Interventional Radiology Notes
Interventional Radiology
Objectives
- Radiation Protection:
- Describe measures for radiation protection for patients and personnel during Interventional Radiology (IR).
- Minimally Invasive Procedures:
- Explain why minimally invasive procedures are often more beneficial than traditional surgeries.
- Contrast Media:
- Discuss advantages of non-ionic contrast over ionic contrast.
- Equipment:
- Describe the equipment found in an interventional suite.
Digital Fluoroscopy
- Similar to traditional fluoroscopy but includes:
- Additional computers and monitors.
- Ability to control the fluoroscopy unit from outside the room.
- Key functionalities:
- Data acquisition.
- Post-processing.
- Image display.
- Control Devices:
- Trackball.
- Joystick or mouse.
- X-ray Tube:
- Typically located under the couch.
- Operates in general radiographic mode as pulses.
- Pulse Rates:
- Common rates are 1/sec or 10/sec.
- Each image is projected until a new image is prepared.
- Interrogation Time:
- Time required for the x-ray tube to switch on and reach adequate kVp and mA levels for the next exposure.
- Extinction Time:
- Time required for the exposure to switch off.
- High-Frequency Generators:
- DF systems use high-frequency generators with interrogation/extinction times of <1ms.
- Duty Factor (DF):
- The fraction of time the imaging system is exposing.
- Example: A fluoro machine exposing 10ms every second has a DF of 10%.
- Dose Reduction:
- Low DF decreases patient dose.
- Pulsing at the lowest rate possible is recommended.
Image Receptor Types in Digital Fluoroscopy
- Charge-Coupled Device (CCD)
- Flat Panel Detector
CCD
- Mechanism:
- A layer of crystalline silicon, when illuminated by output phosphor, creates an electrical signal.
- A pixel at the base of the Detector Element (DEL) captures this signal.
- Automatic Brightness Control (ABC):
- When ABC is used, the entire CCD signal is sampled and drives the ABC system.
- Benefits:
- Small size.
- Does not demonstrate distortion.
- Higher Detective Quantum Efficiency (DQE) and Signal-to-Noise Ratio (SNR).
- Good spatial resolution if pixel size is small.
- Unlimited life.
- No maintenance.
- Linear response.
Flat Panel Detector
- Capture Elements:
- Cesium Iodide (CsI) or amorphous Silicon (a-Si) with a Thin-Film Transistor (TFT).
- Function:
- Works like a flat panel detector in radiography but with faster interrogation/extinction times.
- Advantages:
- Smaller/lighter and more easily manipulated than Image Intensifiers (II).
- Distortion-free images.
- High DQE.
- Improved consistent image quality.
- Rectangular image area.
- Insensitive to external magnetic fields, allowing for image-guided catheter navigation.
- A magnetic tip catheter is introduced into a vessel and manipulated using external steering magnets.
What is Interventional Radiology?
- Interventional radiology is a specialized field using imaging guidance to perform minimally invasive procedures.
Types of Interventional Radiology (IR) Procedures
- Angiography (1930s):
- Opacification of vessels through injection of contrast media.
- Transbrachial (1960s).
- Transfemoral (1960s).
- Opacification of vessels through injection of contrast media.
- Angioplasty
- Thrombolysis
- Embolization
- Vascular Stents
- Biopsies
- Completed through vascular access.
Digital Subtraction Angiography (DSA)
Minicomputers and microprocessors in Digital Radiography (DR) are important for controlling:
- Matrix size
- Dynamic range
- Image acquisition rate
Analog-to-Digital Converter (ADC):
- For data to be compatible, the ADC must have the same dynamic range as the digital system.
- Electrical signal → ADC → Digital Signal → memory → matrix.
- If the image storage is in the primary memory, acquisition and transfer can be as fast as 30 images/sec.
- As matrix size doubles, the time for image acquisition/transfer takes 4x as long, which can limit temporal resolution or frame rate.
Digital Subtraction Methods
- Temporal, energy, or hybrid subtraction.
Temporal Subtraction
- Most common type of DSA.
- Uses computer-assisted techniques.
- Mask Mode:
- A mask image is obtained before contrast injection.
