Medical Gas Therapy - Egan's Chapter 42 Review
Clinical Foundations and Indications for Oxygen Therapy
Acute Hypoxemia Definition and Diagnostic Thresholds:
- Defined pathologically as an abnormally low level of oxygen in the arterial blood.
- Laboratory threshold: Partial pressure of arterial oxygen () less than .
- Pulse oximetry / saturation threshold: Arterial oxygen saturation () less than while breathing ambient room air ().
Hypercapnia Definition:
- Defined as an abnormally increased concentration of carbon dioxide () within the blood.
Cardiopulmonary Compensatory Mechanisms:
- When blood oxygen levels decrease, both the heart and lungs increase their operational workload to compensate for the tissue oxygen deficit.
- Properly titrated and applied oxygen () therapy alleviates cardiopulmonary stress and decreases overall ventilatory demand.
Assessment Methods for Supplemental Oxygen Need:
- Direct arterial blood gas analysis: Evaluation of .
- Noninvasive and co-oximetry saturation measurement: Assessment of or pulse oximetry saturation ().
- Assessment of the comprehensive clinical situation and underlying pathophysiology.
Clinical Indications Beyond Diagnostic Oxygen Levels:
- Severe physical trauma.
- Acute myocardial infarction (MI).
- Administration of anesthesia and monitoring during post-anesthesia recovery.
- Acute chest pain suggestive of cardiac ischemia.
- Documented or suspected hypoxemia.
- Sickle cell anemia crises.
Hazards and Complications of Medical Gas Therapy
Primary Hazards of Oxygen Administration:
- Oxygen toxicity.
- Retinopathy of prematurity (ROP).
- Depression of ventilation (DOV).
- Fire hazard.
- Absorption atelectasis.
Oxygen Toxicity Pathophysiology and Manifestations:
- Target Organs Affected: Primarily damages the lungs (pulmonary system) and the central nervous system (CNS).
- Determinant Risk Factors: Tissue damage is governed by two interacting variables:
- High delivered concentration of oxygen ().
- Total duration and length of exposure to elevated oxygen concentrations.
- Vicious Cycle of Oxygen Toxicity:
- Step 1: Initial oxygen toxicity occurs due to high and prolonged exposure.
- Step 2: Worsening alveolar and capillary damage leads to increased pulmonary shunting.
- Step 3: Increased shunting causes refractory low arterial oxygen tension ().
- Step 4: Clinicians raise the supplemental oxygen concentration () in an attempt to correct the hypoxemia.
- Step 5: The higher exacerbates oxygen toxicity, continuing the destructive cycle.
Retinopathy of Prematurity (ROP):
- Pathological consequence of excessive arterial oxygenation in premature infants.
- May result in permanent visual impairment and irreversible blindness.
Depression of Ventilation (DOV) Considerations:
- Clinical occurrence in Chronic Obstructive Pulmonary Disease (COPD): Depression of ventilation is not typical in all COPD patients when exposed to elevated oxygen concentrations; hypoventilation is not an absolute or universal response across the entire COPD patient population.
Mechanism of Absorption Atelectasis:
- High concentrations of inspired oxygen () wash nitrogen () out of the alveoli.
- Nitrogen normally exerts a "splinting" effect that keeps alveoli patent at low partial pressures; removing nitrogen allows rapid and complete alveolar collapse (absorption atelectasis).
Combustion and Fire Hazard Requirements:
- Oxygen vigorously accelerates combustion.
- Three components required for a fire to start and continue burning (the fire triangle):
- Oxygen ().
- Heat (ignition source).
- Fuel.
Device Classifications and Low-Flow Delivery Systems
Inspiratory Demand and Performance Classifications:
- A delivery system must deliver an output flow that meets or exceeds the patient's peak inspiratory demand to guarantee a stable, precise fraction of inspired oxygen ().
