Ultrasound Resolution and Focusing

Axial and Lateral Resolution Background

  • Resolution: The ability to create accurate images.

Axial Resolution

  • Definition: Measures the ability of a system to display two structures very close together when they are parallel to the sound beam’s main axis.
  • Synonyms: Longitudinal, Axial, Range, Radial, Depth.
  • Units: Distance, measured in millimeters (mm\text{mm}).
  • Typical Value: 0.050.05 – 0.5 mm0.5\text{ mm}. This range indicates normal performance for axial resolution.
  • Adjustability: Not adjustable by the sonographer.
  • Determined by: Spatial pulse length.
    • Shorter pulses improve axial resolution.
  • Numerical Values and Image Quality:
    • Lower numerical values of axial resolution indicate a shorter pulse.
    • Shorter pulses create more accurate images.
    • Image quality is better when axial resolution has a lower numerical value.
  • Formula: Axial Resolution == spatial pulse length (mm)2\frac{\text{spatial pulse length (mm)}}{2}.
  • Factors Determining Pulse Length and Better Axial Resolution:
    • Shorter pulse length leads to better axial resolution.
    • A short pulse is created by:
      • Less ringing.
      • Fewer cycles in the pulse.
      • Higher frequency.
      • Shorter wavelength.
  • Summary of Factors for Better Axial Resolution:
    • Shorter spatial pulse length.
    • Shorter pulse duration.
    • Higher frequencies (which result in shorter wavelengths).
    • Fewer cycles per pulse (meaning less ringing).
    • Lower numerical values representing the axial resolution measurement.

Lateral Resolution

  • Definition: The ability to distinctly identify two structures very close together when they are perpendicular to the sound beam’s main axis.
  • Synonyms: Lateral, Angular, Transverse, Azimuthal.
  • Units: Distance, measured in millimeters (mm\text{mm}).
  • Determined by: The width of the sound beam.
    • Narrower beams have better lateral resolution.
  • Variability with Depth:
    • Beam diameter varies with depth.
    • Consequently, lateral resolution changes with depth.
  • Optimal Location: Lateral resolution is best at the focus of the sound beam.
  • Formula: Lateral Resolution (mm)(\text{mm}) = beam diameter (mm)=\text{ beam diameter (mm)}.
  • Point-Spread Artifact:
    • Occurs when two side-by-side structures are closer together than the beam width.
    • Results in only one wide reflection being seen on the image, rather than two distinct reflections.

Axial vs. Lateral Resolution Comparison

  • Superiority: Axial resolution is generally better than lateral resolution.
    • This is because ultrasound pulses are typically shorter than they are wide.
  • Impact of Higher Frequencies: Higher frequencies improve both axial and lateral resolution.
    • Axial Resolution: Improved overall across the image.
    • Lateral Resolution: Primarily improved in the far field.

Focusing

  • Purpose: Concentration of sound energy into a narrower beam.
  • Effects of Focusing:
    • Beam diameter in the near field and focal zone narrows.
    • Near zone length is reduced.
    • Beam diameter beyond the focal zone widens.
    • Size of the focal zone is reduced.

Comparison of poor and improved resolution images in an abdomen ultrasound, showing the effect of improved focusing.

Methods of Focusing

  • External Focusing: Achieved with a lens.
    • A lens is placed in front of the piezoelectric (PZT) crystal.
    • Focuses sound waves in a manner similar to how a lens focuses light waves.
  • Internal Focusing: Achieved with a curved crystal.
    • A curved crystal concentrates sound energy.
    • This is the most common method of fixed focusing.
    • No lens is used in internal focusing.
    • Synonyms: Conventional, mechanical.
  • Fixed Focusing:
    • Characteristics: Determined by the transducer itself; cannot be changed by the sonographer.
    • Methods: Can be external (with a lens) or internal (with a curved crystal).
  • Phased Array (Electronic) Focusing: Achieved with the electronics of the system.
    • In phased array transducers, the system’s electronics focus the beam.
    • The sonographer can dynamically adjust the focus.
    • This method is never used on single crystal transducers.
    • Meaning of "Phased Array": Implies adjustable or multiple focusing capabilities.

Modifications Due to Focusing

  • Four distinct modifications occur when focusing is applied:
    • The beam diameter in the near field and the focal zone narrows.
    • The near zone length is reduced.
    • The beam diameter beyond the focal zone widens.
    • The size of the focal zone is reduced.