Ultrasound Resolution and Focusing
Axial and Lateral Resolution Background
- Resolution: The ability to create accurate images.
Axial Resolution
- Definition: Measures the ability of a system to display two structures very close together when they are parallel to the sound beam’s main axis.
- Synonyms: Longitudinal, Axial, Range, Radial, Depth.
- Units: Distance, measured in millimeters (mm).
- Typical Value: 0.05 – 0.5 mm. This range indicates normal performance for axial resolution.
- Adjustability: Not adjustable by the sonographer.
- Determined by: Spatial pulse length.
- Shorter pulses improve axial resolution.
- Numerical Values and Image Quality:
- Lower numerical values of axial resolution indicate a shorter pulse.
- Shorter pulses create more accurate images.
- Image quality is better when axial resolution has a lower numerical value.
- Formula: Axial Resolution = 2spatial pulse length (mm).
- Factors Determining Pulse Length and Better Axial Resolution:
- Shorter pulse length leads to better axial resolution.
- A short pulse is created by:
- Less ringing.
- Fewer cycles in the pulse.
- Higher frequency.
- Shorter wavelength.
- Summary of Factors for Better Axial Resolution:
- Shorter spatial pulse length.
- Shorter pulse duration.
- Higher frequencies (which result in shorter wavelengths).
- Fewer cycles per pulse (meaning less ringing).
- Lower numerical values representing the axial resolution measurement.
Lateral Resolution
- Definition: The ability to distinctly identify two structures very close together when they are perpendicular to the sound beam’s main axis.
- Synonyms: Lateral, Angular, Transverse, Azimuthal.
- Units: Distance, measured in millimeters (mm).
- Determined by: The width of the sound beam.
- Narrower beams have better lateral resolution.
- Variability with Depth:
- Beam diameter varies with depth.
- Consequently, lateral resolution changes with depth.
- Optimal Location: Lateral resolution is best at the focus of the sound beam.
- Formula: Lateral Resolution (mm) = beam diameter (mm).
- Point-Spread Artifact:
- Occurs when two side-by-side structures are closer together than the beam width.
- Results in only one wide reflection being seen on the image, rather than two distinct reflections.
Axial vs. Lateral Resolution Comparison
- Superiority: Axial resolution is generally better than lateral resolution.
- This is because ultrasound pulses are typically shorter than they are wide.
- Impact of Higher Frequencies: Higher frequencies improve both axial and lateral resolution.
- Axial Resolution: Improved overall across the image.
- Lateral Resolution: Primarily improved in the far field.
Focusing
- Purpose: Concentration of sound energy into a narrower beam.
- Effects of Focusing:
- Beam diameter in the near field and focal zone narrows.
- Near zone length is reduced.
- Beam diameter beyond the focal zone widens.
- Size of the focal zone is reduced.

Methods of Focusing
- External Focusing: Achieved with a lens.
- A lens is placed in front of the piezoelectric (PZT) crystal.
- Focuses sound waves in a manner similar to how a lens focuses light waves.
- Internal Focusing: Achieved with a curved crystal.
- A curved crystal concentrates sound energy.
- This is the most common method of fixed focusing.
- No lens is used in internal focusing.
- Synonyms: Conventional, mechanical.
- Fixed Focusing:
- Characteristics: Determined by the transducer itself; cannot be changed by the sonographer.
- Methods: Can be external (with a lens) or internal (with a curved crystal).
- Phased Array (Electronic) Focusing: Achieved with the electronics of the system.
- In phased array transducers, the system’s electronics focus the beam.
- The sonographer can dynamically adjust the focus.
- This method is never used on single crystal transducers.
- Meaning of "Phased Array": Implies adjustable or multiple focusing capabilities.
Modifications Due to Focusing
- Four distinct modifications occur when focusing is applied:
- The beam diameter in the near field and the focal zone narrows.
- The near zone length is reduced.
- The beam diameter beyond the focal zone widens.
- The size of the focal zone is reduced.