- Variable-Performance Systems: Oxygen delivery devices that provide only a fraction of the patient's inspiratory flow demand; the remainder of the inspiratory volume is drawn from ambient room air, causing to fluctuate with ventilatory pattern changes.
- High-flow criteria: A true high-flow oxygen delivery system must supply a minimum total output flow of at least .
Nasal Cannula:
- Operates as a low-flow, variable-performance device.
- At an input flow rate of , delivered is approximately ().
Simple Oxygen Mask:
- Minimum operational flow setting for adult application: .
- Flow rates below permit the accumulation and rebreathing of expired carbon dioxide () inside the mask reservoir.
General Complications and Disadvantages of Oxygen Masks:
- Localized skin irritation and breakdown over contact surfaces.
- Risk of pulmonary aspiration if emesis occurs while wearing the appliance.
- Potential for patient suffocation if gas flow is inadvertently occluded or disconnected.
Reservoir and Emergency Oxygen Delivery Devices
Partial-Rebreather Mask:
- Reservoir device equipped with an attached gas reservoir bag.
- Valve configuration: Contains valves (no one-way valves between the bag and mask or on the exhalation ports).
- Delivered oxygen concentration range: ().
Non-Rebreather Mask:
- Utilized for the administration of high concentrations of supplemental oxygen.
- Delivered oxygen concentration range: Approximately .
- Valve configuration: Incorporates one-way valves (flapper valves) that prevent exhaled gases from entering the reservoir bag and prevent ambient room air entrainment through the mask exhalation ports.
- Distinguishing feature: Unlike the partial-rebreather mask (which has zero valves), the non-rebreather mask utilizes one-way valves to prevent rebreathing and room air dilution.
High-Flow Nasal Cannula Classification:
- A high-flow nasal cannula is categorized as an active high-flow delivery device and is NOT classified as a reservoir device.
Emergency and Resuscitation Delivery Systems:
- Standard appliance used in resuscitation or emergency situations to administer close to oxygen (): Bag-mask device (BMD / BVM - Bag-Valve-Mask).
- Initial indicated during acute emergency resuscitation: ().
Air-Entrainment Systems and Flow Dynamics
Operating Principles of Air-Entrainment Masks (AEM):
- Deliver a fixed by utilizing high-velocity oxygen gas passing through a fixed jet nozzle to entrain atmospheric room air through adjustable air-entrainment ports.
- Adjusting Flowmeter Settings: Increasing the delivered oxygen flow from the flowmeter does not alter the delivered ; it proportionally increases the total output flow delivered to the patient.
Effects of Downstream Resistance and Port Obstruction:
- Back pressure generated by downstream resistance or obstructions in the tubing causes: Delivered oxygen concentration () to increase, and total delivered output flow to decrease.
- Direct obstruction of the air-entrainment ports or pathway causes: Delivered oxygen concentration to increase, total system output flow to decrease, and the delivered to become unstable or variable.
Step-by-Step Flow Calculations for Air-Entrainment Systems:
- Calculation Parameters for an AEM delivering oxygen at an input flow of :
- Air-to-Oxygen Entrainment Ratio: An of () utilizes an approximate ratio ( to ).
- Total Flow Multiplier: Total ratio parts equal .
- Total System Output Flow:
- Patient Inspiratory Demand Evaluation:
- Patient Parameters: Respiratory Rate () = ; Tidal Volume () = ().
- Minute Ventilation () Calculation:
- Estimated Peak Inspiratory Demand:
- Performance Verification: Because the total system output flow () exceeds the patient's estimated peak inspiratory demand (), the system functions as a true fixed-performance device and maintains an accurate .
Oxygen Weaning and Titration Protocols
- Weaning Criteria:
- Oxygen weaning protocols begin when clinical indicators demonstrate that oxygenation has stabilized or improved, indicating the patient no longer requires the current level of supplemental therapy.
- Weaning is executed by systematically reducing the delivered while monitoring target saturation parameters.