wk5b quality control

Learning Outcomes

  • Interpret the meaning of quality control.

  • Outline the quality control (QC) checks required for the dose calibrator.

  • Describe the quality control checks required for the gamma camera, PET camera, and the CT component of both SPECT/CT and PET/CT cameras.

Quality Control (QC) vs. Quality Assurance (QA)

  • Quality Control (QC):     * Defined as all activities designed to determine the level of quality delivered.     * Primary focus is on detecting changes in equipment performance from a baseline condition.     * Requires the meticulous keeping of records.

  • Quality Assurance (QA):     * Focussed on planning, documenting, and agreeing on a set of guidelines that are necessary to assure quality.     * Primary Goal: To ensure all procedures are appropriate and all images provide the necessary information for the specific clinical situation.     * Secondary Goal: To minimise patient radiation exposure, limit inconvenience, and optimise costs.

Definition and Importance of Quality Control (QC)

  • QC involves routine assessments of equipment performance, including:     * Dose calibrators.     * Gamma cameras.     * PET cameras.     * CT components of hybrid systems.

  • Objective: To ensure that findings on a patient scan have their origins in the patient and are not artefacts generated by the gamma/PET camera itself.

  • Accuracy: To ensure that the dose displayed on the dose calibrator is accurate.

  • Technologist Role: The technologist must be able to recognise artefacts arising from camera malfunctions or errors occurring due to dose calibrator malfunctions.

  • Mandatory Frequency: QC must be performed on each system every morning before injecting any patient.

Dose Calibrator QC Checks

  • Purpose: Performed to ensure the correct amount of radiation dose is administered to patients.

  • Frequency: Performed daily to ensure readings remain constant from day to day.

  • Measurement Source: A long-lived radioactive source, typically Cesium-137 (137Cs^{137}Cs), is used to confirm the constancy of the response.

  • Protocol: The source is measured on all radionuclide settings used clinically, and results are recorded to make any drift noticeable.

Dose Calibrator QC Schedule
  • High Voltage:     * Frequency: Daily.     * Purpose: To check constancy and ensure the correct operating voltage.

  • Zero Adjustment:     * Frequency: Daily.     * Purpose: To check that the readout is at 00 when no activity is present.

  • Background/Contamination Check:     * Frequency: Daily.     * Purpose: To detect any contamination.

  • Constancy:     * Frequency: Daily.     * Purpose: To check the stability and reproducibility of the ionisation chamber, electrometer, and calibrator nuclide settings.

  • Accuracy:     * Frequency: Yearly.     * Purpose: To check the accuracy of the activity reading.

Gamma Camera Quality Control

  • Many factors within a gamma camera contribute to the final image; a failure in any single factor will affect image quality.

  • Schedule Dependencies:     * The manufacturer's specifications.     * The needs of the individual clinical department.     * National and International standards.

Ideal QC Schedule
  • Daily:     * Visual Inspection.     * Background Check.     * Energy Window/Peak.     * Uniformity Flood.     * CT Tube Warm-up.

  • Weekly:     * High Count Uniformity Flood.     * Spatial Resolution.     * Linearity.

  • Monthly or Longer:     * Centre of Rotation (COR).     * CT Calibrations.

Specific Gamma Camera QC Procedures

Visual Inspection
  • Mechanical damage check.

  • Electrical defects check.

  • Touch pad test.

  • Background check conducted without the collimator on.

Sources Used for QC
  • Point Source:     * 1020MBq10 - 20\,MBq of Technetium-99m (99mTc^{99m}Tc).     * Source positioned at least 4×4 \times the diameter of the crystal away from the crystal.     * Provides a uniform signal across the detector.     * Acquire image for 35 million counts3 - 5 \text{ million counts}.     * Placement varies by manufacturer (Siemens uses a small plastic container; GE uses a needle cap or syringe).     * The source must not be fragmented.

  • Sheet Source:     * A solid sheet source.     * Uses a long-lived radionuclide like Cobalt-57 (57Co^{57}Co).     * 57Co^{57}Co Specifications: T1/2=271 daysT_{1/2} = 271\text{ days}; energy of 120keV120\,keV gamma.

  • Flood Tank:     * A Perspex tank filled with water.     * Approximately 500MBq500\,MBq of 99mTc^{99m}Tc is added.     * Must be mixed thoroughly and cleared of air bubbles.

Photopeak Setting
  • Performed daily, usually automatically by a pre-set program.

  • Goal: Verify that the camera is properly peaked (the photopeak of the radionuclide coincides with the preset photopeak energy window).

  • Usually performed with the same point source used for uniformity checks.

  • Consequences of Incorrect Settings: Degraded uniformity, reduced sensitivity, and increased scatter contribution to the image.

