiate tissue densities.

    • Hounsfield and Cormack awarded the 1979 Nobel Prize in Medicine.

CT Scanner Design

  • Components of the CT System:

    • Computer workstation for scanner operation.

    • Image processing computers.

    • Electronics cabinets.

    • Gantry housing key components needed for x-ray beam production, detection, and data acquisition.

    • Patient imaging table.

Detectors

  • Technology Changes:

    • Solid-state detectors feature very high efficiency in x-ray detection.

    • Multislice scanners utilize a 2D detector array that creates an electronic signal proportional to the detected x-ray energy.

Collimation

  • Purpose:

    • Protect the patient and limit the beam to the size of the active detector array.

    • Prepatient collimators: Restrict beam width and shape for a specific measurement area.

    • Must be perfectly aligned with the tube and detector array to optimize beam location.

  • Postpatient Collimators: Reduce scattered x-rays entering the detector array, enhancing image contrast.

  • Slice Thickness: Ranges from 0.3 mm to several millimeters.

  • Advantages of Thinner Collimation:

    • Less volume averaging and greater spatial resolution.

    • Fewer streaking artifacts from high-density objects.

  • Compromises:

    • Increased quantum mottle.

    • Longer scan times and possibly more slices defined by reconstruction thickness.

Rotation Speed

  • Definition:

    • Time required for the gantry to complete one revolution.

  • Effects on Patient Dose:

    • Faster rotation speeds proportionally reduce patient exposure and imaging artifacts from patient motion.

  • Typical Speeds: Common gantry rotation speeds range from 0.3 to 1.0 seconds, with table motion of about 10 to 20 mm/sec used for oncological PET/CT acquisitions.

Pitch

  • Definition:

    • Ratio of patient movement through the gantry during one 360-degree beam rotation relative to the beam collimation.

    • For single-slice CT:

    • Pitch=table movement per rotationslice collimation\text{Pitch} = \frac{\text{table movement per rotation}}{\text{slice collimation}}

    • Volume Pitch=table movement per rotationtotal slice volume\text{Volume Pitch} = \frac{\text{table movement per rotation}}{\text{total slice volume}}

  • Note: Definitions may vary for multislice CT systems.

Increment

  • Definition:

    • Distance between slices acquired.

  • Users may define how much the slices overlap or if spacing should be introduced between images after data acquisition.

Multislice Helical CT Systems

  • Development:

    • Emerged in the early 1990s to improve patient throughput and reduce motion artifacts.

    • Terminology: "Helical" is preferred as it accurately describes the gantry's 3D path around the patient.

  • Key Technical Developments:

    1. Continuous radiation from the x-ray tube.

    2. Continuous-motion gantry using slip-ring technology.

    3. Continuous motion of the imaging table.

    4. Continuous data transfer from the detector array.

Image Data Acquisition

  • Data from the helical motion of the gantry obtained as projection information representing attenuated radiation.

  • Algorithms subsequently derive the attenuation coefficient for each pixel in the image matrix.

  • Raw data can be reconstructed with different parameters to optimize diagnostic images.

  • Maximum resolution controlled by beam width, determined by detector size and collimator.

  • Typical acquisition for combined SPECT/CT or PET/CT performed in under 1 minute.

  • System slice capacity varies (8, 16, 40, 64, 128 slices or more).

CT Display

  • Wide variation in CT numbers poses challenges for digital display (range from -850 to 500, or 1350 numbers).

  • Human eye can discern approximately 30 to 100 shades of gray; digital imaging is often limited to 8 bits, assigning only 256 intensity values across the grayscale spectrum.

  • Grayscale Windowing:

    • Limited range of CT numbers assigned to display enables full black-to-white transition across a manageable 256 shades of gray.

    • Flexibility: Window centerline and window width can be repositioned on the CT number scale for optimal tissue display.

Image Quality Factors

  • Contrast resolution.

  • Spatial resolution.

  • Focal spot size.

  • Detector element size.

  • Collimation and pitch.

  • Slice thickness and increment.

  • Reconstruction algorithm and kernel.

  • Matrix size and reconstruction field of view (FOV).

  • Image noise.

CT Protocols

  • Types of protocols include:

    • Low-Dose CT for PET Attenuation Correction.

    • Diagnostic CT (Abdomen, Chest, Neck).

    • Contrast Media and Integrated PET/CT Protocols.

CT Artifacts

  • Definition:

    • Any discrepancy in pixel values observed in CT studies.

  • Types:

    • Primary sources: Operator, Scanner, Patient.

    • May be qualitative or quantitative.

CT Radiation Safety

  • Occupational Safety:

    • Implementation of lead-lined walls and control rooms separate from scanners.

  • Patient Safety:

    • Importance of patient verification to minimize unnecessary rescanning, thereby reducing radiation exposure.

CT Quality Control

  • Required by: The Joint Commission (TJC).

  • Adherence to ACR recommendations and vendor-specific tests to ensure standards are met.