wk5b quality control
Learning Outcomes
Interpret the meaning of quality control.
Outline the quality control (QC) checks required for the dose calibrator.
Describe the quality control checks required for the gamma camera, PET camera, and the CT component of both SPECT/CT and PET/CT cameras.
Quality Control (QC) vs. Quality Assurance (QA)
Quality Control (QC): * Defined as all activities designed to determine the level of quality delivered. * Primary focus is on detecting changes in equipment performance from a baseline condition. * Requires the meticulous keeping of records.
Quality Assurance (QA): * Focussed on planning, documenting, and agreeing on a set of guidelines that are necessary to assure quality. * Primary Goal: To ensure all procedures are appropriate and all images provide the necessary information for the specific clinical situation. * Secondary Goal: To minimise patient radiation exposure, limit inconvenience, and optimise costs.
Definition and Importance of Quality Control (QC)
QC involves routine assessments of equipment performance, including: * Dose calibrators. * Gamma cameras. * PET cameras. * CT components of hybrid systems.
Objective: To ensure that findings on a patient scan have their origins in the patient and are not artefacts generated by the gamma/PET camera itself.
Accuracy: To ensure that the dose displayed on the dose calibrator is accurate.
Technologist Role: The technologist must be able to recognise artefacts arising from camera malfunctions or errors occurring due to dose calibrator malfunctions.
Mandatory Frequency: QC must be performed on each system every morning before injecting any patient.
Dose Calibrator QC Checks
Purpose: Performed to ensure the correct amount of radiation dose is administered to patients.
Frequency: Performed daily to ensure readings remain constant from day to day.
Measurement Source: A long-lived radioactive source, typically Cesium-137 (), is used to confirm the constancy of the response.
Protocol: The source is measured on all radionuclide settings used clinically, and results are recorded to make any drift noticeable.
Dose Calibrator QC Schedule
High Voltage: * Frequency: Daily. * Purpose: To check constancy and ensure the correct operating voltage.
Zero Adjustment: * Frequency: Daily. * Purpose: To check that the readout is at when no activity is present.
Background/Contamination Check: * Frequency: Daily. * Purpose: To detect any contamination.
Constancy: * Frequency: Daily. * Purpose: To check the stability and reproducibility of the ionisation chamber, electrometer, and calibrator nuclide settings.
Accuracy: * Frequency: Yearly. * Purpose: To check the accuracy of the activity reading.
Gamma Camera Quality Control
Many factors within a gamma camera contribute to the final image; a failure in any single factor will affect image quality.
Schedule Dependencies: * The manufacturer's specifications. * The needs of the individual clinical department. * National and International standards.
Ideal QC Schedule
Daily: * Visual Inspection. * Background Check. * Energy Window/Peak. * Uniformity Flood. * CT Tube Warm-up.
Weekly: * High Count Uniformity Flood. * Spatial Resolution. * Linearity.
Monthly or Longer: * Centre of Rotation (COR). * CT Calibrations.
Specific Gamma Camera QC Procedures
Visual Inspection
Mechanical damage check.
Electrical defects check.
Touch pad test.
Background check conducted without the collimator on.
Sources Used for QC
Point Source: * of Technetium-99m (). * Source positioned at least the diameter of the crystal away from the crystal. * Provides a uniform signal across the detector. * Acquire image for . * Placement varies by manufacturer (Siemens uses a small plastic container; GE uses a needle cap or syringe). * The source must not be fragmented.
Sheet Source: * A solid sheet source. * Uses a long-lived radionuclide like Cobalt-57 (). * Specifications: ; energy of gamma.
Flood Tank: * A Perspex tank filled with water. * Approximately of is added. * Must be mixed thoroughly and cleared of air bubbles.
Photopeak Setting
Performed daily, usually automatically by a pre-set program.
Goal: Verify that the camera is properly peaked (the photopeak of the radionuclide coincides with the preset photopeak energy window).
Usually performed with the same point source used for uniformity checks.
Consequences of Incorrect Settings: Degraded uniformity, reduced sensitivity, and increased scatter contribution to the image.
Causes of Photopeak Drift: Variations in temperature, high voltage changes, and photomultiplier tube (PMT) drift.