- Contrast is injected, and subsequent images are acquired.
- The computer recognizes the initial mask image and removes that data from the subsequent images, leaving only the contrast-enhanced anatomy.
- Subsequent images can be taken manually or preprogrammed to enhance the exam.
- Remasking is necessary if movement occurs.
- Image Integration:
- Several video frames can be summed in memory to create a single image to compensate for noise due to rapid imaging (33 microseconds).
- This process increases dose by increasing pulses.
Time Interval Mode
- Each subtracted image requires its own new mask.
- Commonly used in cardiac evaluation.
Digital Subtraction Artifacts
- Misregistration Artifact:
- Occurs when motion happens between the mask image and the subsequent image.
- The anatomy is not present in the same pixel in both images.
- Correction:
- Can be corrected through reregistration of the mask by shifting the mask one or more pixels to achieve superimposition again.
- Correction:
- Energy Subtraction:
- Uses two different x-ray beams of two different emission spectrums alternately to provide a subtraction image resulting from differences in photoelectric (PE) interactions.
- Requires alternating pulsing of different kVp beams.
- Hybrid Subtraction:
- Combination of temporal and energy subtraction.
- Mask and subsequent images are created using the energy subtraction technique.
- Produces the highest contrast images.
Roadmapping
- Used in Digital Subtraction Angiography.
- Mask image is acquired and stored.
- Contrast is injected.
- Subsequent subtracted images are obtained.
- The catheter is entered into the vessel.
- An image is formed by the subtraction of the second mask (the vessel with contrast) so the catheter can be visualized advancing.
- It appears as a black guidewire in a white vessel.
Interventional Radiology Suite
Personnel and Equipment
- Expensive, advanced radiographic and fluoroscopic equipment.
- Ceiling-mounted x-ray tube and C-arm are common.
- X-ray tube:
- Designed for magnification and repetitive imaging.
- Large anode and small target angle.
- Small focal spot allows good resolution for small arteries.
- High-frequency generators are popular, but some cases require 3-phase 12-pulse generators.
- The Room:
- Large rooms to accommodate equipment needs.
- Door to allow easy transfer of hospital bed.
- The Personnel:
- Specialized radiographer (2-3 in the room).
- ARRT offers Cardiovascular and Interventional Exams.
- Interventional Radiologist.
- Radiology Nurse.
- Anesthesiologist (when anesthesia is necessary).
- Specialized radiographer (2-3 in the room).
Arterial Access Procedure
- Insertion of needle with a stylet.
- Stylet removed.
- Insertion of guidewire.
- Removal of needle.
- Catheter threaded onto guidewire.
- Common Femoral Artery (CFA) is the most often accessed artery.
Consumable Equipment Details
Guidewires
- Made of stainless steel with a hydrophilic polymer coating.
- Classified by length, size, stiffness, and coating.
- May be J-tipped for areas of arteriosclerosis.
Catheters
- Diameter described in French units (3 Fr = 1mm).
- Come in different shapes.
- Have side holes for contrast injection without whiplash.
- Once inserted, the guidewire can be removed.
- Must be flushed immediately to prevent clotting (heparinized saline).
Patient Preparation and Monitoring
Pre-Procedure
- Patient is referred.
- Meets with interventional radiologist.
- History and physical are acquired.
- Patient begins preparation, including IV hydration and a clear liquid diet.
Monitoring Throughout the Procedure
- Blood Pressure (BP).
- Pulse oximetry.
- Electrocardiogram (ECG).
Post-Procedure
- Compression to access site.
- Immobilization and monitoring for several hours.
Risks of Arteriography
- Continuation of bleeding.
- Contrast media reactions.
- Kidney failure.
Risk Minimization
- Medical examination.
- Allergy history evaluation.
- Kidney function labs.
- Medication evaluation.
Contrast Media: Ionic vs. Non-ionic
- Risks associated with different types of contrast media.
Summary of Interventional Imaging Benefits
- Less invasive
- Less expensive
- Faster
- Often no hospital stay
- Has replaced many surgical procedures
- Highly reliant on skilled interventional radiologists and technologists
- Highly team-dependent