Image Guided Radiotherapy (IGRT) – Routine 2D X-ray Image Verification
Introduction to IGRT & 2D X-ray Image Verification
Image Guided Radiotherapy (IGRT)- Definition: Use of medical imaging to guide ever-increasingly accurate radiotherapy (RT) at all stages of the patient pathway.
Stages impacted
Initial staging & disease identification
Enhanced target delineation (contouring/CT data acquisition)
On-treatment image verification & motion management
Treatment delivery monitoring
Post-treatment follow-up / disease surveillance
2D X-ray Image Verification- Core component of IGRT; historically the routine method for verifying patient positioning.
Images acquired in the treatment room (less than or equal to 2 extra minutes)
Compares verification image(s) with reference image(s) (DRR, simulation film or planning CT projection).
Modalities
MV (megavoltage) EPID images
kV (kilovoltage) onboard imager (OBI) images
Purpose- Detect & correct translational / rotational setup errors.
Confirm internal anatomy alignment (bones, fiducials, sometimes soft tissue).
Imaging Technologies Along the RT Pathway
Planar (2D) X-ray Imaging- Single-view planar MV or kV
2D Stereoscopic Imaging- Orthogonal pair (AP & Lat) or oblique pair (non-orthogonal) providing 3-D spatial information
3-D Volumetric X-ray Imaging- kV CBCT, MV CT
Non-ionising Imaging- Ultrasound (US), surface/visual systems, implanted electromagnetic transponders, respiratory surrogates
Motion-management add-ons- 4-D imaging, gating, tracking, breath-hold techniques
Categories of On-treatment 2D Verification
2D Planar Imaging – single projection (MV or kV)
2D Stereoscopic Orthogonal Imaging – two projections 90 degrees apart (AP + Lat)
2D Stereoscopic Oblique Imaging – two non-orthogonal projections (e.g., gantry-mounted kV-kV pair)
Routine 2D Verification Workflow
BEFORE you begin
Review & understand the imaging protocol for the site/fraction.
Confirm acquisition parameters (kV, mAs, MU, gantry angle, panel position, FFD).
Check that the reference image (DRR or CT projection) is correct & matches treatment plan.
Adjust window & level on both reference & verification images to optimise bony contrast.
Acquire verification image(s) -> perform image registration (bony match or fiducials) -> quantify isocentre offsets -> apply couch shift (if less than or equal to tolerance) or seek review.
Bony Match Technique
Principle: Rigid registration of stable radiopaque skeletal landmarks because bone does not deform significantly with respiration or filling.
Two match types
Translational alignment (x, y, z) – always performed.
Rotational alignment (yaw, pitch, roll) – assessed if equipment allows 6-DoF couches or noted for physician review.
Structures employed by anatomical region:
Central Nervous System (CNS)
Orbital ridges, nasal septum
Inner skull vault border, frontal sinuses
Zygoma, occiput, pituitary fossa
Head & Neck (H&N)
Nasal septum, sinuses, maxilla
Base of skull, pituitary fossa
Cervical vertebral bodies, clavicles (ant/post borders)
Posterior wall of trachea
Breast
Breast outline; humeral head
Lung area & anterior lung border (central lung distance, central flash)
Cervical spine, inferior central margin
Chest (Thorax)
Medial & lateral rib edges
Clavicles, sternum, anterior chest wall
Carina & trachea
Lateral & anterior edges of vertebral bodies
5 & 8. Extremities (Upper & Lower)
Edges of long bones (medial/lateral)
Joint spaces, tuberosities, protuberances
Surgical implants/defects (if present)
Abdomen / Abdopelvis
Vertebral bodies (cervical to lumbar): lateral, anterior, posterior borders
Spinous processes
Superior & inferior vertebral endplates
Pelvis
Iliac crests, sacrum
Acetabulum, obturator foramen
Pubic symphysis & superior pubic ramus
Lumbar vertebral bodies
Bony Match Practices (Visualisation Tools)
Reference vs. Verification display options:- Side-by-side (spyglass), overlay (checkerboard), split-screen, subtraction.
Start coarse -> refine; zoom & pan as needed.
Unmatched vs. fully matched examples highlight impact on translational vectors.
Isocentre Offsets & 6 Degrees of Freedom (6-DoF)
Three translational directions
Left/Right (X)
Sup/Inf (Y)
Ant/Post (Z)
Three rotational directions (about isocentre)
Yaw – rotation around the cranio-caudal (Y) axis
Pitch – rotation around the left-right (X) axis
Roll – rotation around the ant-post (Z) axis
Offsets reported in mm (translations) & degrees (rotations).
Typical tolerance values (site- & protocol-specific):- less than or equal to 2 mm translation, less than or equal to 1 degree rotation -> treat; else apply shift or escalate.
2D Planar Single-Image Matching
Provides 2 translational & 1 rotational dimension per image.
Example (single AP kV):- Visible: Left/Right, Sup/Inf, Yaw.
Hidden: Ant/Post, Pitch, Roll (require lateral image).
Recommended: acquire orthogonal pair for complete 3-D info.
2D Stereoscopic Orthogonal Imaging Strategy
Two images 90 degrees apart share the Sup/Inf dimension.
Workflow tips
Match one dimension at a time.
Know which axis each image controls:
AP: L/R, Sup/Inf, Yaw
Lat: Ant/Post, Sup/Inf, Pitch
Always verify the shared dimension (Sup/Inf) in both images after any adjustment.
Address translations before rotations for clarity.
Hangling Independent Dimensions
Variation visible only in L/R, then adjust couch X.
Variation visible only in Ant/Post, then adjust couch Z.
Use image pair to decouple axes & avoid over-correction.
Rotational Offset Assessment
Only two rotations can be clearly visualised with orthogonal 2-D imaging; the third often inferred.
Yaw: easiest on AP view, subtle on Lat.
Pitch: easiest on Lat view, subtle on AP.
Roll: rarely obvious; look for asymmetric rib or pelvic wing appearance.
If significant roll suspected, volumetric imaging or 6-DoF correction recommended.
Combined Translational & Rotational Errors
Real-world setups show mixture of all 6 DoF.
Be systematic:
Perform translational match in primary image.
Cross-check shared dimension.
Evaluate rotations (yaw/pitch) -> decide if within tolerance or requires correction.
Document residuals.
Fiducial Matching
Implanted or superficial markers (e.g., prostate gold seeds, liver coils).
Used only for translational shifts in routine 2-D matching.
Rotations still assessed against adjacent bony anatomy.
Advantages: internal target surrogate independent of skeletal motion.
Cautions: migration, marker clustering, radiographic visibility (kV > MV).
Practical / Ethical / Safety Considerations
ALARA: additional imaging dose must be justified by setup accuracy gains.
Imaging frequency: daily vs. weekly – depends on site motion & protocol.
Documentation: record offsets, applied shifts, staff initials, and patient response.
Training & competency: therapists must understand anatomy, image quality factors, and 6-DoF corrections.
Patient comfort & immobilisation: proper devices reduce inter-fraction variability and imaging burden.
Real-world Relevance & Evolution
Transition from 2-D to 3-D CBCT for many sites, yet 2-D remains valuable for speed, dose and redundancy.
Integration with adaptive RT workflows demands accurate baseline 2-D verification skills.