Uniformity (Daily Flood)
Performed daily.
Checks if the detector response to a uniform irradiation is uniform within defined limits.
Requires both visual and quantitative checks.
Slight variances in camera components (PMTs, electronics) add up to determine overall uniformity.
Images are saved to monitor changes over time and for uniformity correction.
Types: * Intrinsic Uniformity: Performed with the collimator removed, usually using a point source (approx. ). * Extrinsic Uniformity: Performed with the collimator on, using a sheet source or flood tank.
Causes of Non-Uniformity: * Energy window not peaked properly. * PMT malfunction (seen as non-functioning tubes in images). * Cracked crystal (visible as distinct breaks in the flood image). * Software errors. * Collimator damage (visible in extrinsic floods as septal damage). * Poor source preparation (e.g., inadequate mixing of a flood tank).
High Count Flood
SPECT Sensitivity: SPECT images are highly sensitive to small degrees of non-uniformity in the flood field that might not affect planar images.
Non-Uniformity Amplification: A non-uniformity in planar images can increase to a non-uniformity in SPECT (which is unacceptable).
Failure Consequence: Results in "Bullseye" or Ring Artefacts (concentric rings of alternating high and low count densities in transaxial images).
Frequency: Weekly or monthly.
Count Requirement: (compared to the used for daily floods).
Method: Can be intrinsic or extrinsic; data used for uniform correction.
Centre of Rotation (COR)
Definition: The single point around which detectors rotate. Ideally, this should match the centre of projections recorded by detectors at all angles.
Critical Alignment: The electronic COR must align with the mechanical COR for accurate image generation.
Causes of Shift: Electronic malfunction (most common) or mechanical malfunction.
Consequences of Misalignment: * Image distortion. * Loss of resolution. * Ring artefacts. * Specific Visuals: Small misalignment causes blurred images; large misalignment results in a "doughnut" appearance.
Frequency: Monthly on SPECT systems (less frequent on newer models).
Protocol: * Performed for each camera configuration (, , and ). * Performed for each collimator used (LEHR, medium energy, high energy). * Uses point sources of very similar activities. * Software programs correct for pixel misalignment.
Spatial Resolution
Definition: The ability of a system to accurately depict two separate events as two separate entities in space, time, or energy.
Spatial Resolution Metric: The smallest distance between two radioactive sources where they can still be defined as two separate sources. If too close, the system resolves them as one irregular object.
Frequency: Monthly (or longer for newer cameras).
Method: Extrinsically (most common) or intrinsically using a quadrant bar phantom.
Causes of Loss: Septal penetration or increasing source distance from the collimator face.
Quadrant Bar Phantom Specifications: * Divided into 4 quadrants. * Each quadrant has lead strips in a linear pattern. * Each quadrant is oriented at to the adjacent quadrant. * Lead strips are spaced at different distances in each quadrant. * One quadrant's strips are specifically positioned too close for the camera to resolve.
Linearity
Definition: The ability of the camera to accurately determine photon position without displacement relative to the actual entry point.
Principle: A straight-line source of gamma rays should yield a straight line in the image.
Measurement: Uses a PLES (parallel line equal spacing) phantom or a quadrant bar phantom.
Success Criteria: Lines must be straight and unbroken.
PET Quality Control
Visual inspection of the camera.
Synchronisation of clocks in the department and the camera system.
Temperature checks: The sensitivity of the PET camera changes with variations in temperature.
Camera Checks: * Constancy of detector functionality. * Blank scan: Performed to look for evidence of defective detectors.
Calibration Checks (Less regular): * Involves updating and refining calibration values. * Uses Germanium-68 () sources. * Sources can be inbuilt (rod source) or external (phantom or point source).
CT Quality Control for Hybrid Systems
Tube Conditioning (Tube Warm-up): Prevents tube cracking and electrical arcing.
Automatic Function Checks (Air Calibration): Ensures the response to x-ray flux is uniform among the detectors.
CT Number QC: * Performed with a water-filled phantom. * Ensures the CT number for water is accurate. * Includes calculation of noise. * Involves visual inspection for artefacts.