  • Causes of Photopeak Drift: Variations in temperature, high voltage changes, and photomultiplier tube (PMT) drift.

Uniformity (Daily Flood)
  • Performed daily.

  • Checks if the detector response to a uniform irradiation is uniform within defined limits.

  • Requires both visual and quantitative checks.

  • Slight variances in camera components (PMTs, electronics) add up to determine overall uniformity.

  • Images are saved to monitor changes over time and for uniformity correction.

  • Types:     * Intrinsic Uniformity: Performed with the collimator removed, usually using a point source (approx. 18MBq18\,MBq).     * Extrinsic Uniformity: Performed with the collimator on, using a sheet source or flood tank.

  • Causes of Non-Uniformity:     * Energy window not peaked properly.     * PMT malfunction (seen as non-functioning tubes in images).     * Cracked crystal (visible as distinct breaks in the flood image).     * Software errors.     * Collimator damage (visible in extrinsic floods as septal damage).     * Poor source preparation (e.g., inadequate mixing of a flood tank).

High Count Flood
  • SPECT Sensitivity: SPECT images are highly sensitive to small degrees of non-uniformity in the flood field that might not affect planar images.

  • Non-Uniformity Amplification: A 5%5\% non-uniformity in planar images can increase to a 35%35\% non-uniformity in SPECT (which is unacceptable).

  • Failure Consequence: Results in "Bullseye" or Ring Artefacts (concentric rings of alternating high and low count densities in transaxial images).

  • Frequency: Weekly or monthly.

  • Count Requirement: 50100 million counts50 - 100 \text{ million counts} (compared to the 1015 million10 - 15 \text{ million} used for daily floods).

  • Method: Can be intrinsic or extrinsic; data used for uniform correction.

Centre of Rotation (COR)
  • Definition: The single point around which detectors rotate. Ideally, this should match the centre of projections recorded by detectors at all angles.

  • Critical Alignment: The electronic COR must align with the mechanical COR for accurate image generation.

  • Causes of Shift: Electronic malfunction (most common) or mechanical malfunction.

  • Consequences of Misalignment:     * Image distortion.     * Loss of resolution.     * Ring artefacts.     * Specific Visuals: Small misalignment causes blurred images; large misalignment results in a "doughnut" appearance.

  • Frequency: Monthly on SPECT systems (less frequent on newer models).

  • Protocol:     * Performed for each camera configuration (7676^\circ, 9090^\circ, and 180180^\circ).     * Performed for each collimator used (LEHR, medium energy, high energy).     * Uses 15×1 - 5 \times 99mTc^{99m}Tc point sources of very similar activities.     * Software programs correct for pixel misalignment.

Spatial Resolution
  • Definition: The ability of a system to accurately depict two separate events as two separate entities in space, time, or energy.

  • Spatial Resolution Metric: The smallest distance between two radioactive sources where they can still be defined as two separate sources. If too close, the system resolves them as one irregular object.

  • Frequency: Monthly (or longer for newer cameras).

  • Method: Extrinsically (most common) or intrinsically using a quadrant bar phantom.

  • Causes of Loss: Septal penetration or increasing source distance from the collimator face.

  • Quadrant Bar Phantom Specifications:     * Divided into 4 quadrants.     * Each quadrant has lead strips in a linear pattern.     * Each quadrant is oriented at 9090^\circ to the adjacent quadrant.     * Lead strips are spaced at different distances in each quadrant.     * One quadrant's strips are specifically positioned too close for the camera to resolve.

Linearity
  • Definition: The ability of the camera to accurately determine photon position without displacement relative to the actual entry point.

  • Principle: A straight-line source of gamma rays should yield a straight line in the image.

  • Measurement: Uses a PLES (parallel line equal spacing) phantom or a quadrant bar phantom.

  • Success Criteria: Lines must be straight and unbroken.

PET Quality Control

  • Visual inspection of the camera.

  • Synchronisation of clocks in the department and the camera system.

  • Temperature checks: The sensitivity of the PET camera changes with variations in temperature.

  • Camera Checks:     * Constancy of detector functionality.     * Blank scan: Performed to look for evidence of defective detectors.

  • Calibration Checks (Less regular):     * Involves updating and refining calibration values.     * Uses Germanium-68 (68Ge^{68}Ge) sources.     * Sources can be inbuilt (rod source) or external (phantom or point source).

CT Quality Control for Hybrid Systems

  • Tube Conditioning (Tube Warm-up): Prevents tube cracking and electrical arcing.

  • Automatic Function Checks (Air Calibration): Ensures the response to x-ray flux is uniform among the detectors.

  • CT Number QC:     * Performed with a water-filled phantom.     * Ensures the CT number for water is accurate.     * Includes calculation of noise.     * Involves visual inspection for artefacts.