Study Notes for Chapter 11: Fundamentals of CT in PET/CT
Chapter 11: Fundamentals of CT in PET/CT
Objectives
Describe the physics processes involved in the production of x-rays
Describe the role of each component in the x-ray tube and its operation
Discuss the proper adjustment of x-ray tube voltage and current in CT
Name the principal parts of a CT scanner
Discuss the function of each CT scanner component
Describe how a helical CT scanner operates and the component changes that made this technology possible
Discuss how CT image data are acquired and processed
Describe the calculation process of Hounsfield units
Discuss the advantages of multislice CT
Describe CT number values assigned to various tissues and how these values are assigned into meaningful display windowing
Physics of X-Rays
X-Ray Production
Produced after the ionization of an atom leaves an inner shell (typically the K shell) with an electron vacancy.
Outer-shell electrons fill the inner-shell vacancy and release excess energy as an x-ray photon with specific or characteristic energy equal to the energy difference between the two shells.
Characteristic X-Rays
Bremsstrahlung Radiation
German word meaning "braking radiation".
Occurs when electrons pass very near the nucleus of an atom.
X-Ray Tube Design
Components of the X-Ray Tube:
Glass envelope provides a vacuum environment.
Cathode (filament within a focusing cup).
Anode (rotating target).
Window (exit point for x-rays).
X-Ray Production
Operation:
Anode is positively charged.
Electrons are accelerated across the vacuum tube.
X-rays are produced when the electrons hit the anode, mainly through Bremsstrahlung (1%) and Characteristic processes (99% produces heat).
CT x-ray tubes feature high-speed rotating anodes to dissipate higher heat loads.
Current: Three-phase rectified current heats the cathode filament, boiling off electrons.
Voltage Variation
kVp and Energy:
kVp selects the maximum energy of the x-rays produced.
Higher energy x-rays penetrate tissues more deeply.
Typical kVp range for PET/CT: 80 to 140 kVp.
Advantages of higher kVp:
Greater penetration through the body.
Decrease in quantum noise.
Reduction in beam-hardening artifacts.
Compromises:
Higher dose to patient.
Reduces differences in tissue densities.
Current Variation
Current Settings:
The primary setting that can be adjusted in the x-ray tube.
Increased current (in milliamperes, mA) results in a proportional increase in the number of x-rays at all energies.
Typical CT range: 50 to 400 mA.
A wide range is required to adjust scanning parameters for the patient, clinical application, and anatomical area.
Image Quality:
More x-rays reaching the detector results in lower quantum noise.
Higher mAs increases patient dose but decreases image noise and increases contrast resolution.
Compromises: Higher mA increases patient dose.
X-Ray Filter
Purpose:
Absorbs low-energy x-rays that do not contribute to imaging and only increase patient exposure.
Results in "beam hardening," providing a more uniform average energy of the beam.
Can reduce patient radiation exposure up to 50%.
A bow-tie shaped filter may be used to ensure uniform energy distribution across the detector field.
Principles of Computed Tomography
CT History:
Anatomical cross sections referred to as transverse axial tomography.
CT offers minimal tissue superimposition and improved contrast differences compared to planar x-ray techniques.
Developed by Godfrey Hounsfield in the late 1960s with the system creating projection information from multiple angles (projections).
Early studies in humans conducted by Hounsfield and James Ambrose in 1971; now able to differentiate tissue densities.
Hounsfield and Cormack awarded the 1979 Nobel Prize in Medicine.
CT Scanner Design
Components of the CT System:
Computer workstation for scanner operation.
Image processing computers.
Electronics cabinets.
Gantry housing key components needed for x-ray beam production, detection, and data acquisition.
Patient imaging table.
Detectors
Technology Changes:
Solid-state detectors feature very high efficiency in x-ray detection.
Multislice scanners utilize a 2D detector array that creates an electronic signal proportional to the detected x-ray energy.
Collimation
Purpose:
Protect the patient and limit the beam to the size of the active detector array.
Prepatient collimators: Restrict beam width and shape for a specific measurement area.
Must be perfectly aligned with the tube and detector array to optimize beam location.
Postpatient Collimators: Reduce scattered x-rays entering the detector array, enhancing image contrast.
Slice Thickness: Ranges from 0.3 mm to several millimeters.
Advantages of Thinner Collimation:
Less volume averaging and greater spatial resolution.
Fewer streaking artifacts from high-density objects.
Compromises:
Increased quantum mottle.
Longer scan times and possibly more slices defined by reconstruction thickness.
Rotation Speed
Definition:
Time required for the gantry to complete one revolution.
Effects on Patient Dose:
Faster rotation speeds proportionally reduce patient exposure and imaging artifacts from patient motion.
Typical Speeds: Common gantry rotation speeds range from 0.3 to 1.0 seconds, with table motion of about 10 to 20 mm/sec used for oncological PET/CT acquisitions.
Pitch
Definition:
Ratio of patient movement through the gantry during one 360-degree beam rotation relative to the beam collimation.
For single-slice CT:
Note: Definitions may vary for multislice CT systems.
Increment
Definition:
Distance between slices acquired.
Users may define how much the slices overlap or if spacing should be introduced between images after data acquisition.
Multislice Helical CT Systems
Development:
Emerged in the early 1990s to improve patient throughput and reduce motion artifacts.
Terminology: "Helical" is preferred as it accurately describes the gantry's 3D path around the patient.
Key Technical Developments:
Continuous radiation from the x-ray tube.
Continuous-motion gantry using slip-ring technology.
Continuous motion of the imaging table.
Continuous data transfer from the detector array.
Image Data Acquisition
Data from the helical motion of the gantry obtained as projection information representing attenuated radiation.
Algorithms subsequently derive the attenuation coefficient for each pixel in the image matrix.
Raw data can be reconstructed with different parameters to optimize diagnostic images.
Maximum resolution controlled by beam width, determined by detector size and collimator.
Typical acquisition for combined SPECT/CT or PET/CT performed in under 1 minute.
System slice capacity varies (8, 16, 40, 64, 128 slices or more).
CT Display
Wide variation in CT numbers poses challenges for digital display (range from -850 to 500, or 1350 numbers).
Human eye can discern approximately 30 to 100 shades of gray; digital imaging is often limited to 8 bits, assigning only 256 intensity values across the grayscale spectrum.
Grayscale Windowing:
Limited range of CT numbers assigned to display enables full black-to-white transition across a manageable 256 shades of gray.
Flexibility: Window centerline and window width can be repositioned on the CT number scale for optimal tissue display.
Image Quality Factors
Contrast resolution.
Spatial resolution.
Focal spot size.
Detector element size.
Collimation and pitch.
Slice thickness and increment.
Reconstruction algorithm and kernel.
Matrix size and reconstruction field of view (FOV).
Image noise.
CT Protocols
Types of protocols include:
Low-Dose CT for PET Attenuation Correction.
Diagnostic CT (Abdomen, Chest, Neck).
Contrast Media and Integrated PET/CT Protocols.
CT Artifacts
Definition:
Any discrepancy in pixel values observed in CT studies.
Types:
Primary sources: Operator, Scanner, Patient.
May be qualitative or quantitative.
CT Radiation Safety
Occupational Safety:
Implementation of lead-lined walls and control rooms separate from scanners.
Patient Safety:
Importance of patient verification to minimize unnecessary rescanning, thereby reducing radiation exposure.
CT Quality Control
Required by: The Joint Commission (TJC).
Adherence to ACR recommendations and vendor-specific tests to ensure standards are met.